Subtomogram averaging by cryo electron tomography using CRYO ARM™ 300 II for purified and cellular samples
Fumiaki Makino , Haruaki Yanagisawa , Fabian Eisenstein , Tomoko Miyata , Miki Kinoshita , Masahide Kikkawa , Keiichi Namba
AIMS Biophysics ›› 2026, Vol. 13 ›› Issue (2) : 131 -142.
Cryo electron tomography (cryo-ET) enables three-dimensional visualization of biological macromolecules in their near-native environments, bridging the gap between structural and cellular biology. In this study, we evaluated high-resolution subtomogram averaging (STA) using a cryo-TEM (CRYO ARMTM 300 II, JEOL) and PACEtomo software system for both single-particle analysis (SPA) type grids and lamellae prepared by cryo-focused ion beam (cryo-FIB) milling (JIB-4700F, JEOL). Using mouse apoferritin and E. coli ribosomes in thin vitreous ice film, resolutions of 2.04 Å and 2.86 Å were achieved, respectively. Using cryo-FIB lamellae of Salmonella cells, in situ ribosome structures were resolved to 8.2 Å resolution.
cryo-EM / cryo-ET / STA / cryo-FIB-SEM / CryoLameller / CRYO ARM / structural cell biology
| [1] |
Schur FKM (2019) Toward high-resolution in situ structural biology with cryo-electron tomography and subtomogram averaging. Curr Opin Struct Biol 58: 1-9. https://doi.org/10.1016/j.sbi.2019.03.018 |
| [2] |
|
| [3] |
|
| [4] |
|
| [5] |
|
| [6] |
|
| [7] |
|
| [8] |
|
| [9] |
|
| [10] |
|
| [11] |
|
| [12] |
|
| [13] |
|
| [14] |
|
| [15] |
Chreifi G, Chen S, |
| [16] |
|
| [17] |
|
| [18] |
|
/
| 〈 |
|
〉 |