Dosimetric evaluation of ABS, PLA and NR boluses for electron radiotherapy
Huda Haddad
AIMS Biophysics ›› 2026, Vol. 13 ›› Issue (1) : 1 -20.
Purpose. This study aims to compare the dosimetric characteristics of 3D-printed bolus materials acrylonitrile butadiene styrene (ABS), polylactic acid (PLA) with recently introduced natural rubber (NR) bolus. Materials and Methods. We employed Monte Carlo simulation to evaluate ABS, PLA, and NR boluses of thicknesses 0.5, 1.0, and 1.5 cm under 6, 9, and 16 MeV electron beam irradiation. Percentage depth dose (PDD) data was analyzed to evaluate dosimetry parameters. Dosimetric stability under varying air-gap conditions was assessed by analyzing PDD curves under air-gap sizes of (0, 1, 3, 5 mm).Results.NR and ABS showed similar dosimetric profiles, whereas PLA showed enhanced deeper-tissue protection and provided 0.7–10.2% higher surface dose (SD). PLA also exhibited the highest stability, with SD and R90 deviations limited to 1% and 1.2%, respectively, under varying air-gap sizes. At 6 and 9 MeV, a 0.5 cm bolus failed to provide a single dose of 90% of maximum dose for nearly all tested materials Conclusion. NR demonstrated comparable dosimetric performance to ABS, serving as a viable substitute. PLA was optimal for maximizing SD and distal tissue sparing while exhibiting the lowest air-gap sensitivity. For tumors located several centimeters deep from the surface, 16 MeV electron beams were suitable. Furthermore, Dual hotspots were identified a substantial risk of localized normal tissue toxicity.
ABS / bolus / electron bolus / electron radiotherapy / Monte Carlo simulation / natural rubber / PLA
| [1] |
|
| [2] |
|
| [3] |
|
| [4] |
|
| [5] |
|
| [6] |
|
| [7] |
|
| [8] |
|
| [9] |
|
| [10] |
|
| [11] |
|
| [12] |
|
| [13] |
|
| [14] |
|
| [15] |
|
| [16] |
|
| [17] |
|
| [18] |
|
| [19] |
|
| [20] |
|
| [21] |
National Institute of Standard and Technology. Stopping Power and Range Tables for Electrons, Nist. Gov. Available from: https://physics.nist.gov/PhysRefData/Star/Text/ESTAR.html. |
| [22] |
|
| [23] |
|
| [24] |
|
| [25] |
Rogers DWO, Walters B, Kawrakow I (2009) BEAMnrc user’s manual, National Research Council Canada. Institute for National Measurement Standards, PIRS-0509 (A) (revL), Ottawa, Canda. |
| [26] |
|
| [27] |
International Atomic Energy Agency, phase-space database for external beam radiotherapy. Available from: https://www-nds.iaea.org/phsp/phsp.htmlx. |
| [28] |
International Atomic Energy Agency (2000) Absorbed dose determination in external beam radiotherapy: an International Code of Practice for dosimetry based on standards of absorbed dose to water. TRS 398. IAEA, Vienna. |
| [29] |
|
| [30] |
|
| [31] |
|
| [32] |
|
| [33] |
|
| [34] |
Berger MJ, Hubbell JH, Seltzer SM, |
| [35] |
|
| [36] |
Singh S, |
| [37] |
|
| [38] |
|
| [39] |
|