Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine

Weili Zhong , Zhe (Jay) Chen , Min-Young Lee , Fada Guan , Huixiao Chen , Dae Yup Han

Precision Radiation Oncology ›› 2025, Vol. 9 ›› Issue (1) : 35 -42.

PDF
Precision Radiation Oncology ›› 2025, Vol. 9 ›› Issue (1) : 35 -42. DOI: 10.1002/pro6.70006
ORIGINAL ARTICLE

Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine

Author information +
History +
PDF

Abstract

Purpose: The response of various detectors in small fields from a variety of treatment machines has been studied and is summarized in IAEA TRS-483. However, data for the novel RefleXion system remains largely unexplored. This study measured the output correction factors of multiple detectors in small fields for a clinical RefleXion unit.

Methods: The RefleXion machine consists of a binary multi-leaf collimator and two pairs of Y-jaws with clinical openings of 1 and 2 cm. The reference dosimetry is applied to the 10 × 2 cm2 clinical-reference field, and the output factors of different fields are presented relative to the clinical-reference field. The responses of detectors Edge, Razor, Micro-Diamond, A14SL, CC01 and CC03 in rectangular fields from 1.25 × 1 to 20 × 2 cm2 on the RefleXion unit was studied at a depth of 10 cm in an IBA Blue-Phantom-Helix with a 85 cm source-to-surface distance. Gafchromic EBT4 film data in a solid-water phantom were used as the reference to obtain correction factors for the detectors.

Results: In the fields of the 2 cm jaw, all 6 detectors showed similar responses to the film reference within around 0.5% except at the first field width (1.25 cm), where the Edge and Micro-Diamond exhibited over-response and the CC13 showed the volume effect of ion chambers. In the fields of the 1 cm jaw, the Edge and Micro-Diamond had responses close to the film and the same over-response at small field-widths. Significant deviations of the CC13 (∼4%) and the A14SL (∼2.5%) from the film were present over the whole range of field widths.

Conclusions: The small field output correction factors of 6 kinds of detectors were determined for a RefleXion system, conforming to the formalism in TRS-483. All detectors except CC13 fulfil the 5% correction limit recommended by the TRS-483 for output factor measurement.

Keywords

field output factors / output correction factors / RefleXion / small field

Cite this article

Download citation ▾
Weili Zhong, Zhe (Jay) Chen, Min-Young Lee, Fada Guan, Huixiao Chen, Dae Yup Han. Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine. Precision Radiation Oncology, 2025, 9(1): 35-42 DOI:10.1002/pro6.70006

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Absorbed dose determination in external beam radiotherapy: An internation code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS-398, 2000.

[2]

Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999; 26(9): 1847-1870.

[3]

Dosimetry of small static fields used in external beam radiotherapy: An international code of practice for reference and relative dose determination. IAEA TRS-483, 2017.

[4]

Das IJ, Francescon P, Moran JM, et al. Report of AAPM Task Group 155: Megavoltage photon beam dosimetry in small fields and non-equilibrium conditions. Med Phys. 2021; 48(10): e886-e921.

[5]

Shi M, Chuang CF, Kovalchuk N, et al. Small-field measurement and Monte Carlo model validation of a novel image-guided radiotherapy system. Med Phys. 2021; 48(11): 7450-7460.

[6]

Guan F, Chen H, Draeger E, et al. Characterization of GafchromicTM EBT4 film with clinical kV/MV photons and MeV electrons. Prec Radiat Oncol. 2023; 7(2): 84-91.

[7]

Guan F, Wang X, Yang M, et al. Dosimetric response of Gafchromic™ EBT-XD film to therapeutic protons. Prec Radiat Oncol. 2023; 7(1): 15-26.

[8]

Ritt D. Scanning, calibrations, corrections, image manipultions. RIT online virtual training documents, 2020.

[9]

Howard ME, Herman MG, Grams MP. Methodology for radiochromic film analysis using FilmQA Pro and ImageJ. PLoS One. 2020; 15(5): e0233562.

[10]

Ringholz J, Sauer OA, Wegener S. Small field output correction factors at 18 MV. Med Phys. 2023; 50(11): 7177-7191.

RIGHTS & PERMISSIONS

2025 The Author(s). Precision Radiation Oncology published by John Wiley & Sons Australia, Ltd on behalf of Shandong Cancer Hospital & Institute.

AI Summary AI Mindmap
PDF

39

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/