Non-invasive continuous blood pressure monitoring: a review of current applications
Received date: 20 Mar 2012
Accepted date: 15 Nov 2012
Published date: 05 Mar 2013
Copyright
Blood pressure monitoring has come a long way from the initial observations made by Reverend Hales in the 18th century. There are none that deny the importance of monitoring perioperative blood pressure; however, the limited ability of the current prevalent technology (oscillometric blood pressure monitoring) to offer continuous blood pressure measurements leaves room for improvement. Invasive monitoring is able to detect beat-to-beat blood pressure measurement, but the risks inherent to the procedure make it unsuitable for routine use except when this risk is outweighed by the benefits. This review focuses on the discoveries which have led up to the current blood pressure monitoring technologies, and especially the creation of those offering non-invasive but continuous blood pressure monitoring capabilities, including their methods of measurement and limitations.
Elena Chung , Guo Chen , Brenton Alexander , Maxime Cannesson . Non-invasive continuous blood pressure monitoring: a review of current applications[J]. Frontiers of Medicine, 0 , 7(1) : 91 -101 . DOI: 10.1007/s11684-013-0239-5
1 |
American Society of Anesthesiologists. Standards of the American Society of Anesthesiologists: Standards for Basic Anesthetic Monitoring.; Available from: http://www.asahq.org/For-Healthcare-Professionals/~/media/For%2520Members/documents/Standards%2520Guidelines%2520Stmts/Basic%2520Anesthetic%2520Monitoring%25202011.ashx (Accessed on <month>January</month><day>12</day>, 2012)
|
2 |
Cannesson M, Pestel G, Ricks C, Hoeft A, Perel A. Hemodynamic monitoring and management in patients undergoing high risk surgery: a survey among North American and European anesthesiologists. Crit Care2011; 15(4): R197
|
3 |
Bijker JB, van Klei WA, Kappen TH, van Wolfswinkel L, Moons KG, Kalkman CJ. Incidence of intraoperative hypotension as a function of the chosen definition: literature definitions applied to a retrospective cohort using automated data collection. Anesthesiology2007; 107(2): 213-220
|
4 |
Bijker JB, Persoon S, Peelen LM, Moons KG, Kalkman CJ, Kappelle LJ, van Klei WA. Intraoperative hypotension and perioperative ischemic stroke after general surgery: a nested case-control study. Anesthesiology2012; 116(3): 658-664
|
5 |
Monk TG, Saini V, Weldon BC, Sigl JC. Anesthetic management and one-year mortality after noncardiac surgery. Anesth Analg2005; 100(1): 4-10
|
6 |
Saager L, Greenwald SD, Kelley SD, Schubert A, Sessler DI.. Duration of a “triple low” of blood pressure, BIS, and anesthestic concentration predicts poor outcome. Anesthesiology2009; 111: A880
|
7 |
Tassoudis V, Vretzakis G, Petsiti A, Stamatiou G, Bouzia K, Melekos M, Tzovaras G. Impact of intraoperative hypotension on hospital stay in major abdominal surgery. J Anesth2011; 25(4): 492-499
|
8 |
Ramsey M 3rd. Blood pressure monitoring: automated oscillometric devices. J Clin Monit1991; 7(1): 56-67
|
9 |
Smulyan H, Safar ME. Blood pressure measurement: retrospective and prospective views. Am J Hypertens2011; 24(6): 628-634
|
10 |
Kroeker EJ, Wood EH. Comparison of simultaneously recorded central and peripheral arterial pressure pulses during rest, exercise and tilted position in man. Circ Res1955; 3(6): 623-632
|
11 |
