Objective: The biomechanical interaction between coronal alignment and short extension stem use in total knee arthroplasty (TKA) remains insufficiently understood, particularly regarding the balance between tibial component stability and stress shielding. This study aims to investigate the effects of varying bone cut angles on the tibial side and the use of a short extension stem on tibial component displacement and stress distribution. Additionally, it will examine the stresses experienced by both cortical and cancellous bone during primary TKA, employing three-dimensional finite element analysis and elucidating the biomechanical effects of combining kinematic alignment (KA) with a short extension stem in TKA.
Methods: A healthy adult female volunteer was selected to create a tibial model of the left knee joint utilizing three-dimensional reconstruction and finite element analysis software. Six finite element models were developed based on a standard tibial plateau prosthesis and a tibial prosthesis with a 30 mm short extension stem. These models were categorized according to mechanical alignment (MA), KA, and a 3° valgus alignment. The simulations were designed to replicate the loads and constraints experienced during weight-bearing activities of the lower limb. Subsequently, we analyzed the displacement and stress distribution of the tibial prosthesis and examined the stress distribution within both the tibial cortical bone and cancellous bone.
Result: In the KA model, there was a notable increase in displacement, an elevation in mean cortical bone Von Mises stress, and a reduction in mean component Von Mises stress when compared to both the MA and the 3° valgus alignment model. The design of the short extension stem effectively decreased the displacement of the prosthesis but led to increased stress on the prosthesis itself. Additionally, this design led to a reduction in stress on the cortical bone, suggesting a more pronounced stress-shielding effect.
Conclusion: The application of KA in TKA will increase prosthetic displacement and reduce the stress-shielding effect, while the application of short extension stem will reduce prosthetic displacement and increase the stress-shielding effect. The combination of the KA and short extension stem may play a complementary role, but this needs further research to confirm.
| [1] |
J. M. Tarazi, Z. Chen, G. R. Scuderi, and M. A. Mont, “The Epidemiology of Revision Total Knee Arthroplasty,” Journal of Knee Surgery 34, no. 13 (2021): 1396–1401.
|
| [2] |
T. J. M. van Otten and C. J. M. van Loon, “Early Aseptic Loosening of the Tibial Component at the Cement-Implant Interface in Total Knee Arthroplasty: A Narrative Overview of Potentially Associated Factors,” Acta Orthopaedica Belgica 88, no. 1 (2022): 103–111.
|
| [3] |
X. Li, J. Lai, X. Yang, H. Xu, and S. Xiang, “Intra-Articular Injection of Vancomycin After Arthrotomy Closure Following Gentamicin-Impregnated Bone Cementation in Primary Total Knee Arthroplasty Provides a High Intra-Articular Concentration While Avoiding Systemic Toxicity: A Prospective Study,” Journal of Orthopaedic Surgery and Research 19, no. 1 (2024): 856.
|
| [4] |
C. Zhao, Q. Liao, D. Yang, M. Yang, and P. Xu, “Advances in Perioperative Pain Management for Total Knee Arthroplasty: A Review of Multimodal Analgesic Approaches,” Journal of Orthopaedic Surgery and Research 19, no. 1 (2024): 843.
|
| [5] |
S. Hasan, B. L. Kaptein, R. Nelissen, K. T. van Hamersveld, S. Toksvig-Larsen, and P. J. Marang-van de Mheen, “The Influence of Postoperative Coronal Alignment on Tibial Migration After Total Knee Arthroplasty in Preoperative Varus and Valgus Knees: A Secondary Analysis of 10 Randomized Controlled Trials Using Radiostereometric Analysis,” Journal of Bone and Joint Surgery. American Volume 103, no. 24 (2021): 2281–2290.
|
| [6] |
H. X. Liu, P. Shang, X. Z. Ying, and Y. Zhang, “Shorter Survival Rate in Varus-Aligned Knees After Total Knee Arthroplasty,” Knee Surgery, Sports Traumatology, Arthroscopy 24, no. 8 (2016): 2663–2671.
|
| [7] |
B. S. Lee, H. I. Cho, S. I. Bin, J. M. Kim, and B. K. Jo, “Femoral Component Varus Malposition Is Associated With Tibial Aseptic Loosening After TKA,” Clinical Orthopaedics and Related Research 476, no. 2 (2018): 400–407.
|
| [8] |
C. Riviere, F. Iranpour, E. Auvinet, et al., “Alignment Options for Total Knee Arthroplasty: A Systematic Review,” Orthopaedics & Traumatology, Surgery & Research 103, no. 7 (2017): 1047–1056.
|
| [9] |
J. Bellemans, W. Colyn, H. Vandenneucker, and J. Victor, “The Chitranjan Ranawat Award: Is Neutral Mechanical Alignment Normal for All Patients? The Concept of Constitutional Varus,” Clinical Orthopaedics and Related Research 470, no. 1 (2012): 45–53.
