Intracardiac echocardiography-guided transseptal puncture in anteroinferior native septum site for patients undergoing atrial fibrillation ablation with atrial septal closure devices: A case series

Evan Jim Gunawan , Hui Chai , Zhanyang Wei , Le Li , Zhicheng Hu , LiKun Zhou , Yu Xia , Xu Meng , Lingmin Wu , Lihui Zheng , Ligang Ding , Yan Yao

Brain & Heart ›› 2025, Vol. 3 ›› Issue (3) : 5119

PDF (842KB)
Brain & Heart ›› 2025, Vol. 3 ›› Issue (3) :5119 DOI: 10.36922/bh.5119
CASE SERIES
research-article
Intracardiac echocardiography-guided transseptal puncture in anteroinferior native septum site for patients undergoing atrial fibrillation ablation with atrial septal closure devices: A case series
Author information +
History +
PDF (842KB)

Abstract

This case series describes the safety and feasibility of an anteroinferior transseptal puncture (TSP) in patients implanted with atrial septal defect (ASD) closure devices for atrial fibrillation (AF) ablation. Eight AF ablation procedures were performed in seven patients (58.4 ± 8.7 years; two males) with symptomatic AF and percutaneous ASD closure. Typically, the TSP site for patients with an ASD closure device is located in the inferior-posterior region of the native septum. However, with larger devices (>26 mm), limited space in this area can pose challenges for transseptal access. Using transthoracic echocardiography and 3D cardiac computed tomography, TSP sites were pre-determined, targeting the anteroinferior native septum. Closure device diameters averaged 23.8 ± 10.8 mm (range: 10 - 40 mm), with five cases involving devices ≥26 mm. The procedures were guided by fluoroscopy and intracardiac echocardiography, achieving a 100% success rate without TSP-related complications. These findings suggest the anteroinferior TSP approach is a safe and viable alternative for patients with large ASD closure devices.

Keywords

Atrial septal defect / Atrial septal closure device / Transseptal puncture / Atrial fibrillation / Ablation

Cite this article

Download citation ▾
Evan Jim Gunawan, Hui Chai, Zhanyang Wei, Le Li, Zhicheng Hu, LiKun Zhou, Yu Xia, Xu Meng, Lingmin Wu, Lihui Zheng, Ligang Ding, Yan Yao. Intracardiac echocardiography-guided transseptal puncture in anteroinferior native septum site for patients undergoing atrial fibrillation ablation with atrial septal closure devices: A case series. Brain & Heart, 2025, 3(3): 5119 DOI:10.36922/bh.5119

登录浏览全文

4963

注册一个新账户 忘记密码

Acknowledgments

None.

Funding

This study was supported by the National High Level Hospital Clinical Research Funding (2023-GSP-GG-22) and the Clinical and Translational Medicine Research Project, Chinese Academy of Medical Sciences (2023- LC05).

Conflict of interest

Yan Yao is the Editor-in-Chief of this journal but was not in any way involved in the editorial and peer-review process conducted for this paper, directly or indirectly. Separately, other authors declared that they have no known competing financial interests or personal relationships that could have influenced the work reported in this paper.

Author contributions

Conceptualization: Evan Jim Gunawan, Lihui Zheng, Ligang Ding, Yan Yao

Investigation: Evan Jim Gunawan, Hui Chai, Zhanyang Wei, Ligang Ding

Writing - original draft: Evan Jim Gunawan, Hui Chai, Le Li, Zhicheng Hu, Likun Zhou, Ligang Ding

Writing - review & editing: Evan Jim Gunawan, Yu Xia, Xu Meng, Lingmin Wu, Ligang Ding

Ethics approval and consent to participate

This study was approved by the Ethics Committee of Fuwai Hospital, Chinese Academy of Medical Science (Approval ID: 2022-1705). Written informed consent was obtained from all participating patients before their inclusion in the study.

Consent for publication

Written informed consent was obtained from all patients for the release of their data and/or images included in this publication.

Availability of data

Not applicable.

