A study of calcium ion-selective PVC membrane electrode based on neutral carrier N,n,n’,n’-tetracyclo-3-oxapetanediamide
Ma Yi-long , Rao Xiao-hong , Zhong Shi-ming , Ren Shu , Yu Tian-xiang , Zhen Qiong
Current Medical Science ›› 1992, Vol. 12 ›› Issue (2) : 98 -102.
A study of calcium ion-selective PVC membrane electrode based on neutral carrier N,n,n’,n’-tetracyclo-3-oxapetanediamide
A calcium ion-selective PVC membrane electrode based on neutral carrier n,n,n’,n’-tetracyclohexyl-3-oxapetanediamide, using di-(2-ethylhexyl) phthalate as the plasticizer and potassium tetrakis (4-chlorophenyl) borate as the additive is reported in this paper. The ion selective membrane consists of 1 wt % of the Ca2+ selective ligand, 65 wt % of the plasticizer, 1 wt % of the additive and 33 wt % of poly (vinyl chloride) powder. The electrode has the linear response range of 2.0 × 10™7 —10™1 mol/L with the Nernstian slope of 28 mV/decade at 25 °C and the detection limit of 2.0 × 10™8 mol/L. The response time of the calcium ion-selective electrode is 4s as the concentration of calcium ion is rapidly shifted from 10™5 to 10™4 mol/L. The potential stability and reproducibility are good. The free calcium in blood serum was determined by the calcium ion-selective electrode with satisfactory results.
calcium ion-selective electrode / neutral carrier / PVC membrane / calcium determination in serum