Comparative analysis of pulmonary vein isolation strategies using pulsed field, cryoballoon, and robotic magnetic navigation ablation

Xiao-Yu Liu , Ying Liu , Ling-Ling Qian , Jie Zheng , Ku-Lin Li , Hua-Yan You , Shi-Peng Dang , Xiao-Xi Zhao , Ru-Xing Wang

Vessel Plus ›› 2026, Vol. 10 ›› Issue (1) : 34

PDF
Vessel Plus ›› 2026, Vol. 10 ›› Issue (1) :34 DOI: 10.20517/2574-1209.2025.148
Original Article
Comparative analysis of pulmonary vein isolation strategies using pulsed field, cryoballoon, and robotic magnetic navigation ablation
Author information +
History +
PDF

Abstract

Aim: Pulsed field ablation (PFA) has emerged as a promising strategy for catheter ablation of atrial fibrillation (AF). This study compared perioperative outcomes of pulmonary vein isolation (PVI) performed using robotic magnetic navigation (RMN), cryoballoon (CRYO) ablation, and PFA.

Methods: This retrospective study included patients with AF who underwent PVI using RMN ablation (RMN group, n = 112), CRYO ablation (CRYO group, n = 189), or PFA (PFA group, n = 50). Procedural characteristics, perioperative complications, and postoperative recovery were analyzed.

Results: Total procedure time was longer in the PFA group than in the CRYO group but shorter than in the RMN group (123.9 ± 14.0 vs. 98.3 ± 14.3 vs. 147.9 ± 19.3 min, P < 0.001). However, left atrial procedure time was shortest in the PFA group (26.6 ± 6.4 vs. 44.7 ± 12.1 vs. 95.1 ± 20.5 min, P < 0.001). Fluoroscopy time was comparable between the PFA and CRYO groups, and significantly shorter in the RMN group (15.2 ± 3.0 vs. 15.0 ± 3.1 vs. 7.4 ± 2.5 min, P < 0.001). Although acute procedural success was comparable across groups, first-pass PVI was achieved more frequently with PFA than with CRYO or RMN {50/50 [100%] vs. 170/189 [89.9%] vs. 100/112 [89.3%], P = 0.030}. The overall incidence of perioperative complications did not differ significantly among the three groups.

Conclusion: PFA demonstrated acute safety and efficacy comparable to those of RMN and CRYO ablation for AF, while offering shorter left atrial procedure time and a higher rate of first-pass isolation.

Keywords

Pulsed field ablation / robotic magnetic navigation / cryoballoon / atrial fibrillation

Cite this article

Download citation ▾
Xiao-Yu Liu, Ying Liu, Ling-Ling Qian, Jie Zheng, Ku-Lin Li, Hua-Yan You, Shi-Peng Dang, Xiao-Xi Zhao, Ru-Xing Wang. Comparative analysis of pulmonary vein isolation strategies using pulsed field, cryoballoon, and robotic magnetic navigation ablation. Vessel Plus, 2026, 10 (1) : 34 DOI:10.20517/2574-1209.2025.148

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Johner N,Shah DC.Catheter ablation of atrial fibrillation: time to look beyond iterative pulmonary vein isolation only and “one-size-fits-all” strategies.Indian Pacing Electrophysiol J2025;25:199-206 PMCID:PMC12461723

[2]

Andrade JG, Wells GA, Deyell MW, et al.; EARLY-AF Investigators. Cryoablation or drug therapy for initial treatment of atrial fibrillation.N Engl J Med2021;384:305-15

[3]

Martins Esteves I,Trevisan T,Gomes WF.Pulsed-field ablation versus radiofrequency ablation for paroxysmal atrial fibrillation: a systematic review and meta-analysis.J Arrhythm2025;41:e70158 PMCID:PMC12320123

[4]

Iasiello M,Bianco N.Effects of pulsed radiofrequency source on cardiac ablation.Bioengineering2023;10:227 PMCID:PMC9952521

[5]

Reichlin T, Kueffer T, Badertscher P, et al.; SINGLE SHOT CHAMPION Investigators. Pulsed field or cryoballoon ablation for paroxysmal atrial fibrillation.N Engl J Med2025;392:1497-507

[6]

Reddy VY,Koruth JS.Pulsed field ablation for pulmonary vein isolation in atrial fibrillation.J Am Coll Cardiol2019;74:315-26

[7]

Schmidt B,Neven K.EUropean real-world outcomes with Pulsed field ablatiOn in patients with symptomatic atRIAl fibrillation: lessons from the multi-centre EU-PORIA registry.Europace2023;25:euad185 PMCID:PMC10320231

[8]

Urbanek L,Schaack D.Pulsed field versus cryoballoon pulmonary vein isolation for atrial fibrillation: efficacy, safety, and long-term follow-up in a 400-patient cohort.Circ Arrhythm Electrophysiol2023;16:389-98

[9]

Schmidt B,Tohoku S.5S study: safe and simple single shot pulmonary vein isolation with pulsed field ablation using sedation.Circ Arrhythm Electrophysiol2022;15:e010817

[10]

Reddy VY,Koruth J.Pulsed field ablation in patients with persistent atrial fibrillation.J Am Coll Cardiol2020;76:1068-80

[11]

