Objective: For atlantoaxial joint disorders, traditional surgical approaches often presented challenges such as significant trauma and prolonged recovery. Therefore, it was crucial to explore safer and more effective surgical alternatives. The primary aim of this study was to investigate the anatomical safety and feasibility of artificial atlanto-odontoid joint (AAOJ) replacement via a transoral pharyngeal approach, through simulated surgical procedures and postoperative anatomical and radiological studies.

Methods: The novel AAOJ replacement surgery was simulated on 18 fresh adult cadaveric head and neck specimens, and relevant anatomical parameters were measured. Postoperatively, the specimens underwent X-ray and CT scans, and software was used to measure the relevant parameters of the fixation screws. The spatial relationships between the atlantoaxial components, fixation screws, and critical anatomical structures were also examined. The comparison of parameters between the left and right sides was conducted using paired-sample t-tests.

Results: The transoral pharyngeal approach provided adequate exposure, clear surgical visualization, and sufficient working space. Anatomical measurements showed that the width of the anterior arch bone window of the atlas was (13.8 ± 0.7) mm; the width of the vertebral body bone window of the axis was (11.0 ± 0.4) mm; the distance between the insertion points for the atlas screws was (28.2 ± 4.0) mm; the distance from the atlas insertion points to the lateral joint edge of the atlanto-axial joint was (5.2 ± 0.9) mm; the distance between the insertion points for the axis screws was (16.8 ± 1.6) mm; and the distance from the axis insertion points to the lateral joint edge of the atlanto-axial joint was (7.7 ± 0.9) mm. Radiological measurements showed that the screw trajectory length of the lateral mass screw in the atlas was (21.5 ± 2.8) mm, the outward insertion angle was (13.2 ± 2.5)°, and the caudal insertion angle was (3.5 ± 1.1)°; for the pedicle screw of the axis, the screw trajectory length was (29.8 ± 2.8) mm, the outward insertion angle was (20.7 ± 2.8)°, and the caudal insertion angle was (16.6 ± 2.7)°. The prosthesis was precisely fitted to the upper cervical spine, with adequate safety distances between the atlantoaxial components, fixation screws, and critical anatomical structures such as the foramen transversarium, vertebral artery groove, and spinal canal.

Conclusions: The transoral pharyngeal approach for novel AAOJ replacement is anatomically safe and feasible.

Keywords

arthroplasty / artificial atlanto-odontoid joint / atlantoaxial instability / transoral pharyngeal approach / upper cervical spine

Cite this article

Download citation ▾
Shengguang Lei, Lingyong Jing, Xijiong Chen, Zhiwei Gan, Jianbin Zhong, Liming Yu, Yong Hu. Design of a Novel Artificial Atlanto-Odontoid Joint and Its Anatomical and Radiological Studies Following Transoral Pharyngeal Approach Arthroplasty. Orthopaedic Surgery, 2025, 17(9): 2670-2679 DOI:10.1111/os.70088

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

P. Palmisciano, Y. Al Fawares, C. Woodhouse, et al., “The Impact of C1 Anterior Arch Preservation on Spine Stability After Odontoidectomy: Systematic Review and Meta-Analysis,” World Neurosurgery 167 (2022): 165-175.

[2]

H. Ansaripour, S. J. Ferguson, and M. Flohr, “In Vitro Biomechanics of the Cervical Spine: A Systematic Review,” Journal of Biomechanical Engineering 144, no. 10 (2022): 100801.

[3]

J. Wölfle-Roos, “Atlas Block and Alar Ligament Lesion-Underestimated or Overrated?,” Schmerz 38, no. 5 (2024): 352-358.

[4]

D. K. Singh, D. Shankar, N. Singh, R. K. Singh, and V. K. Chand, “C2 Screw Fixation Techniques in Atlantoaxial Instability: A Technical Review,” Journal of Craniovertebral Junction & Spine 13, no. 4 (2022): 368-377.

[5]

W. W. Wroe, B. Budde, and J. C. Hsieh, “C1-C2 Wiring With Timed Self-Obsolescence for Complex C2 Body Fracture,” Oper Neurosurg (Hagerstown) 21, no. 3 (2021): E278-e282.

