Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine
Weili Zhong , Zhe (Jay) Chen , Min-Young Lee , Fada Guan , Huixiao Chen , Dae Yup Han
Precision Radiation Oncology ›› 2025, Vol. 9 ›› Issue (1) : 35 -42.
Measurements of Small Field Output Correction Factors on a RefleXion Treatment Machine
Purpose: The response of various detectors in small fields from a variety of treatment machines has been studied and is summarized in IAEA TRS-483. However, data for the novel RefleXion system remains largely unexplored. This study measured the output correction factors of multiple detectors in small fields for a clinical RefleXion unit.
Methods: The RefleXion machine consists of a binary multi-leaf collimator and two pairs of Y-jaws with clinical openings of 1 and 2 cm. The reference dosimetry is applied to the 10 × 2 cm2 clinical-reference field, and the output factors of different fields are presented relative to the clinical-reference field. The responses of detectors Edge, Razor, Micro-Diamond, A14SL, CC01 and CC03 in rectangular fields from 1.25 × 1 to 20 × 2 cm2 on the RefleXion unit was studied at a depth of 10 cm in an IBA Blue-Phantom-Helix with a 85 cm source-to-surface distance. Gafchromic EBT4 film data in a solid-water phantom were used as the reference to obtain correction factors for the detectors.
Results: In the fields of the 2 cm jaw, all 6 detectors showed similar responses to the film reference within around 0.5% except at the first field width (1.25 cm), where the Edge and Micro-Diamond exhibited over-response and the CC13 showed the volume effect of ion chambers. In the fields of the 1 cm jaw, the Edge and Micro-Diamond had responses close to the film and the same over-response at small field-widths. Significant deviations of the CC13 (∼4%) and the A14SL (∼2.5%) from the film were present over the whole range of field widths.
Conclusions: The small field output correction factors of 6 kinds of detectors were determined for a RefleXion system, conforming to the formalism in TRS-483. All detectors except CC13 fulfil the 5% correction limit recommended by the TRS-483 for output factor measurement.
field output factors / output correction factors / RefleXion / small field
| [1] |
Absorbed dose determination in external beam radiotherapy: An internation code of practice for dosimetry based on standards of absorbed dose to water. IAEA TRS-398, 2000. |
| [2] |
|
| [3] |
Dosimetry of small static fields used in external beam radiotherapy: An international code of practice for reference and relative dose determination. IAEA TRS-483, 2017. |
| [4] |
|
| [5] |
|
| [6] |
|
| [7] |
|
| [8] |
|
| [9] |
|
| [10] |
|
2025 The Author(s). Precision Radiation Oncology published by John Wiley & Sons Australia, Ltd on behalf of Shandong Cancer Hospital & Institute.
/
| 〈 |
|
〉 |