Purpose: To develop and validate a fully automated quality assurance method for high-dose rate (HDR) brachytherapy, enabling the precise evaluation of source dwell-position accuracy and dwell-time linearity using radiochromic film analysis and advanced image processing algorithms.
Methods: A novel analytical framework was established to verify HDR source positional precision and dwell-time linearity. Radiochromic films were irradiated under predefined treatment plans. Image preprocessing involved grayscale conversion, median filtering, rotation correction, and absolute coordinate calibration. The centroid of each dwell position was determined using two-dimensional Gaussian fitting. Temporal linearity analysis employed radial grayscale summation within the radii of 250–400 pixels. Linearity was quantified using the coefficient of determination (R2), with R2 approaching 1 indicating optimal performance.
Results: Automated Gaussian fitting achieved sub-millimeter positional accuracy (maximum error ≤ 0.6 mm), eliminating manual intervention. Rotation correction algorithms effectively mitigated positional errors caused by film misalignment. Comparative analysis with commercial software yielded nearly identical results, validating the reliability of the method. Dwell-time analysis demonstrated excellent linearity (R2≥0.999) over summation radii ranging from 150 to 400 dpi.
Conclusions: This study presents a robust and fully automated quality assurance (QA) solution for HDR brachytherapy that simultaneously verifies sub-millimeter-level dwell-position accuracy and temporal linearity. The integration of radiochromic film dosimetry with algorithmic image processing eliminates manual intervention and standardizes QA workflows, representing a substantial advancement in treatment safety and efficiency.
| [1] |
Nath R, Anderson LL, Meli JA, et al. Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group No. 56. Med Phys. 1997; 24: 1557–1598.
|
| [2] |
Toita T, Wada K, Sutani S, et al. Definitive radiotherapy consisting of external beam radiotherapy without central shielding and 3D image-guided brachytherapy for patients with cervical cancer: feasibility for Japanese patients and dose-response analyses for local control in the low-dose range. Jpn J Clin Oncol. 2023; 53(6): 480-488.
|
| [3] |
Kim N, Park W, Cho WK, Cho YS. Clinical outcomes after positron emission tomography/computed tomography-based image-guided brachytherapy for cervical cancer. Asia Pac J Clin Oncol. 2022; 18(6): 743-750.
|
| [4] |
Gutiérrez-Valencia E, Navarro-Domenech I, Zhou K, et al. Partial or focal brachytherapy for prostate cancer: a systematic review and meta-analysis. Br J Radiol. 2025; 98(1167): 354-367.
|
| [5] |
Chitmanee P, Tsang Y, Tharmalingam H, et al. Single-Dose Focal Salvage High Dose Rate Brachytherapy for Locally Recurrent Prostate Cancer. Clin Oncol (R Coll Radiol). 2020; 32(4): 259-265.
|
| [6] |
Anderson B, Vicini F, Wazer D, Arthur D. Breast brachytherapy: Building a bright future on the foundation of a rich history of advancement in technology, technique, and patient-centered care. Brachytherapy. 2023; 22(3): 368-380.
|
| [7] |
Kanani A, Owrangi A, Yazdi M, Fatemi-Ardekani A, Mosleh-Shirazi MA. Development of a multi-purpose quality control phantom for MRI-based treatment planning in high-dose-rate brachytherapy of cervical cancer. J Contemp Brachytherapy. 2023; 15(1): 57–68.
|
| [8] |
Pötter R, Haie-Meder C, Van Limbergen E, et al. Recommendations from gynecological (GYN) GEC-ESTRO working group (II): Concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy—3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol. 2006; 78(1): 67–77.
|
| [9] |
Viswanathan AN, Cormack R, Holloway CL, et al. Magnetic resonance-guided interstitial therapy for vaginal recurrence of endometrial cancer. Int J Radiat Oncol Biol Phys. 2006; 66(1): 91–99.
|
| [10] |
Jones CH. HDR microSelectron quality-assurance studies using a well-type ion chamber. Phys Med Biol. 1995; 40(1): 95–101.
|
| [11] |
Thomadsen BR, Erickson BA, Eifel PJ, et al. A review of safety, quality management, and practice guidelines for high-dose-rate brachytherapy: executive summary. Pract Radiat Oncol. 2014; 4(1): 65–70.
|
| [12] |
Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40. Med Phys. 1994; 21(4): 581–618.
|
| [13] |
Watanabe Y, Murakami N, Inaba K, et al. Automated source tracking with a pinhole imaging system during high-dose-rate brachytherapy treatment. Phys Med Biol. 2018; 63(14):145002.
|
| [14] |
Otani Y, Tanaka S, Yoshida K, et al. High-dose rate intracavitary brachytherapy pretreatment dwell position verification using a transparent applicator. J Appl Clin Med Phys. 2018; 19(5): 266–273.
|
| [15] |
Nose T, Chatani M, Maruyama M, et al. Real-time verification of a high-dose-rate iridium 192 source position using a modified C-arm fluoroscope. Int J Radiat Oncol Biol Phys. 2017; 97(4): 858–865.
|
| [16] |
Tanderup K, Beddar S, Andersen CE, et al. In vivo dosimetry in brachytherapy. Medical Physics. 2013; 40(7):070902 1–15.
|
| [17] |
Evans MDC, Devic S, Podgorsak EB, et al. High dose-rate brachytherapy source position quality assurance using radiochromic film. Med Dosim. 2008; 32(1): 13–15.
|
| [18] |
Walb MC, Dorand JE, Flynn RT, et al. An improved source positioning verification device for HDR brachytherapy quality assurance. Brachytherapy. 2018; 17(4): S60–S61.
|
| [19] |
Houdek PV, Schwade JG, Brismar JS, et al. Dose determination in high dose-rate brachytherapy. Int J Radiat Oncol Biol Phys. 1992; 24(4): 795–801.
|
| [20] |
Awunor OA. Assessment of a source position checking tool for the quality assurance of transfer tubes used in HDR 192Ir brachytherapy treatments. Brachytherapy. 2018; 17(4): 628–633.
|
| [21] |
Rickey DW, Sasaki D, Dunscombe P, et al. A quality assurance tool for high-dose-rate brachytherapy. Med Phys. 2010; 37(6): 2525–2532.
|
| [22] |
Task Group on Accident Prevention and Safety in Radiation Therapy. Prevention of accidental exposures to patients undergoing radiation therapy. A report of the International Commission on Radiological Protection. Ann ICRP. 2000; 30(3): 7-70.
|
| [23] |
Valentin J. International Commission on Radiation Protection. Prevention of high-dose-rate brachytherapy accidents. ICRP Publication 97. Ann ICRP. 2005; 35(2): 1-51.
|
| [24] |
Huq MS, Fraass BA, Dunscombe PB, et al. The report of Task Group 100 of the AAPM: application of risk analysis methods to radiation therapy quality management. Med Phys. 2016; 43(7): 4209–4262.
|
Rights & permissions
2025 The Author(s). Precision Radiation Oncology published by John Wiley & Sons Australia, Ltd on behalf of Shandong Cancer Hospital & Institute.