Simultaneous determination of seven impurities in high-purity cobalt oxide by ICP-AES after matrix separation using 1-nitroso-2-naphthol as a precipitant

Yan Wang , Chun-shan Zhou , Qing-bing Duan , Ping Chen

Journal of Central South University ›› 2002, Vol. 9 ›› Issue (2) : 100 -103.

PDF
Journal of Central South University ›› 2002, Vol. 9 ›› Issue (2) : 100 -103. DOI: 10.1007/s11771-002-0051-x
Article

Simultaneous determination of seven impurities in high-purity cobalt oxide by ICP-AES after matrix separation using 1-nitroso-2-naphthol as a precipitant

Author information +
History +
PDF

Abstract

A method was developed for the simultaneous determination of seven trace impurities (Cd, Mn, Pb, Zn, Cu, Fe and Ni) in high-purity cobalt oxide by ICP-AES. The matrix effect was eliminated by precipitation with 1-nitroso-2-naphthol. The matrix effect of cobalt on the absorptions of trace impurities, the effects of reaction time, pH value, dosage of precipitant on the formation of cobalt-1-nitroso-2-naphthol complex, the effects of hydrochloric acid on the stability of this complex and masking of elements were studied. Recoveries of the impurities in spiked sample are from 90% to 110% with a precision of 1.1%–5.0% RSD. The detection limits of the seven elements are in the range of 0.01–0.24µg/g. The method can be applied to the analysis of high-purity cobalt metal, cobalt oxide and other cobalt compounds.

Keywords

trace impurity / high-purity cobalt oxide / matrix separation / inductively coupled plasma atomic emission spectrometry (ICP-AES)

Cite this article

Download citation ▾
Yan Wang, Chun-shan Zhou, Qing-bing Duan, Ping Chen. Simultaneous determination of seven impurities in high-purity cobalt oxide by ICP-AES after matrix separation using 1-nitroso-2-naphthol as a precipitant. Journal of Central South University, 2002, 9(2): 100-103 DOI:10.1007/s11771-002-0051-x

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Chinese National Standard GB6518-86[S]. Beijing: State Bureau of Standards of People’s Republic of China (in Chinese).

[2]

KujiraiO, YamadaK. Application of cobalt ammine for the simultaneous determination of traces of As, Fe, Ti, V and Zr in high-purity cobalt metal by lanthanum hydroxide coprecipitation and inductively coupled plasma-atomic emission spectrometry[J]. Fresenius J Anal Chem, 1996, 354: 428-431

[3]

WatsonA E. The commissioning of a 5 kW inductively coupled plasma unit for use with a 3.4-meter Ebert spectrograph [J]. Rep-Natl Inst Metall, 1979, 2029: 13-13(S Afr)

[4]

LiFan, FanJian. Recent development of studies on the interferences and their correction methods in ICP-AES[J]. Spectroscopy and Spectral Analysis, 1998, 18(3): 325-328(in Chinese)

[5]

ZhangShu-zhen, WangChun-mei. Determination of impurities in trimethylgalliun by ICP-AES[J]. Spectroscopy and Spectral Analysis, 1997, 17(6): 53-56(in Chinese)

[6]

FanJian, HuangJin-yu, ZouMei-mei. Determination of trace elements in higher pure tungsten by ICP-AES[J]. J Cent South Univ Technol, 1999, 30(1): 71-73(in Chinese)

[7]

ChenShou-chunImportant inorganic chemistry reactions, 19822nd editionShanghai, Shanghai Technology Press: 301-301(in Chinese)

[8]

ZhangShun-wei, TangFu-long, ZhangTaiHandbook of modern chemistry reagent II chemistry analytical reagent (in Chinese) [M], 1987, Beijing, Chemistry Industry Press: 502-504