A potentiometric cobalt-based phosphate sensor based on screen-printing technology

Lei ZHU , Xiaohong ZHOU , Hanchang SHI

Front. Environ. Sci. Eng. ›› 2014, Vol. 8 ›› Issue (6) : 945 -951.

PDF (263KB)
Front. Environ. Sci. Eng. ›› 2014, Vol. 8 ›› Issue (6) : 945 -951. DOI: 10.1007/s11783-013-0615-z
RESEARCH ARTICLE
RESEARCH ARTICLE

A potentiometric cobalt-based phosphate sensor based on screen-printing technology

Author information +
History +
PDF (263KB)

Abstract

A potentiometric cobalt-based screen-printing sensor was fabricated by electroplating cobalt on the surface of a screen-printing electrode as the sensitive layer for the determination of dihydrogenphosphate (H2PO4-) in wastewater samples. The electrochemical performance of this sensor was fully examined to determine its detection calibration, detection limit, response time, selectivity, and interference with pH, various ions, and dissolved oxygen (DO). The cobalt-based phosphate sensor showed a phosphate-selective potential response in the range of 10-5 mol·L-1 to 10-1 mol·L-1, yielding a detection limit of 3.16 × 10-6 mol∙L-1and a slope of -37.51 mV∙decade-1 in an acidic solution (pH 4.0) of H2PO4-. DO and pH were found to interfere with sensor responses to phosphate. Ultimately, the performance of the sensor was validated for detecting wastewater samples from the Xiaojiahe Wastewater Treatment Plant against the standard spectrophotometric methods for H2PO4- analysis. The discrepancy between the two methods was generally ±5% (relative standard deviation). Aside from its high selectivity, sensitivity, and stability, which are comparable with conventional bulk Co-wire sensors, the proposed phosphate sensor presents many other advantages, such as low price, compactness, ease of use, and the possibility of integration with other analytical devices, such as flow injectors.

Keywords

phosphate / cobalt / screen-printing technology / electroplate / wastewater

Cite this article

Download citation ▾
Lei ZHU, Xiaohong ZHOU, Hanchang SHI. A potentiometric cobalt-based phosphate sensor based on screen-printing technology. Front. Environ. Sci. Eng., 2014, 8(6): 945-951 DOI:10.1007/s11783-013-0615-z

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Engblom S O. The phosphate sensor. Biosensors and Bioelectronics, 1998, 13(9): 981–994

[2]

Storer R A. In: Annual Book of ASTM standard, American Society for Testing and Materials: West Conshohocken, 1996, 1: 24–27

[3]

Carey C M. Association constant for potassium-acid-phosphate determined by K and HPO4 ion-selective electrodes. Journal of Dental Research, 1997, 76: 668–668

[4]

Liu D, Chen W C, Yang R H, Shen G L, Yu R Q. Yang R H, Shen G L, Yu R Q. Polymeric membrane phosphate sensitive electrode based on binuclear organotin compound. Analytica Chimica Acta, 1997, 338(3): 209–214

[5]

de Marco R, Phan C. Determination of phosphate in hydroponic nutrient solutions using flow injection potentiometry and a cobalt-wire phosphate ion-selective electrode. Talanta, 2003, 60(6): 1215–1221

[6]

Engblom S O. Determination of inorganic phosphate in a soil extract using a cobalt electrode. Plant and Soil, 1999, 206(2): 173–179

[7]

Xiao D, Yuan H Y, Li J, Yu R Q. Surface-modified coalt-based sensor as a phosphate-sensitive electorde. Analytical Chemistry, 1995, 67(2): 288–291

[8]

Chen Z, MarcoD R, Alexander P W. Flow-injection potentiometric detection of phosphates using a metallic cobalt wire ion-selective electrode. Analytical Communications, 1997, 34(3): 93–95

[9]

Zou Z W, Han J Y, Jang A, Bishop P L, Ahn C H. A disposable on-chip phosphate sensor with planar cobalt microelectrodes on polymer substrate. Biosensors and Bioelectronics, 2007, 22(9–10): 1902–1907

[10]

Fanjul-Bolado P, Queipo P, Lamas-Ardisana P J, Costa-García A. Manufacture and evaluation of carbon nanotube modified screen-printed electrodes as electrochemical tools. Talanta, 2007, 74(3): 427–433

[11]

Avramescu A, Andreescu S, Noguer T, Bala C, Andreescu D, Marty J L. Biosensors designed for environmental and food quality control based on screen-printed graphite electrodes with different configurations. Analytical and Bioanalytical Chemistry, 2002, 374(1): 25–32

[12]

Tymecki L, Glab S, Koncki R. Miniaturized, Planar ion-selective electrodes fabricated by means of thick-film technology. Sensors (Basel, Switzerland), 2006, 6(4): 390–396

[13]

Albareda-Sirvent M, Merkoci A, Alegret S. Configurations used in the design of screen-printed enzymatic biosensors. A review. Sensors and Actuators. B, Chemical, 2000, 69(1–2): 153–163

[14]

Shih Y, Zen J M, Kumar A S, Chen P Y. Flow injection analysis of zinc pyrithione in hair care products on a cobalt phthalocyanine modified screen-printed carbon electrode. Talanta, 2004, 62(5): 912–917

[15]

Martinez N A, Messina G A, Bertolino F A, Salinas E, Raba J. Screen-printed enzymatic biosensor modified with carbon nanotube for the methimazole determination in pharmaceuticals formulations. Sensors and Actuators. B: Chemical, 2008, 133(1): 256–262

[16]

Meruva R K, Meyerhoff M E. Mixed potential response mechanism of cobalt electrodes toward inorganic phosphate. Analytical Chemistry, 1996, 68(13): 2022–2026

[17]

Tymecki L, Zwierkowska Z, Koncki R. Screen-printed reference electrodes for potentiometric measurements. Analytica Chimica Acta, 2004, 526(1): 3–11

[18]

Littmann E R, Klotz J R M. Naphthenic acids II Manufacture, properties, and uses. Chemical Reviews, 1942, 30(1): 97–111

[19]

Marco D R, Pejcic B, Chen Z. Flow injection potentiometric determination of phosphate in waste waters and fertilisers using a cobalt wire ion-selective electrode. Analyst (London), 1998, 123(7): 1635–1640

[20]

Fry C H, Langley S E M. Ion-selective Electrodes for Biological Systems. Amsterdam: Harwood Academic, 2001

RIGHTS & PERMISSIONS

Higher Education Press and Springer-Verlag Berlin Heidelberg

AI Summary AI Mindmap
PDF (263KB)

2870

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/