INTRODUCTION
Biomacromolecules, including proteins and nucleic acids, perform diverse physiological activities throughout an organism’s life cycle (Sun
et al.
2021; Zhang
et al.
2018b). Changes in the abundance and distribution of these biomacromolecules can serve as indicators of disease occurrence and therapeutic outcomes (Karimzadeh
et al.
2020). Therefore, analysis of the abundance and spatial distribution of vital biomacromolecules is essential for clinical diagnosis and pharmacological evaluation (Lv
et al.
2013; Sun
et al.
2020).
Mass spectrometry (MS), invented by Francis William Aston in 1919, has evolved into a widely used analytical technique for investigating small molecules and biomacromolecules over the past century. The underlying principle involves ionizing the sample by an ion source to generate charged ions, which are subsequently accelerated by an electric field to form an ion beam directed into a mass analyzer. Within this analyzer, ions with different mass-to-charge ratios experience opposite velocity dispersion due to the combined effects of electric and magnetic fields, ultimately resulting in the generation of a mass spectrum that provides information about their respective masses. With advancements in ion source technologies, various branches of mass spectrometry have emerged including electrospray ionization mass spectrometry (ESI-MS), matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS), inductively coupled plasma mass spectrometry (ICP-MS), secondary ion mass spectrometry (SIMS), among others.
The analytical methods for biomacromolecules by MS can be normally classified into label-free and labeled approaches. Label-free detection of biomacromolecules can be realized by molecular MS, such as ESI-MS and MALDI-MS. Following sample preparation and separation, biomolecule identification can be accomplished through molecular MS combined with peptide sequencing via tandem MS (MS/MS), as well as through database-driven protein identification using bioinformatics (Cristoni and Bernardi
2003).
The classic label-based approaches for analyzing biomacromolecules refer to widely accepted immunoassay methods. For instance, immunofluorescence imaging and flow cytometry analysis are widely employed in biomacromolecule analysis due to the advantages of simplicity, high sensitivity, and rapid response (Sargazi
et al.
2022). Unfortunately, multiple detection is limited by the overlap of emission bandwidth and the quenching of fluorescent dyes. Therefore, it is a great challenge to simultaneously measure the biomacromolecules with the abundance differing in an order of magnitude, or simultaneously measuring kinds of targets (Anyaegbu
et al.
2019; Hulspas
et al.
2009). Additionally, immunofluorescence imaging is susceptibly interfered with by the autofluorescence and scattering light derived from samples, lacking reliable quantification procedures (Chen
et al.
2009; Liu
et al.
2016).
Metal stable isotope labeling, coupled with highly sensitive elemental MS technology, such as ICP-MS and SIMS offers a compelling alternative to fluorescence detection and has proven effective in quantifying various target molecules (Liu
et al.
2014). By replacing fluorescent groups with stable metal isotopes for antibody or ligand labeling, the abundance of each isotope can be correlated with specific antibody probes, allowing for precise measurement of antigen levels. In theory, ICP-MS and SIMS provide high sensitivity and isotopic resolution for over 100 elements, highlighting the great potential of metal stable isotope labeling for advanced multi-dimensional analysis (Han
et al.
2013; Zhang
et al.
2020a).
This review provides a concise introduction of metal stable isotope labeling combined with elemental MS techniques for biomacromolecules analysis (
Fig. 1). First, the constitution and feasibility of exogenous metal stable isotopes probes, normally classified into polymer, small molecule, and inorganic nanoparticle probes, are introduced. Next, we review the current instrumentation state of elemental MS, including ICP-MS and SIMS with the merits of high sensitivity and accuracy, as well as multi-element simultaneous detection ability in the quantification of biomolecules with metal stable isotope labeling. Furthermore, we highlight the practical applications of the combined strategy in biomarkers discovery, clinical pathological analysis, and precision medicine development at molecular, cellular, and tissue levels. Lastly, we conclude with current challenges and prospects for the future directions of the field.
METAL STABLE ISOTOPE LABELED PROBES
Initially, endogenous non-metal atoms such as sulfur, phosphorus, and selenium were employed for quantifying biomacromolecules as endogenous elemental tags. However, their insufficient detection sensitivity is due to the low elemental ionization efficiency or interference from multi-atomic spectra (He
et al.
2013). To effectively detect low-abundance biomolecules, exogenous metal stable isotopes are utilized as the key component of the labeling agents. These metal stable isotope labeled probes, which act as reporters can be normally categorized into three groups, including polymer-based probes, small molecule probes, and inorganic nanoparticle probes,
etc.
