Measurement of frequency stability in tunable lasers by using an F-P interferometer
Kuan-lin Mu, Xiu-rong Ma, Shuang-gen Zhang, Shi-yu Zhang, Xia-yang Wang
Optoelectronics Letters ›› 2014, Vol. 10 ›› Issue (1) : 74-76.
Measurement of frequency stability in tunable lasers by using an F-P interferometer
A method for measuring the frequency stability of tunable laser is proposed by using confocal Fabry-Perot (F-P) interferometer. The F-P interferometer is used to get the output frequency of the laser as a reference, and the method eliminates the need of an independent optical source as a frequency reference. Using this technique, the frequency stability of the tunable external-cavity diode laser (ECDL) is measured to be 2.26×10−9 with an integration time of 20 ms.
Frequency Stability / Tunable Laser / Allan Variance / Optoelectronic Oscillator / Quency Stability
[1] |
|
[2] |
|
[3] |
|
[4] |
|
[5] |
|
[6] |
|
[7] |
|
[8] |
|
[9] |
|
[10] |
|
[11] |
|
[12] |
|
[13] |
|
[14] |
|
[15] |
|
This work has been supported by the Tianjin Special Fund Project for Technology Innovation (No.10FDZDGX00400).
/
〈 |
|
〉 |