Background: Component alignment is a key factor influencing clinical outcomes after total knee arthroplasty (TKA). Previous studies have shown that single-plane alignment can significantly affect knee joint kinematics and biomechanics. However, the comprehensive impact of multiplanar malalignment has been rarely investigated.

Objective: This study aimed to investigate the influence of the multiplanar malalignment combination on the polyethylene tibial liners under gait loading, a primary activity of daily life, as well as the degree of the influence of the alignments on the different planes.

Method: A validated finite element model of a cruciate-retaining knee prosthesis under gait loading was used in this study. Five alignment parameters (−5°, −3°, 0°, 3°, 5°) on each plane (coronal, sagittal, and transverse) were selected to simulate clinical alignment errors, resulting in 125 models combining various alignment errors across the three planes. Boundary and loading conditions were set according to ISO 14243-3:2014. The maximum von Mises stress and contact stress during a gait cycle were recorded for statistical analysis. A polynomial model was used for regression analysis, with the degree of influence of each alignment error on von Mises and contact stress determined by examining the quadratic coefficients.

Results: The highest Mises and contact stress values occurred with alignment errors of 5° varus, 5° flexion, and 5° internal rotation on the coronal, sagittal, and transverse planes, respectively. The lowest stress values were observed with a combination of 3° valgus, 5° flexion, and 0° internal rotation. The regression analysis yielded an R2 value of 0.69 between alignment errors and Mises stress, with quadratic coefficients of 0.096, 0.013, and 0.064 for the coronal, sagittal, and transverse alignments, respectively. For contact stress, the R2 was 0.697, with quadratic coefficients of 0.083, 0.002, and 0.026 for the coronal, sagittal, and transverse alignments, respectively.

Conclusion: The coronal alignment of the lower limb has the most significant impact on both Mises stress and contact stress of the tibial liner, followed by the rotational alignment of the tibial component. In contrast, the sagittal alignment of the femoral component has the least influence.

Keywords

finite element analysis / gait loading / multiplanar malalignment / prosthetic mechanics / total knee arthroplasty

Cite this article

Download citation ▾
Yichao Luan, Min Zhang, Xiang Dong, Hongping Duan, Zhiwei Wang, Zhichang Li, Cheng-Kung Cheng. Comprehensive Impact of Multiplanar Malalignment on Prosthetic Mechanics Under Gait Loading After Total Knee Arthroplasty—A Finite Element Analysis. Orthopaedic Surgery, 2025, 17(7): 2112-2120 DOI:10.1111/os.70068

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

T. Konishi, S. Hamai, H. Tsushima, et al., “Pre- and Postoperative Coronal Plane Alignment of the Knee Classification and Its Impact on Clinical Outcomes in Total Knee Arthroplasty,” Bone & Joint Journal 106-b, no. 10 (2024): 1059-1066.

[2]

Y. H. Kim, J. W. Park, Y. S. Jang, and E. J. Kim, “Minimum 19-Year Clinical Results and Patient Satisfaction After Total Knee Arthroplasty,” Journal of Arthroplasty 39, no. 8 (2024): 2068-2073.

[3]

S. K. Patel, L. T. Buller, E. R. Deckard, and R. M. Meneghini, “Survivorship and Patient Outcomes of Conforming Bearings in Modern Primary Total Knee Arthroplasty: Mean 3.5 Year Follow-Up,” Journal of Arthroplasty 39, no. 11 (2024): 2737-2744.

[4]

C. A. Kahlenberg, B. U. Nwachukwu, A. S. McLawhorn, M. B. Cross, C. N. Cornell, and D. E. Padgett, “Patient Satisfaction After Total Knee Replacement: A Systematic Review,” HSS Journal 14, no. 2 (2018): 192-201.

[5]

C. E. Scott, C. R. Howie, D. MacDonald, and L. C. Biant, “Predicting Dissatisfaction Following Total Knee Replacement: A Prospective Study of 1217 Patients,” Journal of Bone and Joint Surgery 92, no. 9 (2010): 1253-1258, https://doi.org/10.1302/0301-620X.92B9.24394.

[6]

Y. C. Lin, S. H. Lee, C. H. Chang, Y. K. Lin, C. C. Hu, and S. H. Lin, “Isolated Aseptic Loosening in Total Knee Arthroplasty: A Comprehensive 10-Year Review of Partial vs. Total Component Revisions,” BMC Musculoskeletal Disorders 25, no. 1 (2024): 806.

