In vitro evaluation of artificial joints: a comprehensive review

Gang Shen , Ju-Fan Zhang , Feng-Zhou Fang

Advances in Manufacturing ›› 2019, Vol. 7 ›› Issue (1) : 1 -14.

PDF
Advances in Manufacturing ›› 2019, Vol. 7 ›› Issue (1) : 1 -14. DOI: 10.1007/s40436-018-00244-z
Article

In vitro evaluation of artificial joints: a comprehensive review

Author information +
History +
PDF

Abstract

Development of artificial joints is a great progress for joint replacement operations of human being, but the short longevity of prostheses has concerned both industries and researchers since the advent of modern implants. Thus, continuous improvements have been made alongside the clinical applications. The methodologies to inspect the potential properties of new designs have also seen many advances. The test machines have evolved from easy pin-on-disk configuration to the modern joint simulator. Besides, various wear traces provided by testing machines are investigated greatly, and multi-directional files are recommended as motion profiles for in vitro testing. The typical testing parameters, like sliding speed or loading profile, are discussed in this article, and their working mechanisms are described in detail. Meanwhile, the calf or bovine serum has been regarded as the gold standard for testing lubricant, and the future trend of fluid will be focused on investigating the effects of lubricant composites on the tribological properties. Hard particles and proteins are discussed as well, and their effects on the wear mechanisms are the focus. Finally, various approaches to measure the wear rate, wear factor and the thickness of lubricant film are presented, and suggestions are given for the development in future.

Keywords

Artificial joint / Methodology / Testing parameter / Wear rate

Cite this article

Download citation ▾
Gang Shen, Ju-Fan Zhang, Feng-Zhou Fang. In vitro evaluation of artificial joints: a comprehensive review. Advances in Manufacturing, 2019, 7(1): 1-14 DOI:10.1007/s40436-018-00244-z

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Kang CW, Fang FZ. State of the art of bioimplants manufacturing: part I. Adv Manuf, 2018, 6(1): 20-40.

[2]

Kang CW, Fang FZ. State of the art of bioimplants manufacturing: part II. Adv Manuf, 2018, 6(2): 137-154.

[3]

Kurtz S, Ong K, Lau E, et al. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am, 2007, 89(4): 780-785.

[4]

Rahaman MN, Yao A, Bal BS, et al. Ceramics for prosthetic hip and knee joint replacement. J Am Ceram Soc, 2007, 90(7): 1965-1988.

[5]

Ching HA, Choudhury D, Nine MJ, et al. Effects of surface coating on reducing friction and wear of orthopaedic implants. Sci Technol Adv Mater, 2014, 15(1): 014402

[6]

Fang FZ. Atomic and close-to-atomic scale manufacturing—a trend in manufacturing development. Front Mech Eng, 2016, 11(4): 325-327.

[7]

Fang FZ. Manufacturing III (Invited lecture), 2014, Dublin: University College Dublin

[8]

Yamane S, Moro T, Kyomoto M, et al. Wear resistance of vitamin E-blended polyethylene tibial insert by MPC polymer grafting. Bone Joint J, 2017, 99(s 6): 104

[9]

Choudhury D, Rebenda D, Sasaki S, et al. Enhanced lubricant film formation through micro-dimpled hard-on-hard artificial hip joint: An in-situ observation of dimple shape effects. J Mech Behav Biomed Mater, 2018, 81: 120-129.

[10]

Koseki H, Tomita M, Yonekura A, et al. Effect of carbon ion implantation on the tribology of metal-on-metal bearings for artificial joints. Int J Nanomed, 2017, 12: 4111-4116.

[11]

Chen Y, Li Y, Kurosu S, et al. Analysis of run-in-stage wear behavior and contact mechanics of metal-on-metal hip joint bearings with different radial clearances. Mater Trans, 2015, 56(6): 826-834.

[12]

Affatato S. Wear of orthopaedic implants and artificial joints, 2012, Amsterdam: Elsevier

[13]

Bowsher J, Shelton J. A hip simulator study of the influence of patient activity level on the wear of crosslinked polyethylene under smooth and roughened femoral conditions. Wear, 2001, 250(1–12): 167-179.