Polymer-based probes
Metal chelating polymers (MCPs) typically incorporate functional groups such as 1,4,7,10-tetraazacyclododecane-1,4,7,10-tetraacetic acid (DOTA) or diethylenetriaminepentaacetic acid (DTPA) to effectively chelate metal stable isotopes as probes (Kumar
et al.
2023; Schwarz
et al.
2014). These chelators form stable complexes with lanthanide (Ln) metal ions and exhibit low dissociation constants (Han
et al.
2018). For cell labeling, minimal isotopic exchange occurs between different Ln metal conjugates, allowing the simultaneous coupling of multiple Ln metals to polymers. Initially introduced by Canadian company DVS Sciences by the name Maxpar reagents, MCP reagents are currently marketed by Standard BioTools Inc., which is formerly Fluidigm Inc.
In the meantime, Maxpar X8 Antibody Labeling Kit was developed by Standard BioTools as a commercially available metal isotope labeled probe that utilizes the thiol-maleimide coupling to covalently link MCP with antibodies containing disulfide bonds, such as immunoglobulin G (IgGs). The disulfide bonds in the antibody hinge region are reduced by tris(2-carboxyethyl) phosphine (TCEP) to generate thiols, which then selectively react with maleimide-functionalized MCPs through Michael addition (Arnett
et al.
2023; Lou
et al.
2007). Although this bioconjugation has some advantages, it poses challenges for other antibody types like IgMs, IgEs, and IgAs. To overcome this limitation, copper-free click chemistry has been introduced as an innovative conjugation strategy (Allo
et al.
2018). Terminal azides are enrolled to functionalize MCPs and selectively react with cyclooctyne-functionalized antibodies. The orthogonal chemistry broadens the range of antibody classes and is compatible with the thiol-maleimide conjugation method for preparing bispecific probes.
MCPs covalently conjugated antibodies offer numerous advantages, rendering them an optimal choice for metal stable isotope labeled probes. Nevertheless, the probes necessitate further refinement. Firstly, the available range of metal stable isotopes is limited. Currently, chelating agents such as DTPA and DOTA employed in MCPs preparation predominantly chelate Ln metals (Cho
et al.
2019), thereby impeding the application of other elements. Therefore, the introduction of novel chelating agents would expand the range of isotopes accessible and empower MCPs-antibodies for multiparametric detection (Zhang
et al.
2022a; Zhang
et al.
2022b). Meanwhile, both the capacity for polymers to chelate isotope ions and the potential for antibodies to couple with polymers require enhancement. So far, MCPs can bind 20–50 metal ions per polymer, resulting in chelating approximately 100–250 metal ions per antibody. This quantity is sufficient to detect and quantify highly abundant biomarkers, approximately 10
4−10
7 per cell. The highly branched polymers with abundant chelating groups would facilitate greater incorporation of metal atoms, significantly augmenting the analytical sensitivity and accuracy for low abundance biomolecules.
Small molecule probes
Metal chelates belong to a class of non-polymeric reagents, that utilize functional groups to chelate metal ions (Delgado-Gonzalez and Sanchez-Martin
2021). Apart from Ln elements, metal chelates can incorporate other metal stable isotopes. For instance, new barcode reagents have been developed by chelating six isotopes of palladium through a bifunctional molecule known as isothiocyanobenzyl-EDTA (Zunder
et al.
2015). These reagents expand the number of measurement channels available for MS and avoid interference with the Ln metal-antibody as labels for measurement.
The DNA metal intercalator, as another small molecule probe can label double-stranded DNA molecules through electrostatic attraction, groove binding, and insertion between nucleotide base units (Ornatsky
et al.
2008). It is a complex comprising transition metals such as Ir and Rh, two non-intercalating ligands, and one intercalating ligand (Liu and Sadler
2011; Rana
et al.
2021). In cells stained with DNA metal intercalator, the metal signal is associated with the nucleus. It is therefore most commonly used to identify and count cells. Cisplatin, as a specific DNA metal intercalator, is a well-established chemotherapy drug targeting DNA, which can penetrate non-living cells and form platinum-sulfur bonds with protein through nucleophilic groups (
e.
g., R–SH and R–S–R). This property renders it suitable as a cell viability probe (Devine
et al.
2021). Furthermore, cisplatin can directly label proteins by conjugation with antibodies (Mei
et al.
2016). Antibodies labeled with cisplatin are compatible with MCPs-antibodies in labeling and washing procedures, without any overlap of isotopic signals in MS detection.
Inorganic nanoparticles probes
The typical inorganic nanoparticles as MS probes comprise metal nanoparticles, metal nanoclusters (MNCs), quantum dots, metal oxide nanoparticles, upconversion nanoparticles, and metal-organic framework nanoparticles (MOFs). Thereinto, metal nanoparticles consist of densely packed metal atoms, enabling highly sensitive detection of biomolecules (Pichaandi
et al.