Bailey RH, Knaus VL, Bauer JH. Aneroid sphygmomanometers. An assessment of accuracy at a university hospital and clinics. Arch Intern Med1991; 151(7): 1409-1412
|
12 |
Marks LA, Groch A. Optimizing cuff width for noninvasive measurement of blood pressure. Blood Press Monit2000; 5(3): 153-158
|
13 |
Finnie KJ, Watts DG, Armstrong PW. Biases in the measurement of arterial pressure. Crit Care Med1984; 12(11): 965-968
|
14 |
Glasser SP, Ramsey MR 3rd. An automated system for blood pressure determination during exercise. Circulation1981; 63(2): 348-353
|
15 |
Bur A, Herkner H, Vlcek M, Woisetschläger C, Derhaschnig U, Delle Karth G, Laggner AN, Hirschl MM. Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients. Crit Care Med2003; 31(3): 793-799
|
16 |
Alpert BS. Oscillometric blood pressure values are algorithm-specific. Am J Cardiol2010; 106(10): 1524, author reply 1524-1525
|
17 |
Amoore JN, Geake WB, Scott DH. Oscillometric non-invasive blood pressure measurements: the influence of the make of instrument on readings? Med Biol Eng Comput1997; 35(2): 131-134
|
18 |
Gorback MS, Quill TJ, Graubert DA. The accuracy of rapid oscillometric blood pressure determination. Biomed Instrum Technol1990; 24(5): 371-374
|
19 |
Sy WP. Ulnar nerve palsy possibly related to use of automatically cycled blood pressure cuff. Anesth Analg1981; 60(9): 687-688
|
20 |
Bause GS, Weintraub AC, Tanner GE. Skin avulsion during oscillometry. J Clin Monit1986; 2(4): 262-263
|
21 |
Scheer B, Perel A, Pfeiffer UJ. Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care2002; 6(3): 199-204
|
22 |
Brzezinski M, Luisetti T, London MJ. Radial artery cannulation: a comprehensive review of recent anatomic and physiologic investigations. Anesth Analg2009; 109(6): 1763-1781
|
23 |
Hamilton WF. D.P., An experimental study of the standing waves in the pulse propagated through the aorta. Am J Physiol1938; (125): 48-59
|
24 |
O’Rourke MF, Pauca A, Jiang XJ. Pulse wave analysis. Br J Clin Pharmacol2001; 51(6): 507-522
|
25 |
Avolio AP, Van Bortel LM, Boutouyrie P, Cockcroft JR, McEniery CM, Protogerou AD, Roman MJ, Safar ME, Segers P, Smulyan H. Role of pulse pressure amplification in arterial hypertension: experts’ opinion and review of the data. Hypertension2009; 54(2): 375-383
|
26 |
Laurent S, Cockcroft J, Van Bortel L, Boutouyrie P, Giannattasio C, Hayoz D, Pannier B, Vlachopoulos C, Wilkinson I, Struijker-Boudier H ; European Network for Non-invasive Investigation of Large Arteries. Expert consensus document on arterial stiffness: methodological issues and clinical applications. Eur Heart J2006; 27(21): 2588-2605
|
27 |
Lamia B, Chemla D, Richard C, Teboul JL. Clinical review: interpretation of arterial pressure wave in shock states. Crit Care2005; 9(6): 601-606
|
28 |
Cannesson M, Le Manach Y, Hofer CK, Goarin JP, Lehot JJ, Vallet B, Tavernier B. Assessing the diagnostic accuracy of pulse pressure variations for the prediction of fluid responsiveness: a “gray zone” approach. Anesthesiology2011; 115(2): 231-241
|
29 |
Peňáz J. Photoelectric measurement of blood pressure, volume and flow in the finger. Digest of the 10th International Conference on Medical and Biological Engineering-Dresden, 1973
|
30 |
Imholz BP, Wieling W, van Montfrans GA, Wesseling KH. Fifteen years experience with finger arterial pressure monitoring: assessment of the technology. Cardiovasc Res1998; 38(3): 605-616
|
31 |
Wesseling KH, Settels JJ, van der Hoeven GM, Nijboer JA, Butijn MW, Dorlas JC. Effects of peripheral vasoconstriction on the measurement of blood pressure in a finger. Cardiovasc Res1985; 19(3): 139-145