Della Rocca DG,Bala G.Transient inferior ST-segment elevation and ventricular fibrillation after cavotricuspid isthmus pulsed-field ablation.JACC Clin Electrophysiol2023;9:704-6

[12]

Reddy VY, Gerstenfeld EP, Natale A, et al.; ADVENT Investigators. Pulsed field or conventional thermal ablation for paroxysmal atrial fibrillation.N Engl J Med2023;389:1660-71

[13]

Hara S,Hachiya H.Long-term outcomes after second-generation cryoballoon ablation of paroxysmal atrial fibrillation - feasibility of a single short freeze strategy without bonus applications.Int J Cardiol2020;306:90-4

[14]

Liu XY,Zheng J.The impact of left atrial size in catheter ablation of atrial fibrillation using remote magnetic navigation.Cardiol Res Pract2018;2018:3096261 PMCID:PMC6304809

[15]

Krzowski B,Peller M.Pulmonary vein isolation with radiofrequency ablation and pulsed-field ablation: follow-up with real-world data.Pol Arch Intern Med2025;135:17062

[16]

Badertscher P,Reichlin T.Pulsed-field ablation vs. thermal ablation for paroxysmal atrial fibrillation: a systematic review and meta-analysis of randomized controlled trials.Europace2025;27:euaf179 PMCID:PMC12395552

[17]

Ijaz R,Qazi M.Comparison between efficacy and safety of remote magnetic navigation and manual catheter navigation for atrial fibrillation ablation: an updated meta-analysis and systematic review.J Innov Card Rhythm Manag2025;16:6307-28 PMCID:PMC12233321

[18]

Noten AME,Ernst S.Best practices in robotic magnetic navigation-guided catheter ablation of cardiac arrhythmias, a position paper of the Society for Cardiac Robotic Navigation.Front Cardiovasc Med2024;11:1431396 PMCID:PMC11466809

[19]

Isenegger C,Jordan F.Pulsed-field ablation versus cryoballoon ablation in patients with persistent atrial fibrillation.Int J Cardiol Heart Vasc2025;59:101684 PMCID:PMC12076779

[20]

Maurhofer J,Madaffari A.Pulsed-field vs. cryoballoon vs. radiofrequency ablation: a propensity score matched comparison of one-year outcomes after pulmonary vein isolation in patients with paroxysmal atrial fibrillation.J Interv Card Electrophysiol2024;67:389-97 PMCID:PMC10902096

[21]

Keževičiūtė M,Sudavičienė D.Comparative studies of different ablation techniques for atrial fibrillation.Rev Cardiovasc Med2025;26:33490 PMCID:PMC12135642

[22]

Della Rocca DG,Magnocavallo M.; HRMC Investigators. Pulsed electric field, cryoballoon, and radiofrequency for paroxysmal atrial fibrillation ablation: a propensity score-matched comparison.Europace2023;26:euae016 PMCID:PMC11119976

[23]

de Campos MCAV,Daher RF.Pulsed-field ablation versus thermal ablation for atrial fibrillation: a meta-analysis.Heart Rhythm O22024;5:385-95 PMCID:PMC11228281

[24]

Kaddoura R,Asaad N.Pulsed-field ablation versus radiofrequency or cryoballoon thermal ablation in atrial fibrillation: a systematic review and meta-analysis.Postgrad Med J2025;101:690-703

[25]

Vetta G,Parlavecchio A.Multielectrode catheter-based pulsed electric field vs. cryoballoon for atrial fibrillation ablation: a systematic review and meta-analysis.Europace2024;26:euae293

[26]

Ekanem E, Reddy VY, Schmidt B, et al.; MANIFEST-PF Cooperative. Multi-national survey on the methods, efficacy, and safety on the post-approval clinical use of pulsed field ablation (MANIFEST-PF).Europace2022;24:1256-66 PMCID:PMC11271404

[27]

Verma A, Haines DE, Boersma LV, et al.; PULSED AF Investigators. Pulsed field ablation for the treatment of atrial fibrillation: PULSED AF pivotal trial.Circulation2023;147:1422-32 PMCID:PMC10158608

[28]

Mountantonakis SE,Patel H.Incidence of laboratory-defined severe intravascular haemolysis across commercially available pulsed field ablation technologies for atrial fibrillation.Europace2025;27:euaf185 PMCID:PMC12455042

[29]

Mohanty S,Compagnucci P.Acute kidney injury resulting from hemoglobinuria after pulsed-field ablation in atrial fibrillation: is it preventable?.JACC Clin Electrophysiol2024;10:709-15

[30]

Xu Y,Lu Y.Hemolysis after pulsed-field ablation of atrial fibrillation.Heart Rhythm2025;22:1103-9

[31]

Krishan S,Agarwal S.Comparative efficacy and procedural outcomes of pulsed field ablation vs. thermal ablation for paroxysmal atrial fibrillation: a systematic review and meta-analysis of randomized trials.Eur Heart J Open2025;5:oeaf092 PMCID:PMC12345413

[32]

Watfa A,Mdaihly M.Safety of same-day discharge following pulsed-field ablation versus radiofrequency ablation: a comparative analysis.Pacing Clin Electrophysiol2025;48:852-8

PDF

1

Accesses

0

Citation

Detail

Sections
Recommended

/

〈 〉