2019). However, larger particles may lead to spatial hindrance and reduce labeling efficiency despite stronger detection signals. MNCs comprise a concrete number of metal atoms with diameters less than 3 nm. Due to size proximity to Fermi wavelength, MNCs exhibit intriguing properties such as discrete electronic energy levels, chirality, magnetism, fluorescence,
etc. (Chakraborty and Pradeep
2017; Cifuentes-Rius
et al.
2021; Han
et al.
2021). These properties have been successfully applied in the dual-modal detection of biomolecules using a fluorescence microscope and MS (Zhang
et al.
2018a). Interestingly, size matching effects allow MNCs similar in size to proteins to achieve a 1:1 labeling radio when used for protein labeling (Li
et al.
2023a). It is noted that, metal nanoparticles show significant advantages in detecting target biomacromolecules with low expression levels due to the larger number of constituent metal atoms. However, in this case, the nonspecific adsorption capacity is significantly enhanced, which inevitably increases the risk of false positive results. In contrast, MNCs can accurately identify the target in a 1:1 ratio with less nonspecific adsorption, thereby greatly reducing the false positive signals. However, due to the limited number of metal atoms in MNCs, the signal amplification effect is relatively less significant.
As another typical nanometer-sized MS probes, quantum dots are luminescent semiconductor (CdSe, CdS, PbSe,
etc.) particles with unique optoelectronic properties for dual-modal analysis (Sanmartín-Matalobos
et al.
2022). In addition, metal oxide and upconversion nanoparticles have also been shown as potential metal labeling reagents (Ngamcherdtrakul
et al.
2019; Pichaandi
et al.
2017; Zhang
et al.
2020b). For instance, each 5.7 nm tantalum oxide nanoparticle (TaO
2 NP) contains approximately 2700 Ta atoms, while each 10 nm NaHoF
4 nanoparticle contains about 8000 Ho atoms, and they both have much more amounts of atoms than that of MCPs reagent (Pichaandi
et al.
2017; Zhang
et al.
2020b). Similarly, nanoscale MOFs can serve as another class of stable isotope probes to achieve accurate readout of metal signals, which provide a new perspective for these ordered pore structures (Chen
et al.
2022).
ELEMENTAL MASS SPECTROMETRY TECHNOLOGY APPLIED FOR ANALYZING METAL STABLE ISOTOPES LABELED BIOMACROMOLECULES
Elemental MS, ICP-MS and SIMS have long been recognized as highly sensitive instruments for detecting the abundance and distribution of biomolecules. Both techniques have contributed greatly to the field of elements analysis (Comi
et al.
2017). Specifically, ICP-MS lacks sensitivity towards light isotopes, whereas SIMS is an alternative suited for the direct detection of lipids and small molecules but faces challenges in specifically detecting biomacromolecules. In the following section, we mainly introduce the instrument composition, working principle, strengths and shortcomings of the above two MS techniques.
Inductively coupled plasma mass spectrometry
ICP-MS has long been regarded as a reliable and advanced technology in the field of inorganic element analysis, with exceptional characteristics such as high sensitivity, isotope selectivity, and a wide dynamic range (Liu
et al.
2011; Wang
et al.
2010). It integrates with chromatography and laser ablation (LA) systems, and is an irreplaceable analytical tool in life sciences, pharmaceutical research, and environmental monitoring. ICP-MS enables quantitative analysis of naturally occurring elements like sulfur, phosphorus, and metal elements to achieve precise detection of biomolecules (Liu
et al.
2017). By employing exogenous element labeling of biomolecules, background interference can be significantly alleviated. Notably, the utilization of MCPs and nanoparticles as probes for ICP-MS detection has further augmented its analytical sensitivity owing to the signal amplification (Cid-Barrio
et al.
2018; Liu
et al.
2016). It is crucial that, for absolute quantification of biomacromolecule through ICP-MS technique, several criteria should be met. (1) The number of metal atoms in the probe must be accurately determined, which is the basis for accurate quantitative analysis. (2) The determination of the recognition ratio is also indispensable, which involves how to accurately correlate the signal intensity with the actual number of metal atoms in the sample. (3) To avoid the influence of non-specific binding or interference, an effective procedure should be enrolled to separate the redundant probe from the labeled complex.