|
32 |
Kemmotsu O, Ueda M, Otsuka H, Yamamura T, Winter DC, Eckerle JS. Arterial tonometry for noninvasive, continuous blood pressure monitoring during anesthesia. Anesthesiology1991; 75(2): 333-340
|
33 |
Hansen S, Staber M. Oscillometric blood pressure measurement used for calibration of the arterial tonometry method contributes significantly to error. Eur J Anaesthesiol2006; 23(9): 781-787
|
34 |
Janelle GM, Gravenstein N. An accuracy evaluation of the T-Line Tensymeter (continuous noninvasive blood pressure management device) versus conventional invasive radial artery monitoring in surgical patients. Anesth Analg2006; 102(2): 484-490
|
35 |
Dueck R, Goedje O, Clopton P. Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. J Clin Monit Comput2012; 26(2): 75-83
|
36 |
Lane JD, Greenstadt L, Shapiro D, Rubinstein E. Pulse transit time and blood pressure: an intensive analysis. Psychophysiology1983; 20(1): 45-49
|
37 |
Marie GV, Lo CR, Van Jones J, Johnston DW. The relationship between arterial blood pressure and pulse transit time during dynamic and static exercise. Psychophysiology1984; 21(5): 521-527
|
38 |
Drzewiecki GM, Melbin J, Noordergraaf A. Arterial tonometry: review and analysis. J Biomech1983; 16(2): 141-152
|
39 |
Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F. Continuous non-invasive arterial pressure measurement: evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim2010; 29(7-8): 530-535
|
40 |
Imholz BP, van Montfrans GA, Settels JJ, van der Hoeven GM, Karemaker JM, Wieling W. Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre. Cardiovasc Res1988; 22(6): 390-397
|
41 |
Lemson J, Hofhuizen CM, Schraa O, Settels JJ, Scheffer GJ, van der Hoeven JG. The reliability of continuous noninvasive finger blood pressure measurement in critically ill children. Anesth Analg2009; 108(3): 814-821
|
42 |
Maggi R, Viscardi V, Furukawa T, Brignole M. Non-invasive continuous blood pressure monitoring of tachycardic episodes during interventional electrophysiology. Europace2010; 12(11): 1616-1622
|
43 |
Hofhuizen CM, Lemson J, Hemelaar AE, Settels JJ, Schraa O, Singh SK, van der Hoeven JG, Scheffer GJ. Continuous non-invasive finger arterial pressure monitoring reflects intra-arterial pressure changes in children undergoing cardiac surgery. Br J Anaesth2010; 105(4): 493-500
|
44 |
Jeleazcov C, Krajinovic L, Münster T, Birkholz T, Fried R, Schüttler J, Fechner J. Precision and accuracy of a new device (CNAPTM) for continuous non-invasive arterial pressure monitoring: assessment during general anaesthesia. Br J Anaesth2010; 105(3): 264-272
|
45 |
Eckert S, Horstkotte D. Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exercise. Blood Press Monit2002; 7(3): 179-183
|
46 |
Schramm C, Baat L, Plaschke K. Continuous noninvasive arterial pressure: assessment in older and high-risk patients under analgesic sedation. Blood Press Monit2011; 16(6): 270-276
|
47 |
Hahn R, Rinösl H, Neuner M, Kettner SC. Clinical validation of a continuous non-invasive haemodynamic monitor (CNAP™; 500) during general anaesthesia. Br J Anaesth2012108(4):581-585
|
48 |
Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R. Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth2012; 108(2): 202-210
|
49 |
Garnier RP, van der Spoel AG, Sibarani-Ponsen R, Markhorst DG, Boer C. Level of agreement between Nexfin non-invasive arterial pressure with invasive arterial pressure measurements in children. Br J Anaesth2012; 109(4): 609-615
