Effect of surgical factors on the augmentation of cement- injectable cannulated pedicle screw fixation by a novel calcium phosphate-based nanocomposite

Haolin Sun, Chun Liu, Shunlun Chen, Yanjie Bai, Huilin Yang, Chunde Li, Lei Yang

PDF(3666 KB)
PDF(3666 KB)
Front. Med. ›› 2019, Vol. 13 ›› Issue (5) : 590-601. DOI: 10.1007/s11684-019-0710-z
RESEARCH ARTICLE
RESEARCH ARTICLE

Effect of surgical factors on the augmentation of cement- injectable cannulated pedicle screw fixation by a novel calcium phosphate-based nanocomposite

Author information +
History +

Abstract

Bone cement-augmented pedicle screw system demonstrates great efficacy in spinal disease treatments. However, the intrinsic drawbacks associated with clinically used polymethylmethacrylate (PMMA) cement demands for new bone cement formulations. On the basis of our previous studies, a novel injectable and biodegradable calcium phosphate-based nanocomposite (CPN) for the augmentation of pedicle screw fixation was systematically evaluated for its surgical feasibility and biomechanical performance by simulated and animal osteoporotic bone models, and the results were compared with those of clinical PMMA cement. ASTM-standard solid foam and open-cell foam models and decalcified sheep vertebra models were employed to evaluate the augmentation effects of CPN on bone tissue and on the cement-injected cannulated pedicle screws (CICPs) placed in osteoporotic bone. Surgical factors in CICPs application, such as injection force, tapping technique, screw diameter, and pedicle screw loosening scenarios, were studied in comparison with those in PMMA. When directly injected to the solid foam model, CPN revealed an identical augmentation effect to that of PMMA, as shown by the similar compressive strengths (0.73±0.04 MPa for CPN group vs. 0.79±0.02 MPa for PMMA group). The average injection force of CPN at approximately 40–50 N was higher than that of PMMA at approximately 20 N. Although both values are acceptable to surgeons, CPN revealed a more consistent injection force pattern than did PMMA. The dispersing and anti-pullout ability of CPN were not affected by the surgical factors of tapping technique and screw diameter. The axial pullout strength of CPN evaluated by the decalcified sheep vertebra model revealed a similar augmentation level as that of PMMA (1351.6±324.2 N for CPN vs. 1459.7±304.4 N for PMMA). The promising results of CPN clearly suggest its potential for replacing PMMA in CICPs augmentation application and the benefits of further study and development for clinical uses.

Keywords

bone cement / pedicle screw / degenerative spinal diseases / calcium phosphate / injectable

Cite this article

Download citation ▾
Haolin Sun, Chun Liu, Shunlun Chen, Yanjie Bai, Huilin Yang, Chunde Li, Lei Yang. Effect of surgical factors on the augmentation of cement- injectable cannulated pedicle screw fixation by a novel calcium phosphate-based nanocomposite. Front. Med., 2019, 13(5): 590‒601 https://doi.org/10.1007/s11684-019-0710-z