The mass analyzers of ICP-MS encompass quadrupole, sector field, and time-of-flight analyzers. Presently, the major mass spectrometers available on the market are quadrupole mass analyzers. However, quadrupoles belong to sequential mass analyzers and necessitate switching between two
m/
z values for dual isotope ratio measurements. Each switch requires a certain stabilization time to adjust the quadrupole voltage, potentially leading to signal loss during the measurement process. In contrast, inductively coupled plasma time-of-flight mass spectrometry (ICP-TOF-MS) can convert the continuous ion stream generated by ICP into discrete ion packets and precisely analyze each ion packet using TOF. Consequently, it enables the simultaneous acquisition of information from nearly the entire periodic table through transient signals (Au
et al.
2020,
2022; Lockwood
et al.
2024). Tian
et al compared the performance of inductively coupled plasma quadrupole mass spectrometry (ICP-Q-MS), ICP-TOF-MS, and multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) in dual isotope analysis of individual nanoparticle (NP) and cell (Tian
et al.
2023). The results demonstrated that ICP-Q-MS exhibited limited accuracy in determining isotopic ratios within a single NP/cell; whereas MC-ICP-MS encountered challenges in identifying events with minimal metal masses in NPs or cells. In contrast, ICP-TOF-MS has the capability to effectively monitor multiple
m/
z values concurrently while ensuring accurate assessment of elemental/isotopic relative abundances.
In 2009, Tanner and his colleagues at the University of Toronto pioneered the integration of traditional flow cytometry with ICP-TOF-MS, resulting in the development of mass cytometry (CyTOF) (Bandura
et al.
2009). Compared with fluorescence-based flow cytometry, CyTOF effectively overcomes the issue of the overlapping of emission spectra from fluorochromes. This breakthrough enables simultaneous measurement over 40 parameters on individual cell (Bjornson
et al.
2013; Labib and Kelley
2020; Xie and Ding
2022). CyTOF has now played an irreplaceable role in biological research such as monitoring cell signaling pathways, immune cell differentiation, and inflammation-induced mechanisms (Forbester
et al.
2020; Herring
et al.
2018; Krieg
et al.
2018).
Meanwhile, with the continuous advancements of spatial omics, clarifying the spatial distribution of target biomacromolecules within cells or tissues is imperative. To conquer the dilemma, in 1994, Wang
et al. pioneered the combination of LA with ICP-MS for bioimaging (Wang
et al.
1994). In 2005, Hutchinson
et al. successfully resolved the distribution of β-amyloid protein in Alzheimer’s disease plaques by combining LA-ICP-MS and immunohistochemistry (IHC) for biomolecular imaging analysis (Hutchinson
et al.
2005). Subsequently, various metal stable isotope labels and multiplexing methods have been developed to enable simultaneous analysis of multiple biomolecules in tissues. Imaging mass cytometry (IMC) technology integrates kinds of metal stable isotopes antibody labeling for LA-ICP-MS imaging, with high spatial resolution and low sample consumption (Yu
et al.
2020; Zhou
et al.
2021). It provides a novel method for cell mapping and identification of rare cellular subpopulations. Compared to previous single-event research methods, IMC shows significant potential in pathological research by distinguishing between healthy cells and diseased cells by comprehensively describing and understanding their biological characteristics. Notably, LA-ICP-MS requires cells to be mounted on a substrate, allowing for nearly 100% of individual cells available for analysis. In contrast, a significant number of cells are lost during the sample treatment and instrument processing in CyTOF, resulting in fewer than 50% of cells available for analysis. Additionally, LA-ICP-MS has the advantage of being able to analyze biomarkers from single cells spread across the substrate as well as those distributed within tissue samples.
Secondary ion mass spectrometry
SIMS was developed in 1960, with the principle of bombarding a sample’s surface using a high-energy primary ion beam. This process induces the absorption of energy and sputtering of surface molecules, generating detectable secondary ions (Boxer
et al.
2009). SIMS enables the characterization of component and structural features on the sample’s surface. By separating secondary ions with different mass-to-charge ratios through a mass analyzer, an intensity versus mass-to-charge ratio relationship curve can be obtained by counting their respective intensities. This technique offers several advantages, that is, it achieves nanometer-level spatial resolution for precise analysis of micrometer-scale surface areas and captures signals from multiple elements, providing insights into chemical bonding and molecular structures on the sample’s surface (Jiang
et al.
2016).
From the very beginning, SIMS is mainly used to characterize inorganic materials. Following the invention of static SIMS in 1968, its application has been expanded to encompass organic compound and biomolecule analysis. The integration of TOF into static SIMS during the 1990s propelled TOF-SIMS for investigating biomolecules and associated materials. Concurrently, researchers introduced a novel ion beam instrument tailored specifically for dynamic SIMS applications. This innovation was subsequently commercialized by CAMECA, Inc. by the name NanoSIMS. Despite both dynamic and static SIMS relying on secondary ion detection methods for sample analysis, each technique possesses distinct characteristics (Keren
et al.
