Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake

Yue-ying Pan , Shu-chang Zhou , Yu-jin Wang , Qian Li , Ting-ting Zhu , Chun-xia Liu , Han-xiong Guan

Current Medical Science ›› 2020, Vol. 40 ›› Issue (1) : 178 -183.

PDF
Current Medical Science ›› 2020, Vol. 40 ›› Issue (1) : 178 -183. DOI: 10.1007/s11596-020-2162-8
Article

Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake

Author information +
History +
PDF

Abstract

The effect of low voltage and low concentration contrast agent on image quality of coronary CT angiography, radiation dose and iodine intake was evaluated. A total of 121 patients with body mass index (BMI) <26 kg/m2 and heart rate (HR) <70 beats/min were randomly divided into four groups: group A (n=31, 80 kVp, 270 mgI/mL); group B (n=33, 100 kVp, 270 mgI/mL); group C (n=30, 100 kVp, 320 mgI/mL); group D (n=27, 100 kVp, 400 mgI/mL). The automatic current modulation system and the iterative algorithm for reconstruction were adopted in each group. The CT values and SD values of the aortic root (AR), subcutaneous fat, left coronary artery opening (LCA), and right coronary artery opening (RCA) were measured in all groups, the signal-to-noise ratio (SNR) and contrast noise ratio (CNR) were calculated, and effective radiation dose and iodine intake were recorded. The subjective assessment for image quality was performed by two physicians using a 4-point scale. The results were compared using the one-way ANOVA and rank sum tests. The image quality of the four groups met the clinical diagnostic requirements. The CT values of AR in groups A, B, C, and D were 537.6±71.4, 447.2±81.9, 445.2±64.9 and 518.5±94.9 Hu, respectively, with no significant difference between group A and group D, or between group B and group C, while CT values in groups B and C were significantly lower than those in groups A and D (P<0.05). In groups A, B, C, and D, the LCA SNR values were 22.7±9.1, 23.3±9.1, 23.3±7.7 and 26.6±8.9, and the RCA CNR values were 26.9±9.8, 28.5±11.4, 27.7±8.8 and 32.1±10.6, respectively. The AR visual scores in groups A, B, C and D were 3.8±0.2, 3.9±0.3, 3.9±0.3 and 4.0±0.3, respectively. There were no significant differences in SNR, CNR and visual score among the four groups (P>0.05). The radiation doses in groups A, B, C and D were 2.6±1.4, 3.6±1.8, 4.9±3.5 and 4.9±2.8 mSv, respectively. The radiation dose in group A was significantly less than that in the rest three groups (P<0.05). The iodine intakes in groups A, B, C and D were 14.9±1.5, 15.0±1.5, 17.7±2.0 and 18.1±2.5 g, respectively. There was no significant difference in the intake of iodine between groups C and D, or between groups A and B, while iodine intake in groups A and B were significantly reduced as compared with that in groups C and D (P<0.05). It was concluded that for patients with low BMI and controlled HR, compared to 100 kVp tube voltage combined with multiple concentration contrast agents, 80 kVp combined with 270 mgI/mL contrast agent is enough to ensure the quality of the images, and can reduce the radiation dose significantly, while reducing the amount of iodine intake notably, thus reducing the incidence of adverse reaction.

Keywords

tube voltage / contrast agent / coronary CT angiography / radiation dose / iodine intake

Cite this article

Download citation ▾
Yue-ying Pan, Shu-chang Zhou, Yu-jin Wang, Qian Li, Ting-ting Zhu, Chun-xia Liu, Han-xiong Guan. Application of Low Tube Voltage, Low-concentration Contrast Agent Using a 320-row CT in Coronary CT Angiography: Evaluation of Image Quality, Radiation Dose and Iodine Intake. Current Medical Science, 2020, 40(1): 178-183 DOI:10.1007/s11596-020-2162-8

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

LeberAW, JohnsonT, BeckerA, et al.. Diagnostic accuracy of dual source mult-slice CT-coronary angiography in patents with an intermediate pretest likelihood for coronary artery disease. Eur Heart J, 2007, 28(19): 2354-2360

