Determination of trace selenium in electrolytic manganese by graphite furnace atomic absorption spectrometry

Yao Jun , Zhou Fang-qin , MA Cheng-jin , Tuo Yong , Liu Jian-ben , Wu Zhu-qin , Tan Zhu-zhong

Journal of Wuhan University of Technology Materials Science Edition ›› 2003, Vol. 18 ›› Issue (2) : 32 -34.

PDF
Journal of Wuhan University of Technology Materials Science Edition ›› 2003, Vol. 18 ›› Issue (2) : 32 -34. DOI: 10.1007/BF02838796
Article

Determination of trace selenium in electrolytic manganese by graphite furnace atomic absorption spectrometry

Author information +
History +
PDF

Abstract

The effects of four types of graphite tube and five matrix modifiers on the determination of selenium by graphite furnace atomic absorption spectrometry were compared. The results show that platform thernolysis coat graphite tube and magnesium nitrate and cobaltco as matrix modifer can get a high sensitivity and a good recovery. The optimized working conditions and interference in the determination were invesigated. This result is consistent with that of XRF. The recovery is from 100.8% to 102.2%, the relative standard deviation is from 3. 47% to 5.56% (n=9), and the detection limit of selenium is 378 pg (C=44.5 μ/g to 97.3μg/g.). The proposed method can be applied to the rapid determination of selenium in electrolytic manganese.

Keywords

electrolytic manganese / selenium / matrix modifier / atomic absorption spectrometry

Cite this article

Download citation ▾
Yao Jun, Zhou Fang-qin, MA Cheng-jin, Tuo Yong, Liu Jian-ben, Wu Zhu-qin, Tan Zhu-zhong. Determination of trace selenium in electrolytic manganese by graphite furnace atomic absorption spectrometry. Journal of Wuhan University of Technology Materials Science Edition, 2003, 18(2): 32-34 DOI:10.1007/BF02838796

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Jun Yao, Zhizhong Tian, Liangmeng Chen, Zhufeng Yao, Shang Chen, Zhuzhong Tan. Determination of Impurities in Eletrolytic Manganese by Absorption Spectrophotometry. Spectroscopy and Spectral Analysis, 2001, 21(6): 862-864.

[2]

Jun Yao, Shang Chen, Zuobing Xiao, Yong Tuo. Determination of Trace Iron in Electrolytic Manganese by FAAS with EDTA as Releasing Agent. Journal of Instrumental Analysis, 2001, 20(2): 50-51.

[3]

Yonghong Shu, Huakun He. Direct Determination of Trace Selenium in the Extracts of Orange Skin by Platform Graphite Furnace Atomic Absorption Spectrometry. Analytical laboratory, 1996, 15(5): 73-75.

[4]

Yingen Dong, Huijun Shen. Determination of Trace Selenium in Milk Cow with Graphite Furnace Atomic Absorption Spectrophotometry. Spectroscopy and Spectral Analysis, 2002, 22(4): 691-692.

[5]

Shangcai Feng, Changxiu Han. Direct Determination of Trace Selenium in Human Serum by Zeeman Graphite Furnace Atomic Absorption Spectrometry. Analytical laboratory, 2002, 21(4): 52-53.

[6]

Mu Jiapeng. Handbook of Atomic Absorption, 1989, Beijing: Atomic energy Press. 21-22.

[7]

Li Shuxin. The Interference and Its Elimination in Atomic Absorption Spectrometry, 1987, Beijing: Beijing University Press. 76-82.

[8]

Sahuquillo A, Rauret G, Rehnert A, Muntau H. Solid Sample Graphite Furnace Atomic Absorption Spectroscopy for Supporting Arsenic Determination in Sediments Following a Sequential Extraction Procedure. Analytal Chimica Acta, 2003, 476: 15-24.

[9]

Godlewska Beata, Kiewicz Zy, Zaleska Magdalena. Preconcentration of Palladium in a Flow-through Electrochemical Cell for Determination by Graphite Furnace Atomic Absorption Spectrometry. Analytal Chimica Acta, 2002, 462: 305-312.