Five fresh tibias removed from 30–40 years old dead males were osteotomized at one-third supra-medium segment and then fixed by an unilateral adjustable external fixator (UAEF) to be used as a model of external fixation of tibial fracture. For each loading test under different configurations of UAEF, the ratio of the force-displacement was used to determine its stiffness value. The weakest mode of the UAEF device is in AP bending. Under each mode of compression, distraction, enteroposterial and lateral bending and torsion loading, increasing the pin number or pin diameter or pin separation, and/or reduction of pin group separation can significantly increase the stiffness values of UAEF (P<0. 01). These data revealed that UAEF is suitable for the external fixation of tibial fracture.

Keywords

tibia / external fixator / biomechanics

Cite this article

Download citation ▾
Liu Guo-ping, Du Jing-yuan, Wang Hong, Chen Ru-qing, Xu Zhao-xiong. Biomechanical Study on osteotomized tibias fixed with unilateral adjustable external fixator. Current Medical Science, 1995, 15(4): 242-245 DOI:10.1007/BF02887955

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

GoodshipA E, KenwrightJ. The influence of induced micromovement upon the healing of experimental tibial fractures. J Bone Joint Surg, 1985, 67-B: 650-650

[2]

ChaoY S, BriggsB T, McCoyM T. BrookerA F, EdwardsC C. Theoretical and experimental analyses of Hoffmann-Vidal external fixation system. External Fixation—The Current State of the Art, 1979Baltimore/LondonWilliams &. Wilkins345-370

[3]

JorgensenT E. BrookerA F, EdwardsC C. The measurement of stability in healing tibial shaft fractures treated by external fixation. External Fixation—The Current State of the Art, 1979Baltimore/LondonWilliams & Wilkins383-390

[4]

VidalJ. External fixation. Clin Orthop, 1983, 180: 7-7

[5]

1992, 8: 228

[6]

WeberB G, MagerlF. The external fixator, 1989Berlin/Heidelberg/New York/TokyoHuazhong University of Science and Technology27-50

AI Summary AI Mindmap
PDF

85

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/