2019; Nuñez
et al.
2017). NanoSIMS utilizes an energetic primary continuous ion beam, resulting in enhanced sensitivity and spatial resolution. On the other hand, TOF-SIMS employs a pulsed primary ion beam to realize two-dimensional imaging and depth analysis for determining the elemental and molecular composition of samples.
SIMS is also suitable for non-labeled imaging of lipid and metabolite distributions, facilitating direct analysis of the distribution and therapeutic effects of small molecule drugs, metal-containing drugs or biomaterials. Through labeling strategies, SIMS can extend its range of applications. For instance, natural isotopes can be replaced with rare isotopes such as
13C and
15N, which are then incorporated into metabolic intermediates, enabling the tracking and localization of biological synthesis processes (Baboo
et al.
2014). However, a limitation of SIMS is its difficulty in detecting specific proteins. Exogenous labeling is thus more suitable for detecting proteins with larger molecular weight. Although fluorine-19 (
19F) has been utilized as a protein label (Vreja
et al.
2015), its high background signal complicates the detection of biomacromolecules with low abundance (Agüi-Gonzalez
et al.
2021). In contrast, the probes comprising noble metals and Ln metals are scarce in biological samples and easily ionizable. Hence, metal stable isotope probes are well-suited for quantitative analysis and imaging of biomacromolecules using SIMS (Jin
et al.
2022; Li
et al.
2022).
As another rapid technological advancement of SIMS in recent years, the combination of multiple ion beam imaging (MIBI) and metal stable isotope labeling has found widespread application. By simultaneously quantifying dozens or even hundreds of proteins or RNAs, it improves the comprehensive spatial characterization of cellular subtypes. MIBI is particularly well-suited for evaluating intracellular biomolecules and elucidating cell phenotypes, epigenetics, and metabolic states. The high-definition multiple ion beam imaging (HD-MIBI) developed in 2021 achieves an impressive lateral resolution as low as approximately 30 nm (Rovira-Clavé
et al.
2021), rendering this technique potentially indispensable for subcellular resolution analysis of biomolecules. In addition, multiplexed beam imaging by the time of flight (MIBI-TOF) is frequently employed to enable the simultaneous detection of numerous proteins, which facilitates the quantitative analysis, distribution visualization, and spatial proteomics research following the elemental mass tags staining (Keren
et al.
2019; McCaffrey
et al.
2022; Risom
et al.
2022).
APPLICATIONS OF METAL STABLE ISOTOPES LABELING COUPLED WITH ICP-MS AND SIMS FOR BIOMACROMOLECULES ANALYSIS
Using metal stable isotope labeling with MS technology allows for analysis of almost all biological analytes, representing a significant advancement that offers a more comprehensive and precise analytical approach. In the following section, we will emphasize the typical applications of ICP-MS and SIMS in analyzing biomacromolecules at molecular, cellular and tissue levels by different detection methods, respectively.
Table 1 summarizes the recent reports that metal stable isotopes labeling to analyze expression levels and spatial distribution of biomolecules across various levels.
Molecular level
The first work for quantifying biomacromolecules at the molecular level by specific metal probe labelling was reported by Zhang and co-workers for the assessment of thyroid-stimulating hormone (TSH) in human serum (Zhang
et al.
2001). Since then, the use of Ln consists of 14 stable elements as reporters allowing for highly multiplexing biomolecule analysis with good sensitivity and precision in mixed biological samples (Tholey and Schaumlöffel
2010). To achieve higher analytical sensitivity, amplification is a preferred choice. In 2018, as shown in
Fig. 2A, Hu and his colleagues pioneered the integration of tyramine signal amplification with the self-amplification effect of gold nanoparticles (Li
et al.
2018). Through this dual amplification strategy combined with ICP-MS technology, they successfully detected alpha-fetoprotein (AFP) in serum with an impressive detection limit of 1.85 pg/mL. Furthermore, for miRNA detection, they developed an MNAzymes amplification strategy using streptavidin-modified magnetic beads and three Ln-labeled (
159Tb/
165Ho/
175Lu) single-stranded DNA substrates (Kang
et al.
2021). As shown in
Fig. 2B, three types of MNAzymes hybridize with their respective target miRNAs, which induces the cleavage and release of Ln-labeled substrates. The cleaved MNAzymes then cyclically cleave another substrate on the probe to achieve signal amplification.