|
50 |
Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, Immink RV, Jöbsis DA, Hollmann MW, Lahpor JR, de Mol BA, van Lieshout JJ. Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology2012; 116(5): 1092-1103
|
51 |
Akkermans J, Diepeveen M, Ganzevoort W, van Montfrans GA, Westerhof BE, Wolf H. Continuous non-invasive blood pressure monitoring, a validation study of Nexfin in a pregnant population. Hypertens Pregnancy2009; 28(2): 230-242
|
52 |
Eeftinck Schattenkerk DW, van Lieshout JJ, van den Meiracker AH, Wesseling KR, Blanc S, Wieling W, van Montfrans GA, Settels JJ, Wesseling KH, Westerhof BE. Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff. Am J Hypertens2009; 22(4): 378-383
|
53 |
Tanaka H, Thulesius O, Yamaguchi H, Mino M, Konishi K. Continuous non-invasive finger blood pressure monitoring in children. Acta Paediatr1994; 83(6): 646-652
|
54 |
Sipkens LM, Treskes K, Ariese-Beldman K, Veerman DP, Boer C. Application of Nexfin noninvasive beat-to-beat arterial blood pressure monitoring in autonomic function testing. Blood Press Monit2011; 16(5): 246-251
|
55 |
Martina JR, Westerhof BE, Van Goudoever J, De Jonge N, Van Lieshout JJ, Lahpor JR, De Mol BA. Noninvasive blood pressure measurement by the Nexfin monitor during reduced arterial pulsatility: a feasibility study. ASAIO J2010; 56(3): 221-227
|
56 |
Siebig S, Rockmann F, Sabel K, Zuber-Jerger I, Dierkes C, Brünnler T, Wrede CE. Continuous non-invasive arterial pressure technique improves patient monitoring during interventional endoscopy. Int J Med Sci2009; 6(1): 37-42
|
57 |
Chen G, Chung E, Meng L, Alexander B, Vu T, Rinehart J, Cannesson M. Impact of non invasive and beat-to-beat arterial pressure monitoring on intraoperative hemodynamic management. J Clin Monit Comput2012; 26(2): 133-140
|
58 |
Lu Z, Mukkamala R. Continuous cardiac output monitoring in humans by invasive and noninvasive peripheral blood pressure waveform analysis. J Appl Physiol2006; 101(2): 598-608
|
59 |
Rang S, de Pablo Lapiedra B, van Montfrans GA, Bouma BJ, Wesseling KH, Wolf H. Modelflow: a new method for noninvasive assessment of cardiac output in pregnant women. Am J Obstet Gynecol2007; 196(3): 235 e1-8
|
60 |
Bogert LW, Wesseling KH, Schraa O, Van Lieshout EJ, de Mol BA, van Goudoever J, Westerhof BE, van Lieshout JJ. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia2010; 65(11): 1119-1125
|
61 |
Broch O, Renner J, Gruenewald M, Meybohm P, Schöttler J, Caliebe A, Steinfath M, Malbrain M, Bein B. A comparison of the Nexfin® and transcardiopulmonary thermodilution to estimate cardiac output during coronary artery surgery. Anaesthesia2012; 67(4): 377-383
|
62 |
van der Spoel AG, Voogel AJ, Folkers A, Boer C, Bouwman RA. Comparison of noninvasive continuous arterial waveform analysis (Nexfin) with transthoracic Doppler echocardiography for monitoring of cardiac output. J Clin Anesth2012; 24(4): 304-309
|
63 |
Chen G, Meng L, Alexander B, Tran NP, Kain ZN, Cannesson M. Comparison of noninvasive cardiac output measurements using the Nexfin monitoring device and the esophageal Doppler. J Clin Anesth2012; 24(4): 275-283
|
64 |
Fischer MO, Avram R, Cârjaliu I, Massetti M, Gérard JL, Hanouz JL, Fellahi JL. Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgery. Br J Anaesth2012; 109(4): 514-521
|
65 |