|
| [10] |
T. Karasavvidis, C. A. Pagan Moldenhauer, F. S. Haddad, M. T. Hirschmann, M. W. Pagnano, and J. M. Vigdorchik, “Current Concepts in Alignment in Total Knee Arthroplasty,” Journal of Arthroplasty 38, no. 7 Suppl 2 (2023): S29–S37.
|
| [11] |
T. Takahashi, J. Ansari, and H. G. Pandit, “Kinematically Aligned Total Knee Arthroplasty or Mechanically Aligned Total Knee Arthroplasty,” Journal of Knee Surgery 31, no. 10 (2018): 999–1006.
|
| [12] |
A. M. Almaawi, J. R. B. Hutt, V. Masse, M. Lavigne, and P. A. Vendittoli, “The Impact of Mechanical and Restricted Kinematic Alignment on Knee Anatomy in Total Knee Arthroplasty,” Journal of Arthroplasty 32, no. 7 (2017): 2133–2140.
|
| [13] |
T. Tran, P. McEwen, Y. Peng, et al., “Kinematic Alignment in Total Knee Arthroplasty: A Five-Year Prospective, Multicentre, Survivorship Study,” Bone & Joint Open 3, no. 8 (2022): 656–665.
|
| [14] |
Y. Bar Ziv, A. Essa, K. Lamykin, et al., “Minimum 2-Year Radiographic and Clinical Outcomes of Unrestricted Kinematic Alignment Total Knee Arthroplasty in Patients With Excessive Varus of the Tibia Component,” Journal of Personalized Medicine 12, no. 8 (2022): 1206.
|
| [15] |
A. D. Hinman, H. A. Prentice, E. W. Paxton, and M. P. Kelly, “Modular Tibial Stem Use and Risk of Revision for Aseptic Loosening in Cemented Primary Total Knee Arthroplasty,” Journal of Arthroplasty 36, no. 5 (2021): 1577–1583.
|
| [16] |
R. L. Barrack, C. Rorabeck, M. Burt, and J. Sawhney, “Pain at the End of the Stem After Revision Total Knee Arthroplasty,” Clinical Orthopaedics and Related Research 367 (1999): 216–225.
|
| [17] |
H. U. Cameron and Y. B. Jung, “Noncemented Stem Tibial Component in Total Knee Replacement: The 2- to 6-Year Results,” Canadian Journal of Surgery 36, no. 6 (1993): 555–559.
|
| [18] |
B. W. Cho, K. T. Kang, H. M. Kwon, et al., “Biomechanical Effect of Anatomical Tibial Component Design on Load Distribution of Medial Proximal Tibial Bone in Total Knee Arthroplasty: Finite Element Analysis Indicating Anatomical Design Prevents Stress-Shielding,” Bone & Joint Research 11, no. 5 (2022): 252–259.
|
| [19] |
Y. R. Li, Y. H. Gao, C. Yang, et al., “Finite-Element Analysis of the Proximal Tibial Sclerotic Bone and Different Alignment in Total Knee Arthroplasty,” BMC Musculoskeletal Disorders 20, no. 1 (2019): 617.
|
| [20] |
A. Halder, I. Kutzner, F. Graichen, B. Heinlein, A. Beier, and G. Bergmann, “Influence of Limb Alignment on Mediolateral Loading in Total Knee Replacement: In Vivo Measurements in Five Patients,” Journal of Bone and Joint Surgery. American Volume 94, no. 11 (2012): 1023–1029.
|
| [21] |
D. Y. Chong, U. N. Hansen, and A. A. Amis, “Analysis of Bone-Prosthesis Interface Micromotion for Cementless Tibial Prosthesis Fixation and the Influence of Loading Conditions,” Journal of Biomechanics 43, no. 6 (2010): 1074–1080.
|
| [22] |
E. J. Miller, R. F. Riemer, T. L. Haut Donahue, and K. R. Kaufman, “Experimental Validation of a Tibiofemoral Model for Analyzing Joint Force Distribution,” Journal of Biomechanics 42, no. 9 (2009): 1355–1359.
|
| [23] |
H. M. Kwon, H. T. Hong, I. Kim, et al., “Biomechanical Effects of Stem Extension of Tibial Components for Medial Tibial Bone Defects in Total Knee Arthroplasty: A Finite Element Study,” Journal of Knee Surgery 37, no. 13 (2024): 879–886.
|
| [24] |
P. Xing, J. Qu, S. Feng, J. Guo, and T. Huang, “Comparison of the Efficacy of Robot-Assisted Total Knee Arthroplasty in Patients With Knee Osteoarthritis With Varying Severity Deformity,” Journal of Orthopaedic Surgery and Research 19, no. 1 (2024): 872.
|
| [25] |
J. Yang, X. Li, P. Liu, X. Liu, L. Li, and M. Zhang, “The Impact of Patellofemoral Joint Diseases on Functional Outcomes and Prosthesis Survival in Patients Undergoing Unicompartmental Knee Arthroplasty: A Systematic Review and Meta-Analysis,” Journal of Orthopaedic Surgery and Research 19, no. 1 (2024): 840.