Single nanoparticle analysis is an alternative method to enhance sensitivity. An advantage of single particle inductively coupled plasma mass spectrometry (SP-ICP-MS) is the detected signal pulse corresponds to a single NP, enabling rapidly determining the size, size distribution, and concentration of metal NPs in suspension. SP-ICP-MS was successfully employed for assessing serum biomarkers CA724, CA199, and CEA of gastric cancer patients by Huang
et al. (Huang
et al.
2022). As shown in
Fig. 2C, firstly, noble metal NPs, antibody-labeled magnetic microspheres and target proteins were mixed. Then noble metal NPs bound to the target proteins on the surface of magnetic microspheres through sandwich immuno-recognition. Subsequently, magnetic microspheres were separated magnetically, and the unreacted metal NPs were collected and quantified to assess the concentration of the target proteins. It provides an accurate and convenient method for monitoring malignant tumor prognosis and recurrence, especially when combined with biomarker panel strategies.
Cellular level
Cells are the most important research object for revealing the mechanism of disease development and exploring therapeutic approaches. The Maxpar commercial assay kits have been extensively employed in the investigation of various cellular biomarkers. The assay kits typically incorporate Ir-DNA intercalators for nucleic acid labeling, cisplatin for discriminating between live and dead cells, or MCPs conjugated with antibodies for specific biomacromolecule targeting. For instance, Martínez-Diz
et al. utilized the Maxpar Human Monocyte/Macrophage Phenotyping Panel Kit to access peripheral blood samples of COVID-19 patients by CyTOF (Martínez-Diz
et al.
2022). By using a panel comprising 15 Ln metal probes alongside TaqMan probes for genotyping, they correlated several markers such as TMPRSS2, CD45
-, CD163/CD206, and CD33 with COVID-19 invasiveness, providing evidence for patient classification in a clinical setting. Similarly, Maxpar commercial assay kits have also found applications in research areas including immunotherapy and the investigation of mechanisms underlying the onset complications (Krieg
et al.
2018; Shen
et al.
2022). However, as shown in
Fig. 3A, owing to the limited availability of Ln isotopes, Ding and colleagues developed antibody-modified Zr-NMOFs compatible with four additional detection channels alongside MCPs-antibodies to enable simultaneous labeling of multiple proteins (Dang
et al.
2021). They also devised
198Pt-cisplatin that can rapidly penetrate the damaged cells and non-specifically interact with proteins, peptides, and amino acids to label carrier cells (
Li et al. 2023b). The inclusion of carrier cells boosts initial counts during the detection process which helps to counteract the reduced resolution and distorted analysis results when examining target cells. This approach is congruent with viability assays based on cisplatin.
Single-cell analysis uniquely identifies and characterizes cellular heterogeneity. Additionally, different from the targets distributed on the cytomembrane, accurately assessing the cell volume is crucial for determining the quantity of the intracellular analytes in individual cell. Recently, Qin
et al. proposed a novel approach for precise determination of cell volume by detecting ruthenium red (RR) binding with polysaccharides on the cell surface by single-cell (SC)-ICP-TOF-MS analysis (Qin
et al.
2021). The method exhibited species-dependent staining intensity and enabled estimation of cellular volume when cells exhibit regular morphology. Further, as present in
Fig. 3B, Menero-Valdés
et al. employed RR along with three metal clusters to label suspended cells, investigating both the quantity and relative concentration variations of ferritin, metallothionein-2, and transferrin within each cell under two distinct stress conditions (Menero-Valdés
et al.
2023). These findings provide valuable insights into studying cellular heterogeneity and offer a deeper comprehension of biochemical processes at an individual cellular level.
Circulating tumor cells (CTCs) are rare tumor cells that are released from primary tumors and circulate in the bloodstream for remote metastasis, which act as prognostic indicators for breast, prostate, colorectal cancer and other cancers (Lawrence
et al.
2023; Nikanjam
et al.
2022; Sarioglu
et al.
2015). Zhang
et al. developed a sea urchin-inspired single CTC recognition platform (Zhang
et al.
2021). As illustrated in
Fig. 4A, this platform integrated a dual aptamer connected to AuNPs (sea urchin-DMA-AuNPs) that facilitated efficient capture and sorting of CTCs, and utilized ICP-MS for background-free analysis. This innovative platform demonstrated a remarkable 100% CTC capture efficiency, enabling rapid detection of individual CTCs within just one hour from a mere 100 μL of whole blood.
To further achieve non-destructive counting, release and culture of CTCs, Yin
et al. developed a multifunctional platform. This platform effectively captured MCF-7 cells by recognition of mucin 1 (MUC1) on the cell membrane with aptamer-decorated capture probes (Yin
et al.