References

[1]

Spies C, Khandelwal A, Timmermanns I, Schräder R. Incidence of atrial fibrillation following transcatheter closure of atrial septal defects in adults. Am J Cardiol. 2008; 102(7):902-906. doi: 10.1016/j.amjcard.2008.05.045

[2]

Santangeli P, Di Biase L, Burkhardt JD, et al. Transseptal access and atrial fibrillation ablation guided by intracardiac echocardiography in patients with atrial septal closure devices. Heart Rhythm. 2011; 8(11):1669-1675. doi: 10.1016/j.hrthm.2011.06.023

[3]

Saluveer O, Bastani H, Verouhis D, Källner G, Jensen- Urstad M. Transseptal access through an atrial septal defect closure device resulting in open heart surgery. JACC Case Rep. 2022; 4(11):685-687. doi: 10.1016/j.jaccas.2022.04.006

[4]

Li X, Wissner E, Kamioka M, et al. Safety and feasibility of transseptal puncture for atrial fibrillation ablation in patients with atrial septal defect closure devices. Heart Rhythm. 2014; 11(2):330-335. doi: 10.1016/j.hrthm.2013.11.011

[5]

Tzeis S, Gerstenfeld EP, Kalman J, et al. 2024 European Heart Rhythm Association/Heart Rhythm Society/Asia Pacific Heart Rhythm Society/Latin American Heart Rhythm Society expert consensus statement on catheter and surgical ablation of atrial fibrillation. Europace. 2024; 26(4):euae043. doi: 10.1093/europace/euae043

[6]

Lee T, Tsai IC, Fu YC, et al. MDCT evaluation after closure of atrial septal defect with an amplatzer septal occluder. AJR Am J Roentgenol. 2007; 188(5):W431-W439. doi: 10.2214/AJR.06.0709

[7]

Himelfarb JD, Shulman H, Olesovsky CJ, et al. Atrial fibrillation following transcatheter atrial septal defect closure: A systematic review and meta-analysis. Heart. 2022; 108(15):1216-1224. doi: 10.1136/heartjnl-2021-319794

[8]

Prolic Kalinsek T, Rauber M, Podlogar V, et al. Feasibility, efficacy, and safety of intracardiac echocardiography guided transseptal puncture: A single-center retrospective study. Ep Europace. 2023; 25(Supplement_1):euad122.768. doi: 10.1093/europace/euad122.768

[9]

Guo Q, Sang C, Bai R, et al. Transseptal puncture in patients with septal occluder devices during catheter ablation of atrial fibrillation. EuroIntervention. 2022; 17(13):1112-1119. doi: 10.4244/EIJ-D-21-00340

[10]

Sang CH, Dong JZ, Long DY, et al. Transseptal puncture and catheter ablation of atrial fibrillation in patients with atrial septal occluder: Initial experience of a single centre. Europace. 2018; 20(9):1468-1474. doi: 10.1093/europace/eux282

[11]

Garg J, Shah K, Turagam MK, Janagam P, Natale A, Lakkireddy D. Safety and efficacy of catheter ablation for atrial fibrillation in patients with percutaneous atrial septal closure device: Electrophysiology Collaborative Consortium for Meta-analysis-ELECTRAM investigators. J Cardiovasc Electrophysiol. 2020; 31(9):2328-2334. doi: 10.1111/jce.14616

[12]

Sabiniewicz R, Królak T, Wozniak L, Budrejko S, Raczak G. Transesophageal echocardiography guided transseptal puncture for atrial fibrillation ablation in a patient with a 30 mm atrial septal closure device. Postepy Kardiol Interwencyjnej. 2016; 12(3):285-286. doi: 10.5114/aic.2016.61658

[13]

Hu J, Ding L, Gunawan E, Lai H, Yao Y. Successful treatment of atrial flutter post-radiofrequency ablation for atrial fibrillation following atrial septal defect occlusion: A case report of pulsed field ablation. Eur Heart J Case Rep. 2024;8:ytae558. doi: 10.1093/ehjcr/ytae558

PDF (842KB)

18

Accesses

0

Citation

Detail

Sections
Recommended

/