Measurement of frequency stability in tunable lasers by using an F-P interferometer

Kuan-lin Mu, Xiu-rong Ma, Shuang-gen Zhang, Shi-yu Zhang, Xia-yang Wang

Optoelectronics Letters ›› 2014, Vol. 10 ›› Issue (1) : 74-76.

Optoelectronics Letters ›› 2014, Vol. 10 ›› Issue (1) : 74-76. DOI: 10.1007/s11801-014-3154-3
Article

Measurement of frequency stability in tunable lasers by using an F-P interferometer

Author information +
History +

Abstract

A method for measuring the frequency stability of tunable laser is proposed by using confocal Fabry-Perot (F-P) interferometer. The F-P interferometer is used to get the output frequency of the laser as a reference, and the method eliminates the need of an independent optical source as a frequency reference. Using this technique, the frequency stability of the tunable external-cavity diode laser (ECDL) is measured to be 2.26×10−9 with an integration time of 20 ms.

Keywords

Frequency Stability / Tunable Laser / Allan Variance / Optoelectronic Oscillator / Quency Stability

Cite this article

Download citation ▾
Kuan-lin Mu, Xiu-rong Ma, Shuang-gen Zhang, Shi-yu Zhang, Xia-yang Wang. Measurement of frequency stability in tunable lasers by using an F-P interferometer. Optoelectronics Letters, 2014, 10(1): 74‒76 https://doi.org/10.1007/s11801-014-3154-3

References

[1]
YoungB C, CruzF C, ItanoW M, BergquistJ C. Physical Review Letters, 1999, 82: 3799
CrossRef Google scholar
[2]
SellinP B, StricklandN M, BöttgerT, CarlstenJ L, ConeR L. Physical Review B, 2001, 63: 155111-1
CrossRef Google scholar
[3]
TaralloM G, PoliN, SchioppoM, SutyrinD, TinoG M. Appl. Phys. B, 2011, 103: 17
CrossRef Google scholar
[4]
LiuT, MaX, ZhangS, RenG. Chinese Journal of Lasers, 2012, 39: 0408006-1
[5]
OzdurI, MandridisD, PirachaM U, AkbulutM, HoghooghiN, DelfyettP J. Photonics Technology Letters, 2011, 23: 263
CrossRef Google scholar
[6]
WangW, MaX-R, ChenL, ZhangS-G, ZhaoJ. Chinese Physics Letters, 2013, 29: 100601-1
[7]
TrommsdorffH P, CorvalA, von LaueL. Pure & Appl. Chem., 1995, 67: 191
CrossRef Google scholar
[8]
BöttgerT, PrydeG J, ThielC W, ConeR L. Journal of Luminescence, 2007, 127: 83
CrossRef Google scholar
[9]
ChenL, MaX-R, WangW, ZhangS-G, MuK-L, WangX-Y, ZhangS-Y. Chin. Phys. B, 2013, 22: 06213-1
[10]
SunH, LiY, ChenG. Relations between the Standard Variance and the Allan VarianceInternational Conference on Computational and Information Sciences, 2010, 66
[11]
LiuB, LiangW, SunH, LiJ. International Forum on Strategic Technology, 2011, 7: 390
[12]
DengS-g, JiangY-s, LiZ-d, SangF. Optoelectronics Letters, 2010, 6: 380
CrossRef Google scholar
[13]
WangC-b, ZhangW-g, RuanJ, ShangJ-b, YanA-d. Optoelectronics Letters, 2012, 8: 84
CrossRef Google scholar
[14]
BjorklundG C, LevensonM D. Appl. Phys. B, 1983, 32: 145
CrossRef Google scholar
[15]
BlackE D. Am. J. Phys, 2001, 69: 79
CrossRef Google scholar

This work has been supported by the Tianjin Special Fund Project for Technology Innovation (No.10FDZDGX00400).