2020). Subsequently, the starting primer hybridized with the aptamer was replaced by MUC1 and released into solution. The released primer was separated from the captured cells and hybridized with Tb-labeled substrate on detection probes. This process was followed by releasing a large amount of nicked Tb fragments through the nicking endonuclease-assisted amplification, enabling subsequent ICP-MS detection. On the other hand, after digestion by nucleases, the captured cells were detached from the probe and were continuously cultured. Finally, ICP-MS counting demonstrated a recovery rate of 52.7% at a detection limit of 87 for MCF-7 cells, while maintaining a cellular viability of 74.3%. Moreover, the demand for faster single-cell detection has driven advancements in the ICP-TOF-MS technique toward higher speed. As present in
Fig. 4B, Löhr
et al. developed a single-cell array and integrated it with LA-ICP-TOF-MS for quantitative analysis and isotopic fingerprinting, achieving high-throughput detection of 550 cells per hour (Löhr
et al.
2019). In this work, the precise operation and arrangement of single cells have been successfully achieved by the single-cell array. In detail, Ir-DNA and mDOTA-Ho were used to stain the DNA and protein of the suspension cells, respectively. With the aid of aligned droplets containing the same probe at known concentrations and the LA-ICP-TOF-MS technique, both probes can be quantified at the single-cell level. In addition, this work examined the isotopic fingerprint of single cells, and employed Ir-DNA as a biomarker to assess cell cycle status. It was found that Zn and P contents increased as the cell period moved from G1 to S and G2. This indicates that elemental biomarkers are related to changes in the content of endogenous isotopes.
Tissue level
Conventional histopathologic analysis methods, such as immunofluorescence, primarily employ fluorescently labeled antibodies for tissue section staining and subsequent microscopic examination to assess the distribution of target analytes within the tissue. However, this method allows for the analysis of a limited number of biomarkers, it becomes valueless when dealing with an increased number of biomarkers due to spectral overlap issues (Angelo
et al.
2014). Alternatively, metalstable isotope barcoding circumvents spontaneous fluorescence interference, thereby reducing background signals. In a typical process, histopathologic tissues are primarily acquired through fresh freezing, formalin fixation, or paraffin embedding. Then metal stable isotope-coded probes label the multiplexed biomarkers expressed on tissues. Lastly, high-resolution MS techniques enable simultaneous detection of numerous metal isotopes labeled-antibodies recognized biomacromolecules.
IMC refers to a technology that ablates tissue sections using a laser, brings the ablated component through an inert gas flow into an inductively coupled plasma ion source, and performs isotopic analysis via a mass flow cytometer with a TOF mass analyzer. Recently, Erreni
et al. developed a panel of 28 markers specifically designed for detecting the spatial distribution, as well as the interactions of tumor cells and immunocytes in the tumor microenvironment (TME) at the frozen tissue sections level (Erreni
et al.
2024). This panel enables effective cell segmentation and phenotypic clustering, facilitating IMC analysis of TME cell distribution in different pancreatic ductal adenocarcinoma (PDAC) mouse models. Moreover, additional markers can be incorporated to expand this panel, unraveling the complexity of PDAC and providing insights into prognosis or treatment options for this disease.
As another representative study, Ijsselsteijn
et al. developed a panel of 40 markers for determining major immune cell subsets by IMC, as well as investigating tumor-immune cell interactions within the tumor-immune microenvironment on formalin-fixed paraffin-embedded (FFPE) tissues (Ijsselsteijn
et al.
2019). To ensure full depth penetration of the probes into the tissue, the antibodies were allowed for precise control over incubation temperature and duration time to maximize the antibodies’ performance, while maintaining signal intensity and specificity for antigen detection.
MIBI-TOF MS has a higher spatial resolution than ICP-TOF-MS down to 250 nm, and is therefore used for high-resolution elemental analysis of tissues. Vijayaragavan
et al. validated 39 antibodies to identify brain tissue-specific targets through conventional IHC, the targets were then labeled and imaged with MIBI-TOF MS (Vijayaragavan
et al.
2022). Human brain FFPE samples were stained with 36 distinct metal isotopes labeled antibodies, followed by images of various brain regions at different stages of Alzheimer’s disease using a 36-dimensional MIBI-TOF model. This method enabled the identification of cell types and protein pathologies associated with diverse stages of Alzheimer’s disease in the hippocampus.
As the pioneer researchers in this field, Nolan and colleagues at Stanford University have developed a robust method known as PANINI, which integrates nucleic acid and protein imaging (Jiang
et al.