References

[1]
Hitchon PW, Brenton MD, Coppes JK, From AM, Torner JC. Factors affecting the pullout strength of self-drilling and self-tapping anterior cervical screws. Spine 2003; 28(1): 9–13
CrossRef Pubmed Google scholar
[2]
Lai DM, Shih YT, Chen YH, Chien A, Wang JL. Effect of pedicle screw diameter on screw fixation efficacy in human osteoporotic thoracic vertebrae. J Biomech 2018; 70: 196–203
CrossRef Pubmed Google scholar
[3]
Weiser L, Huber G, Sellenschloh K, Viezens L, Püschel K, Morlock MM, Lehmann W. Time to augment?! Impact of cement augmentation on pedicle screw fixation strength depending on bone mineral density. Eur Spine J 2018; 27(8): 1964–1971
CrossRef Pubmed Google scholar
[4]
Pishnamaz M, Lange H, Herren C, Na HS, Lichte P, Hildebrand F, Pape HC, Kobbe P. The quantity of bone cement influences the anchorage of augmented pedicle screws in the osteoporotic spine: a biomechanical human cadaveric study. Clin Biomech (Bristol, Avon) 2018; 52: 14–19
CrossRef Pubmed Google scholar
[5]
Paré PE, Chappuis JL, Rampersaud R, Agarwala AO, Perra JH, Erkan S, Wu C. Biomechanical evaluation of a novel fenestrated pedicle screw augmented with bone cement in osteoporotic spines. Spine 2011; 36(18): E1210–E1214
CrossRef Pubmed Google scholar
[6]
Choma TJ, Frevert WF, Carson WL, Waters NP, Pfeiffer FM. Biomechanical analysis of pedicle screws in osteoporotic bone with bioactive cement augmentation using simulated in vivo multicomponent loading. Spine 2011; 36(6): 454–462
CrossRef Pubmed Google scholar
[7]
Janssen I, Ryang YM, Gempt J, Bette S, Gerhardt J, Kirschke JS, Meyer B. Risk of cement leakage and pulmonary embolism by bone cement-augmented pedicle screw fixation of the thoracolumbar spine. Spine J 2017; 17(6): 837–844
CrossRef Pubmed Google scholar
[8]
Padányi Csaba, Misik F, Papp Z, Vitanovics D, Balogh A, Veres R, Lipóth L, Banczerowski P. Treatment of osteoporotic vertebral compression fractures with PMMA-augmented pedicle screw fixation. Ideggyogy Sz 2015; 68(1-2): 52–58 (in Hungarian)
Pubmed
[9]
Moon BJ, Cho BY, Choi EY, Zhang HY. Polymethylmethacrylate-augmented screw fixation for stabilization of the osteoporotic spine: a three-year follow-up of 37 patients. J Korean Neurosurg Soc 2009; 46(4): 305–311
CrossRef Pubmed Google scholar
[10]
Liu D, Sheng J, Wu HH, Kang X, Xie QY, Luo Y, Zhou JJ, Zheng W. Biomechanical study of injectable hollow pedicle screws for PMMA augmentation in severely osteoporotic lumbar vertebrae: effect of PMMA distribution and volume on screw stability. J Neurosurg Spine 2018; 29(6): 639–646
CrossRef Pubmed Google scholar
[11]
Galbusera F, Volkheimer D, Reitmaier S, Berger-Roscher N, Kienle A, Wilke HJ. Pedicle screw loosening: a clinically relevant complication? Eur Spine J 2015; 24(5): 1005–1016
CrossRef Pubmed Google scholar
[12]
María SF, Claudia T, Victor Hugo P, Marcela V, Luis V, María Luisa F, Luis C, Valenzuela GJ. Perinatal neuroendocrine regulation. Development of the circadian time-keeping system. J Biomech 2004; 186(2): 169–173
[13]
Fukuda C, Goto K, Imamura M, Neo M, Nakamura T. Bone bonding ability and handling properties of a titania-polymethylmethacrylate (PMMA) composite bioactive bone cement modified with a unique PMMA powder. Acta Biomater 2011; 7(10): 3595–3600
CrossRef Pubmed Google scholar
[14]
Piñera AR, Duran C, Lopez B, Saez I, Correia E, Alvarez L. Instrumented lumbar arthrodesis in elderly patients: prospective study using cannulated cemented pedicle screw instrumentation. Eur Spine J 2011; 20 (Suppl 3):408–414 PMID: 21850421 DOI: 10.1007/s00586-011-1907-2
[15]
Breusch SJ, Kühn KD. Bone cements based on polymethylmethacrylate. Orthopade 2003; 32(1): 41–50 (in German)
CrossRef Pubmed Google scholar
[16]
Mar DE, Clary SJ, Burton DC, McIff TE. Effect of spinous process tether tension for prophylactic treatment of proximal junctional kyphosis in adult spinal deformity surgery. Spine J 2017; 17(10): S190
CrossRef Google scholar
[17]
Sun H, Liu C, Liu H, Bai Y, Zhang Z, Li X, Li C, Yang H, Yang L. A novel injectable calcium phosphate-based nanocomposite for the augmentation of cannulated pedicle-screw fixation. Int J Nanomedicine 2017; 12: 3395–3406
