Non-invasive prosthesis pose visualization in joint replacement surgery based on multi-sensor fusion

Shaojie Su , Yangsheng Chen , Zhenwei Wei , Zhaofeng Lin

The Journal of China Universities of Posts and Telecommunications ›› 2026, Vol. 33 ›› Issue (2) : 102 -114.

PDF (3660KB)
The Journal of China Universities of Posts and Telecommunications ›› 2026, Vol. 33 ›› Issue (2) :102 -114. DOI: 10.19682/j.cnki.1005-8885.2026.1001
Others
research-article
Non-invasive prosthesis pose visualization in joint replacement surgery based on multi-sensor fusion
Author information +
History +
PDF (3660KB)

Abstract

With the aging population, demand for artificial joint replacement is growing. In total hip arthroplasty (THA), implanting the acetabular cup into the safe zone (inclination: 40° ± 10°, anteversion: 15° ± 10°) is critical for success, but existing technologies suffer from low accuracy or radiation invasiveness. To address this, a non-invasive prosthesis pose visualization method based on multi-sensor fusion is proposed, aiming for real-time, radiation-free and high-precision monitoring. A complete hardware-software system is constructed, consisting of an inertial measurement unit (IMU), a microprocessor, a bluetooth transceiver and a computer. An extended Kalman filter (EKF) is employed for IMU data fusion, and a non-invasive registration method is proposed to establish the coordinate transformation, enabling calculation and visualization of the cup's inclination and anteversion. Experiments showed that the root mean square error (RMSE) of self-pose estimation was within 0.423°, the RMSE of the inclination and anteversion were 0.278° and 0.296° respectively, and the average error vector magnitude was 0.373°, all within the safe zone tolerance, rendering the proposed method promising as an effective surgical tool.

Keywords

total hip arthroplasty (THA) / pose visualization / inertial measurement unit (IMU) / data fusion / non-invasive registration

Cite this article

Download citation ▾
Shaojie Su, Yangsheng Chen, Zhenwei Wei, Zhaofeng Lin. Non-invasive prosthesis pose visualization in joint replacement surgery based on multi-sensor fusion. The Journal of China Universities of Posts and Telecommunications, 2026, 33 (2) : 102-114 DOI:10.19682/j.cnki.1005-8885.2026.1001

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

SIDDIQI A, LEVINE B R, SPRINGER B D . Highlights of the 2021 American joint replacement registry annual report. Arthroplasty Today, 2022, 13: 205-207.

[2]

CIEREMANS D, SHAH A, SLOVER J, et al. Trends in complications and outcomes in patients aged 65 years and younger undergoing total hip arthroplasty: Data from the american joint replacement registry. JAAOS: Global Research and Reviews, 2023, 7(3): 780-785.

[3]

JOINT SURGERY GROUP OF ORTHOPAEDIC SOCIETY OF CHINESE MEDICAL ASSOCIATION. Expert consensus on perioperative management of day-case hip and knee arthroplasty in China. Chinese Journal of Geriatric Orthopaedics and Rehabilitation (Electronic Edition), 2023, 9(6): 321-332. (in Chinese)

[4]

LEWINNEK G E, LEWIS J L, TARR R, et al. Dislocations after total hip-replacement arthroplasties. Journal of Bone & Joint Surgery, 1978, 60(2): 217-220.

[5]

NUTT J L, PAPANIKOLAOU K, KELLETT C F . (ii) Complications of total hip arthroplasty. Orthopaedics and Trauma, 2013, 27(5): 272-276.

[6]

ZHANG L, DENG S H, DONG S N, et al. Unrestricted or low-restriction exercise on the intervention effect of postoperative dislocation after total hip arthroplasty: A meta-analysis. Chinese Journal of Rehabilitation Medicine, 2024, 39(9): 1352-1357. (in Chinese)

[7]

VAN DUREN B H, ROYECA J M, CUNNINGHAM C M, et al. Can the use of an inclinometer improve acetabular cup inclination in total hip arthroplasty? A review of the literature. HIP International, 2021, 31(5): 609-617.

[8]

MAILLOT C, HARMAN C, VILLET L, et al. Modern cup alignment techniques in total hip arthroplasty: A systematic review. Orthopaedics & Traumatology: Surgery & Research, 2019, 105(5): 907-913.

[9]

SPENCER-GARDNER L, PIERREPONT J, TOPHAM M, et al. Patient-specific instrumentation improves the accuracy of acetabular component placement in total hip arthroplasty. Bone & Joint Journal, 2016, 98-B(10): 1342-1346.

[10]

MCARTHUR B A, VULCANO E, CROSS M, et al. Acetabular component orientation in total hip arthroplasty: The impact of obesity. Hip international, 2014, 24(3): 263-269.

[11]

RYAN J A, JAMALI A A, BARGAR W L . Accuracy of computer navigation for acetabular component placement in THA. Clinical Orthopaedics and Related Research, 2010, 468(1): 169-177.

[12]

MIURA T, YAMAMOTO N, SHIROSHITA A, et al. Comparison of implant placement accuracy between manual, robot-assisted, computer-navigated, augmented reality navigated, patient-specific instrumentation, and accelerometer navigated total hip arthroplasty: Asystematic review and network meta-analysis. JBJS Review, 2024, 12(11): Article e24.00120.

[13]

TANG H, ZHOU Y X, MAI B J, et al. Component orientation measurements in total hip arthroplasty using an inertial measurement unit-based smart trial system. Arthroplasty, 2025, 7: Article 36.

[14]

SU S J, ZHOU Y X, WANG Z H, et al. Monocular vision-and IMU-based system for prosthesis pose estimation during total hip replacement surgery. IEEE Transactions on Biomedical Circuits and Systems, 2017, 11(3): 661-670.

[15]

CHEN H, YANG Z C, ZHANG J L, et al. An IMU-based real-time measuring system for acetabular prosthesis implant angles in THR surgeries. IEEE Sensors Journal, 2021, 21(17): 19407-19415.

[16]

MURRAY D W . The definition and measurement of acetabular orientation. The Bone & Joint Journal, 1993, 75(2): 228-232.

[17]

CORKE P. Representing position and orientation. Robotics, Vision and Control. Springer Tracts in Advanced Robotics, Vol 146. Cham, Switzerland: Springer Nature SwitzerlandSpringer, 2023: 31-48.

[18]

MARKLEY F L, CRASSIDIS J L . Euler angles. Fundamentals of Spacecraft Attitude Determination and Control. Space Technology Library, Vol 33. New York, NY, USA: Springer, 2014.

[19]

SABATINI A M. Quaternion-based extended Kalman filter for determining orientation by inertial and magnetic sensing. IEEE Transactions on Bio-Medical Engineering, 2006, 53(7): 1346-1356.

PDF (3660KB)

4

Accesses

0

Citation

Detail

Sections
Recommended

/

〈 〉