[6]

J. J. Rasouli, A. Momin, B. Fiani, J. Shao, and T. A. Moore, “C1-C2 Arthrodesis With C1 Lateral Mass and C2 Parsicle Screws Combined With Interlaminar Arthrodesis and Wiring,” Clinical Spine Surgery 35, no. 2 (2022): 76-79.

[7]

S. Diaz, S. Zenkhri, P. Omoumi, et al., “C2 Odontoid Fracture Associated With C1-C2 Rotatory Dislocation: A Retrospective Analysis of 2 Surgical Techniques,” World Neurosurgery 180 (2023): e460-e467.

[8]

Z. L. Cao, S. Z. Zhong, J. F. Liu, C. D. Xu, and Z. H. Li, “Anatomic Study on Designation of Artificial Atlanto-Odontoid Joint,” Chinese Journal of Traumatology 19, no. 1 (2003): 27-29.

[9]

Z. L. Cao, S. Z. Zhong, J. F. Liu, et al., “Biomechanical Evaluation of the Replacement With Artiflcial Atlanto-Odontoid Joint,” Orthopedic Journal of China 11, no. 3 (2003): 97-99.

[10]

Z. L. Cao, S. Z. Zhong, J. F. Liu, et al., “Three-Dimensional Movements of the Atlanto-Axis Joint After Replacement With Artificial Atlanto-Odontoid Joint,” Chinese Journal of Spine and Spinal Cord 13, no. 2 (2003): 29-31.

[11]

M. S. Tan, G. B. Zhang, H. Y. Wei, et al., “Anatomic Study and Designation of Artificial Semi-Atlanto-Odontoid Joint,” Chinese Journal of Spine and Spinal Cord 40, no. 10 (2004): 25-28.

[12]

Y. Hu, H. Xie, and S. H. Yang, “Advancement of Study of Artificial Atlanto-Odontoid Joint,” China Journal of Orthopaedics and Traumatology 10, no. 5 (2007): 358-360.

[13]

B. Lu, G. Q. Li, J. W. Li, L. T. Li, Y. Xie, and G. X. Bai, “Biomechanical Study of Atlas Odontoid Process Replacement,” Orthopedic Journal of China 16, no. 18 (2008): 1408-1411.

[14]

B. Lu, G. Q. Li, J. W. Li, L. T. Li, Y. Xie, and G. X. Bai, “Atlantoaxial Anatomy Measurement-Based Design of Atlas Odontoid Process Artificial Joint,” Chinese Journal of Tissue Engineering Research 12, no. 17 (2008): 3237-3240.

[15]

Y. Hu, S. H. Yang, H. Xie, R. M. Xu, and W. H. Ma, “Quantitative Anatomic Study of Atlanto-Odontoid Joint and Designation of Artificial Atlanto-Odontoid Joint,” Chinese Journal of Traumatology 22, no. 11 (2006): 858-863.

[16]

Y. Hu, Y. J. Gu, X. F. He, et al., “Biomechanical Evaluation of Stability and Three-Dimensional Movements of the Atlantoaxial Joint After Artificial Atlanto-Odontoid Joint Arthroplasty,” Orthopaedic Surgery 2, no. 2 (2010): 111-118.

[17]

J. Chen, H. Liu, and Z. M. Yang, “The Clinical Anatomy Study on Upper Cervical Spine and the Design of Artificial Atlanto-Axial Joint,” Chinese Journal of Clinical Anatomy 17, no. 2 (1999): 56-58.

[18]

Y. Hu, X. F. He, R. M. Xu, W. H. Ma, W. D. Zhao, and D. Z. Liang, “Biomechanical Study of Atlas-Axis Components Pull-Out Strength Following Artificial Atlanto-Odontoid Joint Arthroplasty,” Chinese Journal of Experimental Surgery 26, no. 4 (2009): 522-524.

[19]

Y. Hu, Z. T. Chu, S. F. Shen, et al., “Biomechanical Properties of Novel Lateral Hole Pedicle Screws and Solid Pedicle Screws: A Comparative Study in the Beagle Dogs,” Orthopedic Surgery 15, no. 1 (2022): 328-336.

