Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise
Author information +
History +
PDF (5245KB)

Abstract

Aim: Structured light plethysmography (SLP), is a contactless optical system developed to monitor breathing patterns by analyzing chest-wall movement. It has not been thoroughly validated against other non-invasive motion analysis systems under different breathing conditions. This study therefore aimed to evaluate the criterion-validity of the SLP compared to the respiratory inductive plethysmography (RIP) at rest and after exercise.

Methods: Adults underwent two simultaneous 5-minute recordings from both devices, conducted at rest and following submaximal exercise on a cycle ergometer. Timing indices and thoracoabdominal (TA) movement parameters were examined. Measurement agreement between SLP and RIP was assessed using Bland-Altman plots at rest, after exercise, and for exercise-induced changes.

Results: Fifty adults (mean age 29.3 ± 6.8 years; 30 males) participated. Α total of 3,395 and 4,295 breath cycles were analyzed at rest and post-exercise, respectively. Over 92% of differences in timing parameters under both conditions were within the 95% limits of agreement (LOA) and their mean differences were found close to zero across a wide range of breath cycle magnitudes (rest: 2.62–8.06 s; post-exercise: 2.16–6.16 s). For ΤΑ movement parameters, the mean bias between devices at rest was 0.31 for ribcage amplitude (RCampi) and 0.23 for abdominal amplitude (ABampi), with LOA ranging from −0.06 to 0.66 and −0.06 to 0.52, respectively. A trend towards greater discrepancies for the individual measurements of RCampi and ABampi at higher magnitudes of TA movements was noted, especially post-exercise. A good average agreement between the devices was found for RCampi/ABampi both at rest [mean difference: 0.03, standard deviation (SD): 0.21] and after exercise (mean difference: 1.10, SD: 0.24).

Conclusions: The SLP is an accurate method to quantify and measure timing indices and the ratio of the ribcage motion to the abdominal motion under different breathing conditions.

Keywords

Breathing pattern assessment / digital monitoring method / structured light plethysmography / respiratory inductive plethysmography

Cite this article

Download citation ▾
Panagiotis Sakkatos. Evaluation of a digital fully-contactless optical device to quantify and record breathing pattern components at rest and after exercise. Exploration of Digital Health Technologies, 2025, 3 (1) : 101156 DOI:10.37349/edht.2025.101156

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Krohn F, Novello M, van der Giessen RS, De Zeeuw CI, Pel JJM, Bosman LWJ. The integrated brain network that controls respiration. Elife. 2023; 12:e83654.

[2]

Boulding R, Stacey R, Niven R, Fowler SJ. Dysfunctional breathing: a review of the literature and proposal for classification. Eur Respir Rev. 2016; 25:287-94.

[3]

Sakkatos P, Bruton A, Barney A. Changes in quantifiable breathing pattern components predict asthma control: an observational cross-sectional study. Asthma Res Pract. 2021; 7:5.

[4]

Fiamma MN, Samara Z, Baconnier P, Similowski T, Straus C. Respiratory inductive plethysmography to assess respiratory variability and complexity in humans. Respir Physiol Neurobiol. 2007; 156:234-9.

[5]

Massaroni C, Carraro E, Vianello A, Miccinilli S, Morrone M, Levai IK, et al. Optoelectronic Plethysmography in Clinical Practice and Research: A Review. Respiration. 2017; 93:339-54.

[6]

Houssein A, Ge D, Gastinger S, Dumond R, Prioux J. Estimation of respiratory variables from thoracoabdominal breathing distance: a review of different techniques and calibration methods. Physiol Meas. 2019; 40:03TR01.

[7]

Levai IK, Sidoroff V, Iles R. An introduction to the non-invasive noncontact assessment of respiratory function. Respir Ther. 2012; 7:42-4.

[8]

Motamedi-Fakhr S, Iles R, Barney A, de Boer W, Conlon J, Khalid A, et al. Evaluation of the agreement of tidal breathing parameters measured simultaneously using pneumotachography and structured light plethysmography. Physiol Rep. 2017; 5:e13124.

[9]

Niérat MC, Dubé BP, Llontop C, Bellocq A, Layachi Ben Mohamed L, Rivals I, et al. Measuring Ventilatory Activity with Structured Light Plethysmography (SLP) Reduces Instrumental Observer Effect and Preserves Tidal Breathing Variability in Healthy and COPD. Front Physiol. 2017; 8:316.

