In Vitro validation of tissue doppler left ventricular regional wall velocities by using a novel balloon phantom

Wu Ying , Timothy Irvine , Yoshiki Mori , Li Xiaokui , David J. Sahn

Current Medical Science ›› 2001, Vol. 21 ›› Issue (23) : 337 -340.

PDF
Current Medical Science ›› 2001, Vol. 21 ›› Issue (23) : 337 -340. DOI: 10.1007/BF02886574
Article

In Vitro validation of tissue doppler left ventricular regional wall velocities by using a novel balloon phantom

Author information +
History +
PDF

Abstract

To investigate the validity and accuracy of tissue Doppler imaging (TDI) using a novel balloon phantom, validation of TDI myocardial velocity measurements has been carried out indirectly from conventional M-mode images. However it is not a true and independent gold standard. We described a new TDI validation method by using a specially developed left ventricular balloon model mounted in a water bath and constructed using two pear-shaped balloons. It was connected to a pulsatile flow pump at 8 stroke volumes (50–85 ml/beat). The displacement and velocity of the balloon walls were recorded simultaneously by video imaging and TDI on a GE-Vingmed System Five with a 5 MHz phased array probe at the highest frame rates available. Conventional M-mode and 2-D imaging verified that our balloon model mimicked the shape and wall motion of left ventricle. There was a good correlation and agreement between the maximum video excursion of the anterior and posterior walls of the phantom and the results of the temporal integration of digital distance data by TDI (Anterior wall: r=0. 97, SEE=0. 24 mm, s=0. 04±0. 24 mm; Posterior wall: r=0. 95, SEE = 0. 22 mm, −x±s−0. 03±0. 24 mm). Analysis of the velocity profile by the TDI method showed that the velocity at each measured point was correlated well with the velocity obtained from the video images (Anterior wall: r=0. 97, SEE = 0. 30 mm, −x±s= 0. 04±0. 28 mm; Posterior wall: r=0. 97, SEE = 0. 30 mm, −x±s = 0. 04 + 0. 28 mm). Our balloon model provided a new independent method for the validation of TDI data. This study demonstrated that the present TDI system is reliable for measuring wall motion distance and velocity.

Keywords

tissue Doppler imaging / in vitro / left ventricular regional velocity

Cite this article

Download citation ▾
Wu Ying, Timothy Irvine, Yoshiki Mori, Li Xiaokui, David J. Sahn. In Vitro validation of tissue doppler left ventricular regional wall velocities by using a novel balloon phantom. Current Medical Science, 2001, 21(23): 337-340 DOI:10.1007/BF02886574

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

MiyatakeK, YamagishiM, TanakaN, et al. . New method for evaluating left ventricular wall motion by color-coded tissue Doppler imaging: In vitro and in vivo studies. JACC, 1995, 25(3): 717-717

[2]

McDickenW N, SutherlandG R, MoranC M, et al. . Color Doppler velocity imaging of the myocardium. Ultrasound Med Biol, 1992, 18654-654

[3]

BlandJ M, AltmanD G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet, 1986, 1(8476): 307-307

[4]

UematsuM, MiyatakeK, TanakaN, et al. . Myocardial velocity gradient as a new indicator of regional left ventricular contraction: Detection by a two-dimensional tissue Doppler imaging technique. JACC, 1995, 26(1): 217-217

[5]

OhteN, NaritaH, HashimotoT, et al. . Evaluation of left ventricular early diastolic performance by color tissue Doppler imaging of the mitral annulus. Am J Cardiol, 1998, 82(11): 1414-1414

[6]

GorcsanJ, DeswalA, MankadS, et al. . Quantification of the myocardial response to low-dose dobutamine using tissue Doppler echocardiographic measures of velocity and velocity gradient. Am J Cardiol, 1998, 81(5): 615-615

[7]

NaguehS F, MikatiI, KopelenH A, et al. . Doppler estimation of left ventricular filling pressure in sinus tachycardia. A new application of tissue Doppler imaging. Circulation, 1998, 98(16): 1644-1644

[8]

GorcsanJ, StrumD P, MandarinoW A, et al. . Quantitative assessment of alterations in regional left ventricular contractility with color-coded tissue Doppler echocardiography. Comparison with sonomicrometry and pressurevolume relations. Circulation, 1997, 95(10): 2123-2123

[9]

AbeM, OkiT, TabataT, et al. . Difference in the diastolic left ventricular wall motion velocities between aortic and mitral regurgitation by pulsed tissue Doppler imaging. J Am Soc Echocardiogr, 1999, 12(1): 15-15

[10]

OkiT, IuchiA, TabataT, et al. . Left ventricular systolic wall motion velocities along the long and short axes measured by pulsed tissue Doppler imaging in patients with atrial fibrillation. J Am Soc Echocardiogr, 1999, 12(2): 121-121

[11]

GaliutoL, IgnoneG, DemariaA N. Contraction and relaxation velocities of the normal left ventricle using pulsed-wave tissue Doppler echocardiography. Am J Cardiol, 1998, 81(5): 609-609

[12]

BartelT, MullerS, SchurgerD, et al. . Quantitative tissue Doppler echocardiography in comparison with Mmode measurements in health probands. Zeitschrift fur Kardiologie, 1997, 86(2): 131-131