2022). As shown in
Fig. 5, this method employs protease-free branched-chain amplification of nucleic acids, tyramide signal amplification combined with hapten deposition for MIBI imaging to simultaneously quantify the expressing levels of viral DNA (vDNA), viral RNA (vRNA), and 31 proteins within tissue sections. It was utilized to analyze viral reservoirs and immune responses in lymphoid tissues from simian immunodeficiency virus SIV-infected and uninfected controls, revealing synergistic immune events in the lymphoid tissue reservoirs of retroviral infection. In the study of viral infection mechanisms, simultaneous analysis of proteins and nucleic acids can determine the state of infected cells and virus-host interactions, and can also reveal the immune regulatory process of SIV-infected tissues.
Additionally, Nolan and his colleagues developed EpicTags, a combined labeling technique using six palladium isotopes that merged with MIBI (EpicMIBI) for in situ tracking of barcodes within the tissue microenvironment (Rovira-Clavé
et al.
2022). EpicTags enable barcode encoding for 20 cell lines and were employed to dissect spatial components of cell lineages and phenotypes in xenograft models of small cell lung cancer. EpicMIBI facilitates in situ detection of tumor heterogeneity related to both cell-intrinsic and cell-extrinsic processes. Its capability to track barcoded cancer cells in situ, along with their surrounding spatial environment, opens new opportunities for understanding tumor differentiation.
Despite the notable advantages of MS imaging, its speed has consistently posed a significant challenge, hindering its clinical application. One effective approach to address this issue involves employing fluorescence analysis to identify regions of interest (ROI) initially, followed by subsequent MS analysis. As illustrated in
Fig. 6, Yu
et al. have developed a series of blue, green, and red fluorescent Ln-doped multicolor carbon nanodots (MC-Cdots (Ln)) as dual-mode probes for both fluorescence and mass detection. These nanodots were combined with Maxpar antibody reagents for tissue analysis (Yu
et al.
2021). By rapidly identifying ROIs by traditional immunofluorescence and subsequent multiplex IMC analysis, this strategy reduces blind scanning time for IMC by 90% while compensating for its low resolution. This methodology offers a promising solution for clinical applications that necessitate rapid imaging.
Further, an “all in one” probe was fabricated to facilitate a multi-modal accurate diagnosis of early-stage Alzheimer’s disease. Recently, Lu
et al. employed gold NCs that penetrate the blood–brain barrier (BBB) and target connective tissue growth factor (CTGF) expressed in the brain to analyze CTGF in early-stage APP/PS1 transgenic mice (Lu
et al.
2024). The abnormal overexpression of CTGF predate Aβ deposition could be a potential biomarker of early-stage AD. Moreover, combined with ICP-MS quantitative techniques, through
in vivo NIR-II imaging,
in vitro visible fluorescence and peroxidase-like colorimetric imaging, the multi-modal analysis of CTGF on postmortem brain slices was conducted to differentiate AD patients from healthy individuals. In this case, the detection methods of different modalities can complement each other and provide valuable information for a comprehensive understanding of the AD pathological process. With NIR-II imaging, the probe can be well home to the elevated CTGF
in vivo, enabling noninvasive and real-time early elevation of CTGF. The spatial distribution and overall expression level of CTGF in mouse brain sections can meanwhile be evaluated by fluorescence and peroxidase-like colorimetric imaging. Thereafter, CTGF expression analysis on brain pathological tissue level was performed by LA-ICP-TOF-MS, verifying the consistency of cell analysis and molecular imaging analysis, and finally providing the “gold standard” for diagnosis.
SUMMARY AND PERSPECTIVES
Biologists heavily rely on detecting biomolecular abundance and distribution to deepen their understanding of life processes. Accurate quantification and clarifying the distribution of biomolecules is crucial for both fundamental biological research and practical clinical diagnostics. Accompanied by the advancements of metal stable isotope labeling and exogenous tags, including polymer, small molecule, and inorganic nanoparticle probes, elemental mass spectrometry instruments have emerged as indispensable tools for analyzing various target biomacromolecules. The review briefly summarizes the applications of ICP-MS and SIMS with elemental tags for quantifying biomacromolecules including proteins and nucleic acids. Particularly in the field of multiparametric biomacromolecule analysis, it outperforms traditional spectroscopic methods regarding sensitivity and resolution. However, we still face many unresolved issues and challenges. Although there are currently nearly 50 isotopes used for biomacromolecule labeling, there are still many non-biological element isotopes that have not been applied. To address these issues, new labeling reagents need to be designed to utilize more elemental isotopes while optimizing the specificity and efficiency of labeling. Additionally, continuous improvement and breakthroughs in detection methodology are also needed to achieve lower detection limits, higher resolution, faster speed, and greater throughput. We anticipate metal stable isotope labeling strategies combined with elemental mass spectrometry technique playing a more confounding role in routine laboratory and clinical settings, leading to a deeper understanding of physiological processes across different fields.