[10]

Alhuthail E, Stockley J, Coney A, Cooper B. Measurement of breathing in patients with post-COVID-19 using structured light plethysmography (SLP). BMJ Open Respir Res. 2021; 8:e001070.

[11]

Motamedi-Fakhr S, Wilson RC, Iles R. Tidal breathing patterns derived from structured light plethysmography in COPD patients compared with healthy subjects. Med Devices (Auckl). 2016; 10:1-9.

[12]

Fleck D, Curry C, Donnan K, Logue O, Graham K, Jackson K, et al. Investigating the clinical use of structured light plethysmography to assess lung function in children with neuromuscular disorders. PLoS One. 2019; 14:e0221207.

[13]

Elshafie G, Kumar P, Motamedi-Fakhr S, Iles R, Wilson RC, Naidu B. Measuring changes in chest wall motion after lung resection using structured light plethysmography: a feasibility study. Interact Cardiovasc Thorac Surg. 2016; 23:544-7.

[14]

Lauhkonen E, Cooper BG, Iles R. Mini review shows that structured light plethysmography provides a non-contact method for evaluating breathing patterns in children. Acta Paediatr. 2019; 108:1398-405.

[15]

Hmeidi H, Motamedi-Fakhr S, Chadwick E, Gilchrist FJ, Lenney W, Iles R, et al. Tidal breathing parameters measured using structured light plethysmography in healthy children and those with asthma before and after bronchodilator. Physiol Rep. 2017; 5:e13168.

[16]

Clarenbach CF, Senn O, Brack T, Kohler M, Bloch KE. Monitoring of ventilation during exercise by a portable respiratory inductive plethysmograph. Chest. 2005; 128:1282-90.

[17]

Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al. Standardisation of spirometry. Eur Respir J. 2005; 26:319-38.

[18]

American Thoracic Society; American College of Chest Physicians. ATS/ACCP Statement on cardiopulmonary exercise testing. Am J Respir Crit Care Med. 2003; 167:211-77.

[19]

Romei M, Mauro AL, D’Angelo MG, Turconi AC, Bresolin N, Pedotti A, et al. Effects of gender and posture on thoraco-abdominal kinematics during quiet breathing in healthy adults. Respir Physiol Neurobiol. 2010; 172:184-91.

[20]

Sakkatos P. Recording and analysis of breathing patterns: a physiological marker in asthma management [dissertation]. Southampton (UK):University of Southampton; 2019.

[21]

Hollier CA, Harmer AR, Maxwell LJ, Menadue C, Willson GN, Black DA, et al. Validation of respiratory inductive plethysmography (LifeShirt) in obesity hypoventilation syndrome. Respir Physiol Neurobiol. 2014; 194:15-22.

[22]

De Boer W, Lasenby J, Cameron J, Wareham R, Ahmad S, Roach C, et al. SLP: a zero-contact non-invasive method for pulmonary function testing. BMVC Press. 2010:1-12.

[23]

Tehrany R, DeVos R, MCSP, Bruton A. Breathing pattern recordings using respiratory inductive plethysmography, before and after a physiotherapy breathing retraining program for asthma: A case report. Physiother Theory Pract. 2018; 34:329-35.

[24]

Wilcox RR. New methods for comparing groups: strategies for increasing the probability of detecting true differences. Curr Dir Psychol Sci. 2005; 14:272-5.

[25]

Krouwer JS. Why Bland-Altman plots should use X, not (Y+X)/2 when X is a reference method. Stat Med. 2008; 27:778-80.

[26]

Smith I, Mackay J, Fahrid N, Krucheck D. Respiratory rate measurement: a comparison of methods. Br J Healthc Assist. 2011; 5:125-31.

[27]

De Troyer A, Boriek AM. Mechanics of the respiratory muscles. Compr Physiol. 2011; 1:1273-300.

[28]

Motamedi-Fakhr S, Iles R, Barker N, Alexander J, Cooper BG. Reference equations for tidal breathing parameters using structured light plethysmography. ERJ Open Res. 2021; 7:00050-2021.

PDF (5245KB)

0

Accesses

0

Citation

Detail

Sections
Recommended

/

〈 〉