[14]

Chen Y, Li Y, Koizumi Y, et al. Effects of carbon addition on wear mechanisms of CoCrMo metal-on-metal hip joint bearings. Mater Sci Eng C, 2017, 76: 997-1004.

[15]

Wilches L, Uribe J, Toro A. Wear of materials used for artificial joints in total hip replacements. Wear, 2008, 265(1–2): 143-149.

[16]

Ghosh S, Choudhury D, Das NS, et al. Tribological role of synovial fluid compositions on artificial joints—a systematic review of the last 10 years. Lubr Sci, 2014, 26(6): 387-410.

[17]

Fisher J, Dowson D, Hamdzah H, et al. The effect of sliding velocity on the friction and wear of UHMWPE for use in total artificial joints. Wear, 1994, 175(1–2): 219-225.

[18]

Bowsher J, Shelton J (2000) The influence of stumbling on the wear of ultra-high molecular weight polyethylene. In: Proceedings of the 6th world biomaterials congress transactions, 2000, p 867

[19]

Bigsby R, Hardaker C, Fisher J. Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant. Proc Inst Mech Eng Part H J Eng Med, 1997, 211(3): 265-269.

[20]

Kurtz SM. UHMWPE biomaterials handbook: Ultra high molecular weight polyethylene in total joint replacement and medical devices, 2009, London: Academic Press

[21]

Baykal D, Siskey R, Haider H, et al. Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers. J Mech Behav Biomed Mater, 2014, 31: 117-134.

[22]

McKellop H, Clarke I, Markolf K, et al. Wear characteristics of UHMW polyethylene: A method for accurately measuring extremely low wear rates. J Biomed Mater Res Part A, 1978, 12(6): 895-927.

[23]

Shen G, Fang FZ, Kang CW. Tribological performance of bioimplants: A comprehensive review. Nanotechnol Precis Eng, 2018, 1(2): 107-122.

[24]

Archard J. Contact and rubbing of flat surfaces. J Appl Phys, 1953, 24(8): 981-988.

[25]

Wang A. A unified theory of wear for ultra-high molecular weight polyethylene in multi-directional sliding. Wear, 2001, 248(1–2): 38-47.

[26]

Wang A, Polineni V, Essner A, et al. The significance of nonlinear motion in the wear screening of orthopaedic implant materials. J Test Eval, 1997, 25(2): 239-245.

[27]

Saikko V, Ahlroos T. Type of motion and lubricant in wear simulation of polyethylene acetabular cup. Proc Inst Mech Eng Part H J Eng Med, 1999, 213(4): 301-310.

[28]

Pooley CM, Tabor D. Friction and molecular structure: The behaviour of some thermoplastics. Proc R Soc Lond A, 1972, 329(1578): 251-274.

[29]

Bragdon CR, O’Connor DO, Lowenstein JD, et al. A new pin-on-disk wear testing method for simulating wear of polyethylene on cobalt-chrome alloy in total hip arthroplasty. J Arthroplasty, 2001, 16(5): 658-665.

[30]

Turell M, Friedlaender G, Wang A, et al. The effect of counterface roughness on the wear of UHMWPE for rectangular wear paths. Wear, 2005, 259(7–12): 984-991.

[31]

Turell M, Wang A, Bellare A. Quantification of the effect of cross-path motion on the wear rate of ultra-high molecular weight polyethylene. Wear, 2003, 255(7–12): 1034-1039.

[32]

Bennett D, Humphreys L, O’brien S, et al. The influence of wear paths produced by hip replacement patients during normal walking on wear rates. J Orthop Res, 2008, 26(9): 1210-1217.

[33]

Davey S, Orr J, Buchanan F, et al. Measurement of molecular orientation in retrieved ultra-high-molecular-weight polyethylene (UHMWPE) hip sockets using Fourier-transform infrared spectroscopy. Strain, 2004, 40(4): 203-210.

[34]

Saikko V, Calonius O, Keränen J. Effect of slide track shape on the wear of ultra-high molecular weight polyethylene in a pin-on-disk wear simulation of total hip prosthesis. J Biomed Mater Res B Appl Biomater, 2004, 69(2): 141-148.