[2]

BrennerDJ, HallEJ. Computed tomography-an increasing source of radiation exposure. N Engl J Med, 2007, 357(22): 2277-2284

[3]

FromAM, BartholmaiBJ, WilliamsAW, et al.. Mortality associated with nephropathy after radiographic contrast exposure. Mayo Clin Proc, 2008, 83(10): 1095-1100

[4]

AndreiniD, MushtaqS, ConteE, et al.. Coronary CT angiography with 80 kV tube voltage and low iodine concentration contrast agent in patients with low body weight. J Cardiovasc Comput Tomogr, 2016, 10(4): 322-326

[5]

FaggioniL, NeriE, SbragiaP, et al.. 80-kV pulmonary CT angiography with 40 mL of iodinated contrast material in lean patients: comparison of vascular enhancement with iodixanol (320 mg I/mL) and iomeprol (400 mg I/mL). Am J Roentgenol, 2012, 199(6): 1220-1225

[6]

ModicaMJ, KanalKM, GunnML. The obese emergency patient: imaging challenges and solutions. Radiographics, 2011, 31(3): 811-823

[7]

LeipsicJ, LabountyTM, HeilbronB, et al.. Adaptive statistical iteratve reconstruction: assessment of image noise and image quality in coronary CT angiography. Am J Roentgenol, 2010, 195(3): 649-654

[8]

Van der MolenAJ, JoemaiRM, GeleijnsJ. Performance of longitudinal and volumetric tube current modulation in a 64-slice CT with different choices of acquisition and reconstruction parameters. Physica Med, 2012, 28(4): 319-326

[9]

ChenMY, ShanbhagSM, AraiAE. Submillisievert median radiation dose for coronary angiography with a second generation 320-detector row CT scanner in 107 consecutive patients. Radiology, 2013, 267(1): 76-85

[10]

ZhangLJ, QiL, WangJ, et al.. Feasibility of prospectively ECG-triggered high-pitch coronary CT angiography with 30 mL iodinated contrast agent at 70 kVp: initial experience. Eur Radiol, 2014, 24(7): 1537-1546

[11]

Bongartz G, Golding SJ, Jurik AG, et al. European guidelines for multislice computed tomography: appendix c funded by the European Commission contract number FIGM-CT2000-20078-CT-TIP2004. 2015.

[12]

OdaS, UtsunomiyaD, YukiH, et al.. Low contrast and radiation dose coronary CT angiography using a 320-row system and a refined contrast injection and timing method. J Cardiovasc Comput Tomogr, 2015, 9(1): 19-27

[13]

HigashigaitoK, HusarikDB, BarthelmesJ, et al.. Computed tomography angiography of coronary artery bypass grafts: low contrast media volume protocols adapted to tube voltage. Invest Radiol, 2016, 51(4): 241-248

[14]

ZhengM, WuY, WeiM, et al.. Low-concentration contrast medium for 128-slice dual-source CT coronary angiography at a very low radiation dose using prospectively ECG-triggered high-pitch spiral acquisition. Acad Radiol, 2015, 22(2): 195-202

[15]

WuQ, WangY, KaiH, et al.. Application of 80-kVp tube voltage, low-concentration contrast agent and iterative reconstruction in coronary CT angiography: evaluation of image quality and radiation dose. Int J Clin Pract, 2016, 70(Suppl 9B): B50-55

[16]

ZhangC, YuY, ZhangZ, et al.. Imaging quality evaluation of low tube voltage coronary CT angiography using low concentration contrast medium. PloS One, 2015, 10(3): e0120539

[17]

ZhangLJ, QiL, De CeccoCN, et al.. High-pitch coronary CT angiography at 70 kVp with low contrast medium volume: comparison of 80 and 100 kVp high–pitch protocols. Medicine (Baltimore), 2014, 93(22): e92

AI Summary AI Mindmap
PDF

110

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/