[7]

S. P. Ryan, J. B. Stambough, J. I. Huddleston, and B. R. Levine, “Highlights of the 2023 American Joint Replacement Registry Annual Report,” Arthroplasty Today 26 (2024): 101325, https://doi.org/10.1016/j.artd.2024.101325.

[8]

H. Achakri, Y. Ben-Shlomo, A. Blom, et al., “National Joint Registry Annual Reports,” in The National Joint Registry 20th Annual Report 2023 (National Joint Registry, 2022).

[9]

S. J. Gelderman, H. P. van Jonbergen, L. van Steenbergen, E. Landman, and Y. V. Kleinlugtenbelt, “Patients Undergoing Revisions for Total Knee Replacement Malposition Are Younger and More Often Female: An Analysis of Data From the Dutch Arthroplasty Register,” Journal of Orthopaedics 40 (2023): 70-73.

[10]

K. Thiele, C. Perka, G. Matziolis, H. O. Mayr, M. Sostheim, and R. Hube, “Current Failure Mechanisms After Knee Arthroplasty Have Changed: Polyethylene Wear Is Less Common in Revision Surgery,” Journal of Bone and Joint Surgery 97, no. 9 (2015): 715-720, https://doi.org/10.2106/JBJS.M.01534.

[11]

A. S. Panni, F. Ascione, M. Rossini, et al., “Tibial Internal Rotation Negatively Affects Clinical Outcomes in Total Knee Arthroplasty: A Systematic Review,” Knee Surgery, Sports Traumatology, Arthroscopy 26, no. 6 (2018): 1636-1644.

[12]

Y. Luan, M. Zhang, T. Ran, et al., “Correlation Between Component Alignment and Short-Term Clinical Outcomes After Total Knee Arthroplasty,” Frontiers in Surgery 9 (2022): 991476.

[13]

Z. Li, C. I. Esposito, C. N. Koch, Y. Y. Lee, D. E. Padgett, and T. M. Wright, “Polyethylene Damage Increases With Varus Implant Alignment in Posterior-Stabilized and Constrained Condylar Knee Arthroplasty,” Clinical Orthopaedics and Related Research 475, no. 12 (2017): 2981-2991.

[14]

D. M. Fang, M. A. Ritter, and K. E. Davis, “Coronal Alignment in Total Knee Arthroplasty: Just How Important Is It?,” Journal of Arthroplasty 24, no. 6 Suppl (2009): 39-43.

[15]

Z. W. Wang, L. Wen, Y. C. Luan, et al., “Restoration of Joint Inclination in Total Knee Arthroplasty Offers Little Improvement in Joint Kinematics in Neutrally Aligned Extremities,” Frontiers in Bioengineering and Biotechnology 9 (2021): 673275.

[16]

C. Fang, Y. Luan, Z. Wang, L. Shao, T. Qu, and C. K. Cheng, “Moderate External Rotation of Tibial Component Generates More Natural Kinematics Than Internal Rotation After Total Knee Arthroplasty,” Frontiers in Bioengineering and Biotechnology 10 (2022): 910311.

[17]

D. S. Suh, K. T. Kang, J. Son, O. R. Kwon, C. Baek, and Y. G. Koh, “Computational Study on the Effect of Malalignment of the Tibial Component on the Biomechanics of Total Knee Arthroplasty: A Finite Element Analysis,” Bone & Joint Research 6, no. 11 (2017): 623-630.

[18]

Y. G. Koh, H. T. Hong, H. Y. Lee, H. J. Kim, and K. T. Kang, “Influence of Variation in Sagittal Placement of the Femoral Component After Cruciate-Retaining Total Knee Arthroplasty,” Journal of Knee Surgery 34, no. 4 (2021): 444-451.

[19]

C. J. Betzle, K. E. Bringe, J. V. Horberg, J. T. Moskal, and J. W. Mann, “Effect of Tibial Slope on Coronal Alignment in Total Knee Arthroplasty,” Journal of Knee Surgery 36, no. 4 (2023): 445-449.

[20]

D. P. D'Lima, P. Huang, P. Suryanarayan, A. Rosen, and D. D. D'Lima, “Tibial Tray Rotation and Posterior Slope Increase Risk for Outliers in Coronal Alignment,” Bone & Joint Journal 102-b, no. 6_SA (2020): 43-48.

[21]

Y. Luan, H. Wang, C. Fang, et al., “Mechanical Alignment Tolerance of a Cruciate-Retaining Knee Prosthesis Under Gait Loading-A Finite Element Analysis,” Frontiers in Bioengineering and Biotechnology 11 (2023): 1148914.