McCarthy T, Telec N, Dennis A, Griffiths J, Buettner A. Ability of non-invasive intermittent blood pressure monitoring and a continuous non-invasive arterial pressure monitor (CNAP™) to provide new readings in each 1-min interval during elective caesarean section under spinal anaesthesia. Anaesthesia2012; 67(3): 274-279
|
66 |
Kemmotsu O, Ueda M, Otsuka H, Yamamura T, Okamura A, Ishikawa T, Winter DC, Eckerle JS. Blood pressure measurement by arterial tonometry in controlled hypotension. Anesth Analg1991; 73(1): 54-58
|
67 |
Sato T, Nishinaga M, Kawamoto A, Ozawa T, Takatsuji H. Accuracy of a continuous blood pressure monitor based on arterial tonometry. Hypertension1993; 21(6 Pt 1): 866-874
|
68 |
Weiss BM, Spahn DR, Rahmig H, Rohling R, Pasch T. Radial artery tonometry: moderately accurate but unpredictable technique of continuous non-invasive arterial pressure measurement. Br J Anaesth1996; 76(3): 405-411
|
69 |
Kemmotsu O, Ohno M, Takita K, Sugimoto H, Otsuka H, Morimoto Y, Mayumi T. Noninvasive, continuous blood pressure measurement by arterial tonometry during anesthesia in children. Anesthesiology1994; 81(5): 1162-1168
|
70 |
Harasawa K, Yamane M, Morimoto Y, Kemmotsu O. Performance of a tonometer for arterial pressure measurement during anesthesia in the elderly. J Anesth2004; 18(4): 316-317
|
71 |
Papaioannou TG, Stamatelopoulos KS, Gialafos E, Vlachopoulos C, Karatzis E, Nanas J, Lekakis J. Monitoring of arterial stiffness indices by applanation tonometry and pulse wave analysis: reproducibility at low blood pressures. J Clin Monit Comput2004; 18(2): 137-144
|
72 |
O’Rourke MF, Seward JB. Central arterial pressure and arterial pressure pulse: new views entering the second century after Korotkov. Mayo Clin Proc2006; 81(8): 1057-1068
|
73 |
O’Rourke MF, Adji A. An updated clinical primer on large artery mechanics: implications of pulse waveform analysis and arterial tonometry. Curr Opin Cardiol2005; 20(4): 275-281
|
74 |
Saugel B, Fassio F, Hapfelmeier A, Meidert AS, Schmid RM, Huber W. The T-Line TL-200 system for continuous non-invasive blood pressure measurement in medical intensive care unit patients. Intensive Care Med2012; 38(9): 1471-1477
|
75 |
Szmuk P, Pivalizza E, Warters RD, Ezri T, Gebhard R. An evaluation of the T-Line Tensymeter continuous noninvasive blood pressure device during induced hypotension. Anaesthesia2008; 63(3): 307-312
|
76 |
Poon CC, Zhang YT. Cuff-less and noninvasive measurements of arterial blood pressure by pulse transit time. Conf Proc IEEE Eng Med Biol Soc2005; 6: 5877-5880
|
77 |
Ahlstrom C, Johansson A, Uhlin F, Länne T, Ask P. Noninvasive investigation of blood pressure changes using the pulse wave transit time: a novel approach in the monitoring of hemodialysis patients. J Artif Organs2005; 8(3): 192-197
|
78 |
Fung P, Dumont G, Ries C, Mott C, Ansermino M. Continuous noninvasive blood pressure measurement by pulse transit time. Conf Proc IEEE Eng Med Biol Soc2004; 1: 738-741
|
79 |
Lansdorp B, Ouweneel D, de Keijzer A, van der Hoeven JG, Lemson J, Pickkers P. Non-invasive measurement of pulse pressure variation and systolic pressure variation using a finger cuff corresponds with intra-arterial measurement. Br J Anaesth2011; 107(4): 540-545
|
80 |
Monnet X, Dres M, Ferré A, Le Teuff G, Jozwiak M, Bleibtreu A, Le Deley MC, Chemla D, Richard C, Teboul JL. Prediction of fluid responsiveness by a continuous non-invasive assessment of arterial pressure in critically ill patients: comparison with four other dynamic indices. Br J Anaesth2012; 109(3): 330-338
|
/
〈 | 〉 |