A potentiometric cobalt-based phosphate sensor based on screen-printing technology
Lei ZHU, Xiaohong ZHOU, Hanchang SHI
A potentiometric cobalt-based phosphate sensor based on screen-printing technology
A potentiometric cobalt-based screen-printing sensor was fabricated by electroplating cobalt on the surface of a screen-printing electrode as the sensitive layer for the determination of dihydrogenphosphate () in wastewater samples. The electrochemical performance of this sensor was fully examined to determine its detection calibration, detection limit, response time, selectivity, and interference with pH, various ions, and dissolved oxygen (DO). The cobalt-based phosphate sensor showed a phosphate-selective potential response in the range of 10-5 mol·L-1 to 10-1 mol·L-1, yielding a detection limit of 3.16 × 10-6 mol∙L-1and a slope of -37.51 mV∙decade-1 in an acidic solution (pH 4.0) of . DO and pH were found to interfere with sensor responses to phosphate. Ultimately, the performance of the sensor was validated for detecting wastewater samples from the Xiaojiahe Wastewater Treatment Plant against the standard spectrophotometric methods for analysis. The discrepancy between the two methods was generally ±5% (relative standard deviation). Aside from its high selectivity, sensitivity, and stability, which are comparable with conventional bulk Co-wire sensors, the proposed phosphate sensor presents many other advantages, such as low price, compactness, ease of use, and the possibility of integration with other analytical devices, such as flow injectors.
phosphate / cobalt / screen-printing technology / electroplate / wastewater
[1] |
Engblom S O. The phosphate sensor. Biosensors and Bioelectronics, 1998, 13(9): 981–994
CrossRef
Pubmed
Google scholar
|
[2] |
Storer R A. In: Annual Book of ASTM standard, American Society for Testing and Materials: West Conshohocken, 1996, 1: 24–27
|
[3] |
Carey C M. Association constant for potassium-acid-phosphate determined by K and HPO4 ion-selective electrodes. Journal of Dental Research, 1997, 76: 668–668
|
[4] |
Liu D, Chen W C, Yang R H, Shen G L, Yu R Q. Yang R H, Shen G L, Yu R Q. Polymeric membrane phosphate sensitive electrode based on binuclear organotin compound. Analytica Chimica Acta, 1997, 338(3): 209–214
CrossRef
Google scholar
|
[5] |
de Marco R, Phan C. Determination of phosphate in hydroponic nutrient solutions using flow injection potentiometry and a cobalt-wire phosphate ion-selective electrode. Talanta, 2003, 60(6): 1215–1221
CrossRef
Pubmed
Google scholar
|
[6] |
Engblom S O. Determination of inorganic phosphate in a soil extract using a cobalt electrode. Plant and Soil, 1999, 206(2): 173–179
CrossRef
Google scholar
|
[7] |
Xiao D, Yuan H Y, Li J, Yu R Q. Surface-modified coalt-based sensor as a phosphate-sensitive electorde. Analytical Chemistry, 1995, 67(2): 288–291
CrossRef
Google scholar
|
[8] |
Chen Z, MarcoD R, Alexander P W. Flow-injection potentiometric detection of phosphates using a metallic cobalt wire ion-selective electrode. Analytical Communications, 1997, 34(3): 93–95
CrossRef
Google scholar
|
[9] |
Zou Z W, Han J Y, Jang A, Bishop P L, Ahn C H. A disposable on-chip phosphate sensor with planar cobalt microelectrodes on polymer substrate. Biosensors and Bioelectronics, 2007, 22(9–10): 1902–1907
CrossRef
Pubmed
Google scholar
|
[10] |
Fanjul-Bolado P, Queipo P, Lamas-Ardisana P J, Costa-García A. Manufacture and evaluation of carbon nanotube modified screen-printed electrodes as electrochemical tools. Talanta, 2007, 74(3): 427–433
CrossRef
Pubmed
Google scholar
|
[11] |
Avramescu A, Andreescu S, Noguer T, Bala C, Andreescu D, Marty J L. Biosensors designed for environmental and food quality control based on screen-printed graphite electrodes with different configurations. Analytical and Bioanalytical Chemistry, 2002, 374(1): 25–32
CrossRef
Pubmed
Google scholar
|
[12] |
Tymecki L, Glab S, Koncki R. Miniaturized, Planar ion-selective electrodes fabricated by means of thick-film technology. Sensors (Basel, Switzerland), 2006, 6(4): 390–396
CrossRef
Google scholar
|
[13] |
Albareda-Sirvent M, Merkoci A, Alegret S. Configurations used in the design of screen-printed enzymatic biosensors. A review. Sensors and Actuators. B, Chemical, 2000, 69(1–2): 153–163
CrossRef
Google scholar
|
[14] |
Shih Y, Zen J M, Kumar A S, Chen P Y. Flow injection analysis of zinc pyrithione in hair care products on a cobalt phthalocyanine modified screen-printed carbon electrode. Talanta, 2004, 62(5): 912–917
CrossRef
Pubmed
Google scholar
|
[15] |
Martinez N A, Messina G A, Bertolino F A, Salinas E, Raba J. Screen-printed enzymatic biosensor modified with carbon nanotube for the methimazole determination in pharmaceuticals formulations. Sensors and Actuators. B: Chemical, 2008, 133(1): 256–262
CrossRef
Google scholar
|
[16] |
Meruva R K, Meyerhoff M E. Mixed potential response mechanism of cobalt electrodes toward inorganic phosphate. Analytical Chemistry, 1996, 68(13): 2022–2026
CrossRef
Pubmed
Google scholar
|
[17] |
Tymecki L, Zwierkowska Z, Koncki R. Screen-printed reference electrodes for potentiometric measurements. Analytica Chimica Acta, 2004, 526(1): 3–11
CrossRef
Google scholar
|
[18] |
Littmann E R, Klotz J R M. Naphthenic acids II Manufacture, properties, and uses. Chemical Reviews, 1942, 30(1): 97–111
CrossRef
Google scholar
|
[19] |
Marco D R, Pejcic B, Chen Z. Flow injection potentiometric determination of phosphate in waste waters and fertilisers using a cobalt wire ion-selective electrode. Analyst (London), 1998, 123(7): 1635–1640
CrossRef
Google scholar
|
[20] |
Fry C H, Langley S E M. Ion-selective Electrodes for Biological Systems. Amsterdam: Harwood Academic, 2001
|
/
〈 | 〉 |