[22]

A. Gomoll, T. Wanich, and A. Bellare, “J-Integral Fracture Toughness and Tearing Modulus Measurement of Radiation Cross-Linked UHMWPE,” Journal of Orthopaedic Research 20, no. 6 (2002): 1152-1156.

[23]

C. K. Fitzpatrick, C. Maag, C. W. Clary, A. Metcalfe, J. Langhorn, and P. J. Rullkoetter, “Validation of a New Computational 6-DOF Knee Simulator During Dynamic Activities,” Journal of Biomechanics 49, no. 14 (2016): 3177-3184.

[24]

M. A. Ritter, K. E. Davis, J. B. Meding, J. L. Pierson, M. E. Berend, and R. A. Malinzak, “The Effect of Alignment and BMI on Failure of Total Knee Replacement,” Journal of Bone and Joint Surgery 93, no. 17 (2011): 1588-1596.

[25]

Y. Luan, H. Wang, M. Zhang, et al., “Comparison of Navigation Systems for Total Knee Arthroplasty: A Systematic Review and Meta-Analysis,” Frontiers in Surgery 10 (2023): 1112147.

[26]

R. N. Tandogan, N. P. Kort, E. Ercin, et al., “Computer-Assisted Surgery and Patient-Specific Instrumentation Improve the Accuracy of Tibial Baseplate Rotation in Total Knee Arthroplasty Compared to Conventional Instrumentation: A Systematic Review and Meta-Analysis,” Knee Surgery, Sports Traumatology, Arthroscopy 30, no. 8 (2022): 2654-2665, https://doi.org/10.1007/s00167-021-06495-x.

[27]

K. Lei, L. Liu, X. Chen, Q. Feng, L. Yang, and L. Guo, “Navigation and Robotics Improved Alignment Compared With PSI and Conventional Instrument, While Clinical Outcomes Were Similar in TKA: A Network Meta-Analysis,” Knee Surgery, Sports Traumatology, Arthroscopy 30, no. 2 (2022): 721-733.

[28]

W. Rostoker and J. O. Galante, “Contact Pressure Dependence of Wear Rates of Ultra High Molecular Weight Polyethylene,” Journal of Biomedical Materials Research 13, no. 6 (1979): 957-964.

[29]

R. S. Pascaud, W. T. Evans, P. J. McCullagh, and D. P. FitzPatrick, “Influence of Gamma-Irradiation Sterilization and Temperature on the Fracture Toughness of Ultra-High-Molecular-Weight Polyethylene,” Biomaterials 18, no. 10 (1997): 727-735.

[30]

S. M. Kurtz, C. M. Rimnac, L. Pruitt, C. W. Jewett, V. Goldberg, and A. A. Edidin, “The Relationship Between the Clinical Performance and Large Deformation Mechanical Behavior of Retrieved UHMWPE Tibial Inserts,” Biomaterials 21, no. 3 (2000): 283-291.

[31]

S. M. Kurtz, “2 - From Ethylene Gas to UHMWPE Component: The Process of Producing Orthopedic Implants,” in UHMWPE Biomaterials Handbook, 3rd ed., ed. S. M. Kurtz (William Andrew Publishing, 2016), 7-20.

[32]

O. K. Muratoglu, C. R. Bragdon, M. Jasty, D. O. O'Connor, R. S. Von Knoch, and W. H. Harris, “Knee-Simulator Testing of Conventional and Cross-Linked Polyethylene Tibial Inserts,” Journal of Arthroplasty 19, no. 7 (2004): 887-897.

[33]

J. C. Huot, D. W. Van Citters, J. H. Currier, B. H. Currier, M. B. Mayor, and J. P. Collier, “Evaluating the Suitability of Highly Cross-Linked and Remelted Materials for Use in Posterior Stabilized Knees,” Journal of Biomedical Materials Research Part B, Applied Biomaterials 95, no. 2 (2010): 298-307, https://doi.org/10.1002/jbm.b.31714.

[34]

X. H. Wang, H. Li, X. Dong, F. Zhao, and C. K. Cheng, “Comparison of ISO 14243-1 to ASTM F3141 in Terms of Wearing of Knee Prostheses,” Clinical Biomechanics (Bristol, Avon) 63 (2019): 34-40.

[35]

L. Bauer, M. Kistler, A. Steinbrück, et al., “Different ISO Standards' Wear Kinematic Profiles Change the TKA Inlay Load,” Applied Sciences 11, no. 7 (2021): 3161.

RIGHTS & PERMISSIONS

2025 The Author(s). Orthopaedic Surgery published by Tianjin Hospital and John Wiley & Sons Australia, Ltd.

AI Summary AI Mindmap