|
| [26] |
Y. Liu, Z. Xing, B. Wu, et al., “Association of MRI-Based Knee Osteoarthritis Structural Phenotypes With Short-Term Structural Progression and Subsequent Total Knee Replacement,” Journal of Orthopaedic Surgery and Research 19, no. 1 (2024): 699.
|
| [27] |
S. Nakamura, Y. Tian, Y. Tanaka, et al., “The Effects of Kinematically Aligned Total Knee Arthroplasty on Stress at the Medial Tibia: A Case Study for Varus Knee,” Bone & Joint Research 6, no. 1 (2017): 43–51.
|
| [28] |
Q. H. Zhang, A. Cossey, and J. Tong, “Stress Shielding in Periprosthetic Bone Following a Total Knee Replacement: Effects of Implant Material, Design and Alignment,” Medical Engineering & Physics 38, no. 12 (2016): 1481–1488.
|
| [29] |
J. R. Martin, C. D. Watts, D. L. Levy, T. M. Miner, B. D. Springer, and R. H. Kim, “Tibial Tray Thickness Significantly Increases Medial Tibial Bone Resorption in Cobalt-Chromium Total Knee Arthroplasty Implants,” Journal of Arthroplasty 32, no. 1 (2017): 79–82.
|
| [30] |
Z. H. Zhang, Y. S. Qi, B. G. Wei, H. R. Bao, and Y. S. Xu, “Application Strategy of Finite Element Analysis in Artificial Knee Arthroplasty,” Frontiers in Bioengineering and Biotechnology 11 (2023): 1127289.
|
| [31] |
R. Lozano, V. Campanelli, S. Howell, and M. Hull, “Kinematic Alignment More Closely Restores the Groove Location and the Sulcus Angle of the Native Trochlea Than Mechanical Alignment: Implications for Prosthetic Design,” Knee Surgery, Sports Traumatology, Arthroscopy 27, no. 5 (2019): 1504–1513.
|
| [32] |
C. P. Walsh, S. Han, C. D. Canham, J. L. Gonzalez, P. Noble, and S. J. Incavo, “Total Knee Arthroplasty in the Osteoporotic Tibia: A Biomechanical Evaluation of the Role of Stem Extensions and Cementing Techniques,” Journal of the American Academy of Orthopaedic Surgeons 27, no. 10 (2019): 370–374.
|
| [33] |
S. Xie, N. Conlisk, D. Hamilton, C. Scott, R. Burnett, and P. Pankaj, “Metaphyseal Cones in Revision Total Knee Arthroplasty: The Role of Stems,” Bone & Joint Research 9, no. 4 (2020): 162–172.
|
| [34] |
A. Heidari, M. H. Ebrahimzadeh, M. Daliri, A. Moradi, M. Sahebi, and M. Sadeghi, “Tibial Stem Extension Versus Standard Configuration in Primary Cemented Total Knee Arthroplasty: Systematic Review and Meta-Analysis,” Journal of Orthopaedic Surgery and Research 20, no. 1 (2025): 13.
|
| [35] |
Q. Wan, A. Zhang, Y. Liu, et al., “Stem to Prevent Periprosthetic Fracture After Notching in Total Knee Arthroplasty,” International Journal for Numerical Methods in Biomedical Engineering 40, no. 10 (2024): e3842.
|
| [36] |
J. L. Prudhon, R. Verdier, and J. H. Caton, “Primary Cementless Total Knee Arthroplasty With or Without Stem Extension: A Matched Comparative Study of Ninety Eight Standard Stems Versus Ninety Eight Long Stems After More Than Ten Years of Follow-Up,” International Orthopaedics 43, no. 8 (2019): 1849–1857.
|
| [37] |
A. C. Filip, S. A. Cuculici, S. Cristea, et al., “Tibial Stem Extension Versus Standard Configuration in Total Knee Arthroplasty: A Biomechanical Assessment According to Bone Properties,” Medicina (Kaunas, Lithuania) 58, no. 5 (2022): 634.
|
| [38] |
D. Wautier, S. Ftaita, and E. Thienpont, “Radiolucent Lines Around Knee Arthroplasty Components: A Narrative Review,” Acta Orthopaedica Belgica 86, no. 1 (2020): 82–94.
|
| [39] |
K. T. Kang, J. Son, O. R. Kwon, and Y. G. Koh, “Malpositioning of Prosthesis: Patient-Specific Total Knee Arthroplasty Versus Standard Off-The-Shelf Total Knee Arthroplasty,” Journal of the American Academy of Orthopaedic Surgeons Global Research & Reviews 1, no. 4 (2017): e020.
|
Rights & permissions
2026 The Author(s). Orthopaedic Surgery published by Tianjin Hospital and John Wiley & Sons Australia, Ltd.