Reliability and Validity of Hand-Held Dynamometer and Hand-Held Sphygmomanometer for Testing Shoulder Isometric External and Internal Rotator Muscles Strength

Francesco Bettariga , Nicola Francesco Lopomo , Fabio Civera , Stefano Giuseppe Lazzarini , Lisa Mantovani , Luca Maestroni

Journal of Science in Sport and Exercise ›› : 1 -8.

PDF
Journal of Science in Sport and Exercise ›› : 1 -8. DOI: 10.1007/s42978-023-00232-1
Original Article

Reliability and Validity of Hand-Held Dynamometer and Hand-Held Sphygmomanometer for Testing Shoulder Isometric External and Internal Rotator Muscles Strength

Author information +
History +
PDF

Abstract

Background

Shoulder strength evaluation is a recommended procedure in musculoskeletal rehabilitation.

Aim

To examine hand-held sphygmomanometer (HHS) and hand-held dynamometer (HHD) intra- and inter-rater reliability during isometric shoulder external and internal rotation strength testing in prone rotation position in asymptomatic participants, and to compare these two testing modalities.

Design

Reliability study.

Methods

A total of 20 asymptomatic participants (27.7 ± 7.4 years; 77.1 ± 10.1 kg) attended a strength assessment consisting of HHS and HHD tests. Reliability was assessed using the intra-class correlation coefficient (ICC) with 95% confidence intervals (CI), coefficient of variation (CV) with 95%CI, and standard error of measurement (SEM). Pearson correlation and linear regression analysis were used to compare HHS and HHD testing modalities.

Results

“Good” to “excellent” intra (ICC range = 0.896 to 0.979) and inter-rater reliability scores (ICC range = 0.850 to 0.978) were displayed during both HHS and HHD tests during internal and external rotation strength assessments. Linear relationships between HHS and HHD measures were found, with coefficients of determination (R 2) ranging between 0.60 and 0.79.

Conclusion

HHS and HHD resulted to be reliable strength assessment modalities for clinical practice. These assessment modes can be equally valid in assessing intra and inter-limb asymmetries in isometric shoulder rotation strength. The affordability and availability of HHS in ordinary clinical settings can facilitate its implementation in musculoskeletal practice.

Cite this article

Download citation ▾
Francesco Bettariga, Nicola Francesco Lopomo, Fabio Civera, Stefano Giuseppe Lazzarini, Lisa Mantovani, Luca Maestroni. Reliability and Validity of Hand-Held Dynamometer and Hand-Held Sphygmomanometer for Testing Shoulder Isometric External and Internal Rotator Muscles Strength. Journal of Science in Sport and Exercise 1-8 DOI:10.1007/s42978-023-00232-1