CrossRef Pubmed Google scholar
[18]
Chun L, Huiling L, Haolin S, Huilin Y, Chunde L, Lei Y. Development and optimization of biodegradable calcium phosphate-based nanocomposite for the application of spinal fixation. Ferroelectrics 2018; 527(1): 162–169
CrossRef Google scholar
[19]
Liu H, Liu B, Gao C, Meng B, Yang H, Yu H, Yang L. Injectable, biomechanically robust, biodegradable and osseointegrative bone cement for percutaneous kyphoplasty and vertebroplasty. Int Orthop 2018; 42(1): 125–132
CrossRef Pubmed Google scholar
[20]
Fan HT, Zhang RJ, Shen CL, Dong FL, Li Y, Song PW, Gong C, Wang YJ. The biomechanical properties of pedicle screw fixation combined with trajectory bone cement augmentation in osteoporotic vertebrae. Clin Spine Surg 2016; 29(2): 78–85
CrossRef Pubmed Google scholar
[21]
Chen SI, Lin RM, Chang CH. Biomechanical investigation of pedicle screw-vertebrae complex: a finite element approach using bonded and contact interface conditions. Med Eng Phys 2003; 25(4): 275–282
CrossRef Pubmed Google scholar
[22]
Gao C, Liu H, Yang H, Yang L. Fabrication and characterization of injectable calcium phosphate-based cements for kyphoplasty. Mater Technol 2015; 30(sup8): B256–B263
CrossRef Google scholar
[23]
Chen LH, Tai CL, Lee DM, Lai PL, Lee YC, Niu CC, Chen WJ. Pullout strength of pedicle screws with cement augmentation in severe osteoporosis: a comparative study between cannulated screws with cement injection and solid screws with cement pre-filling. BMC Musculoskelet Disord 2011; 12(1): 33
CrossRef Pubmed Google scholar
[24]
Johnson AE, Keller TS. Mechanical properties of open-cell foam synthetic thoracic vertebrae. J Mater Sci Mater Med 2008; 19(3): 1317–1323
CrossRef Pubmed Google scholar
[25]
Keller TS, Nathan M. Height change caused by creep in intervertebral discs: a sagittal plane model. J Spinal Disord 1999; 12(4): 313–324
CrossRef Pubmed Google scholar
[26]
Muthukrishnan S, Mitra S, Ranjith P, Gopalakrishnan S, Parthipan P, Al-Farhood B. High heat resistant blends of poly (methyl methacrylate) and styrenic copolymers via post reactor modification. J Appl Polym Sci 2018; 135(18): 46220
CrossRef Google scholar
[27]
Hsu CC, Chao CK, Wang JL, Hou SM, Tsai YT, Lin J. Increase of pullout strength of spinal pedicle screws with conical core: biomechanical tests and finite element analyses. J Orthop Res 2005; 23(4): 788–794
CrossRef Pubmed Google scholar
[28]
Pfeiffer FM, Abernathie DL, Smith DE. A comparison of pullout strength for pedicle screws of different designs: a study using tapped and untapped pilot holes. Spine 2006; 31(23): E867–E870
CrossRef Pubmed Google scholar
[29]
Carmouche JJ, Molinari RW, Gerlinger T, Devine J, Patience T. Effects of pilot hole preparation technique on pedicle screw fixation in different regions of the osteoporotic thoracic and lumbar spine. J Neurosurg Spine 2005; 3(5): 364–370
CrossRef Pubmed Google scholar
[30]
Fujibayashi S, Takemoto M, Neo M, Matsuda S. Strategy for salvage pedicle screw placement: a technical note. Int J Spine Surg 2013; 7(1): e67–e71
CrossRef Pubmed Google scholar

Acknowledgements

This study was funded by the Youth Clinical Research Project of Peking University First Hospital (No. 2017CR06), the National Natural Science Foundation of China (Nos. 81622032 and 51672184), the Principal Project of Natural Science Research of Jiangsu Higher Education Institutions (No. 17KJA180011), and Jiangsu Innovation and Entrepreneurship Program.

Compliance with ethics guidelines

Haolin Sun, Chun Liu, Shunlun Chen, Yanjie Bai, Huilin Yang, Chunde Li, and Lei Yang declare that they have no conflicts of interest in this study. All institutional and national guidelines for the care and use of laboratory animal were followed. The animal study protocol was approved by the Ethics Committee of Soochow University under case number ECSU-201700031.

RIGHTS & PERMISSIONS

2019 Higher Education Press and Springer-Verlag GmbH Germany, part of Springer Nature
AI Summary AI Mindmap
PDF(3666 KB)

Accesses

Citations

Detail

Sections
Recommended

/