[20]

Y. Hu, X. Chen, Z. Chu, et al., “Micro-Computed Tomography Analysis and Histological Observation of the Screw-Bone Interface of Novel Porous Scaffold Core Pedicle Screws and Hollow Lateral Hole Pedicle Screws: A Comparative Study in Bama Pigs,” World Neurosurgery 190 (2024): e315-e322.

[21]

X. Zou, M. Cai, H. Yang, et al., “Transoral Atlantoaxial Reduction Plate (TARP) Technique for the Treatment of Irreducible Atlantoaxial Dislocation (IAAD) Caused by Old Odontoid Fracture,” Global Spine Journal 15, no. 4 (2025): 2340-2347.

[22]

J. F. Liu, Z. H. Wu, G. Z. Xu, S. Y. Liu, J. H. Qu, and R. Q. Zhong, “Surgical Treatment for Atlantoaxial Fracture or Dislocation,” Chinese Journal of Traumatology 14, no. 3 (1998): 46-48.

[23]

Q. S. Yin, J. F. Liu, H. Xia, Z. H. Wu, and G. Z. Xu, “Experience in Preventing Infection in Transoral Anterior Approach Surgery for Atlantoaxial Joint,” Chinese Journal of Nosocomiology 15, no. 11 (2002): 32-33.

[24]

C. Wang, G. D. Dang, and Z. J. Liu, “Anterior Transarticular Screw Fixation Through C2 Vertebral Body for Atlantoaxial Instability,” Chinese Journal of Orthopaedics 19, no. 8 (1999): 7-9.

[25]

F. Kandziora, R. Pflugmacher, K. Ludwig, G. Duda, T. Mittlmeier, and N. P. Haas, “Biomechanical Comparison of Four Anterior Atlantoaxial Plate Systems,” Journal of Neurosurgery 96, no. 3 Suppl (2002): 313-320.

[26]

Q. S. Yin, F. Z. Ai, H. Xia, Z. H. Wu, X. H. Mai, and J. F. Liu, “Design and Biomechanical Evaluation of Transoralpharyngeal Atlantoaxial Reduction System,” Chinese Journal of Orthopaedic Trauma 6, no. 2 (2004): 54-57.

[27]

Y. Hu, S. H. Yang, R. M. Xu, et al., “Quantitative Anatomic Study and Investigation of Screw Placement Safety in Chinese Atlas,” China Journal of Orthopaedics and Traumatology 20, no. 10 (2007): 679-683.

[28]

T. Obo, T. Fujishiro, M. Mizutani, et al., “Biologic Agents Preserve the C-2 Pedicle in Patients With Rheumatoid Arthritis: A Comparative Imaging Study Using Three-Dimensional Computed Tomography,” World Neurosurgery 149 (2021): e42-e50.

[29]

L. Zhang and H. Wang, “Biomechanical Effects of Lateral Inclination C(1) and C(2) Pedicle Screws on Atlantoaxial Fixation,” Orthopaedic Surgery 13, no. 7 (2021): 2027-2033.

[30]

F. Wu, H. Li, S. Wan, et al., “The Effect of Axis Pedicle and Intra-Axial Vertebral Artery on C (2) Pedicle Screw Placement,” Chinese Journal of Reparative and Reconstructive Surgery 36, no. 7 (2022): 866-872.

[31]

Q. Zang, J. Li, Y. Liu, et al., “Atlantoaxial Non-Fusion Using Biomimetic Artificial Atlanto-Odontoid Joint: Technical Innovation and Initial Biomechanical Study,” Spine 47, no. 11 (2022): 825-832.

[32]

Y. Q. Du, G. Y. Qiao, Y. H. Yin, T. Li, and X. G. Yu, “Posterior Atlantoaxial Facet Joint Reduction, Fixation and Fusion as Revision Surgery for Failed Suboccipital Decompression in Patients With Basilar Invagination and Atlantoaxial Dislocation: Operative Nuances, Challenges and Outcomes,” Clinical Neurology and Neurosurgery 194 (2020): 105793.

RIGHTS & PERMISSIONS

2025 The Author(s). Orthopaedic Surgery published by Tianjin Hospital and John Wiley & Sons Australia, Ltd.

AI Summary AI Mindmap
PDF

21

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/