A new system for rapid measurement of ATP

Shu Baihua , Zhou Yikai , Ren Shu

Current Medical Science ›› 1997, Vol. 17 ›› Issue (3) : 190 -192.

PDF
Current Medical Science ›› 1997, Vol. 17 ›› Issue (3) : 190 -192. DOI: 10.1007/BF02888300
Article

A new system for rapid measurement of ATP

Author information +
History +
PDF

Abstract

The paper introduces a new type instrument for rapid measuring ATP. The system consists of a micromodule ATP sensor and an instrument for measuring weak light transmitted by optic fiber. The micromodule ATP sensor mainly is composed of enzyme membrane, a probe and a bundle of optic fiber. The instrument measuring weak light consists of photomultiplier, high voltage power, pulse amplifier and counter. The instrument was characterized by simple structure, small size, rapid response time (<5 s), high sensitivity (10−12 mol/L), stable performance (measuring the same sample for 50 times, CV < 5 %), long enzyme storage time (>3 months).

Keywords

adenosine triphosphate / luminescent analysis / instrument / sensor

Cite this article

Download citation ▾
Shu Baihua, Zhou Yikai, Ren Shu. A new system for rapid measurement of ATP. Current Medical Science, 1997, 17(3): 190-192 DOI:10.1007/BF02888300

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

StanleyP E. A review of bioluminescent ATP techniques in rapid microbiology. J Biolumin Chemilumin, 1989, 4: 375-375

[2]

StanleyP E. Commercially available luminometers and imaging devices for low-light measurements and kits and reagents utilizing bioluminescence or chemiluminescence: Survey Update I. J Biolumin Chemilumin, 1993, 8: 237-237

[3]

StanleyP E. Commercially available luminometers and imaging devices for low light measurements and kits and reagents utilizing bioluminescence or chemiluminescence: Survey Update I. J Biolumin Chemilumin, 1994, 9: 51-51

[4]

StanleyP E. Commercially available luminometers and imaging devices for low -light measurements and kits and reagents utilizing bioluminescence or chemiluminecence: Survey update I. J Biolumin Chemilumin, 1994, 9: 123-123

[5]

UgarovaN N. Luciferase of luciola mingrelica fireflics kinetics and regulation mechanism. J Biolumin Chemilumin, 1989, 4: 406-406

[6]

Mosbach K. Immobilized enzymes and cell in methods in enzymology vol. 137 Academic Press, INC 1988.