[35]

Korduba L, Wang A. The effect of cross-shear on the wear of virgin and highly-crosslinked polyethylene. Wear, 2011, 271(9–10): 1220-1223.

[36]

Firkins P, Tipper J, Saadatzadeh M, et al. Quantitative analysis of wear and wear debris from metal-on-metal hip prostheses tested in a physiological hip joint simulator. Bio-Med Mater Eng, 2001, 11(2): 143-157.

[37]

Tipper J, Firkins P, Besong A, et al. Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator. Wear, 2001, 250(1–12): 120-128.

[38]

Ali M, Al-Hajjar M, Partridge S, et al. Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings. Proc Inst Mech Eng Part H J Eng Med, 2016, 230(5): 389-397.

[39]

Moro T, Takatori Y, Kyomoto M, et al. Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer. J Orthop Res, 2014, 32(3): 369-376.

[40]

Herrmann S, Kaehler M, Grawe R et al (2015) Physiological-like testing of the dislocation stability of artificial hip joints. In: Wenger P, Flores P (eds) New trends in mechanism and machine science. Springer, Berlin, pp 659–667

[41]

Kawanabe K, Clarke IC, Tamura J, et al. Effects of A–P translation and rotation on the wear of UHMWPE in a total knee joint simulator. J Biomed Mater Res Part A, 2001, 54(3): 400-406.

[42]

Dowson D, Gillis B, Atkinson J. Penetration of metallic femoral components into polymeric tibial components observed in a knee joint simulator, 1985, Washington: ACS Publications

[43]

Tipper J, Galvin A, Williams S, et al. Isolation and characterization of UHMWPE wear particles down to ten nanometers in size from in vitro hip and knee joint simulators. J Biomed Mater Res Part A, 2006, 78(3): 473-480.

[44]

McKellop HA, D’lima D. How have wear testing and joint simulator studies helped to discriminate among materials and designs?. JAAOS J Am Acad Orthop Surg, 2008, 16: S111-S119.

[45]

Affatato S, Ruggiero A, Merola M. Advanced biomaterials in hip joint arthroplasty. A review on polymer and ceramics composites as alternative bearings. Composites Part B Engineering, 2015, 83: 276-283.

[46]

Saikko V. Effect of contact area on the wear of ultrahigh molecular weight polyethylene in noncyclic pin-on-disk tests. Tribol Int, 2017, 114: 84-87.

[47]

Zhang T, Deng Q, Liu B, et al. Wear and corrosion properties of diamond like carbon (DLC) coating on stainless steel, CoCrMo and Ti6Al4V substrates. Surf Coat Technol, 2015, 273: 12-19.

[48]

Nečas D, Vrbka M, Urban F, et al. The effect of lubricant constituents on lubrication mechanisms in hip joint replacements. J Mech Behav Biomed Mater, 2016, 55: 295-307.

[49]

Deng Y, Xiong D. Fabrication and properties of UHMWPE grafted with acrylamide polymer brushes. J Polym Res, 2015, 22(10): 195

[50]

Krismer M. Sports activities after total hip arthroplasty. EFORT Open Rev, 2017, 2(5): 189-194.

[51]

Zhang G, Zhang C, Nardin P, et al. Effects of sliding velocity and applied load on the tribological mechanism of amorphous poly-ether-ether-ketone (PEEK). Tribol Int, 2008, 41(2): 79-86.

[52]

Manhabosco TM, Barboza APM, Batista RJC, et al. Corrosion, wear and wear-corrosion behavior of graphite-like a-C: H films deposited on bare and nitrided titanium alloy. Diam Relat Mater, 2013, 31: 58-64.

[53]

Gispert M, Serro A, Colaco R, et al. Friction and wear mechanisms in hip prosthesis: comparison of joint materials behaviour in several lubricants. Wear, 2006, 260(1–2): 149-158.

[54]

Kahyaoglu O, Unal H. Friction and wear behaviours of medical grade UHMWPE at dry and lubricated conditions. Int J Phys Sci, 2012, 7(16): 2478-2485.

[55]

Dressler MR, Strickland MA, Taylor M, et al. Predicting wear of UHMWPE: decreasing wear rate following a change in direction. Wear, 2011, 271(11–12): 2879-2883.

[56]

Welghtman B, Light D. The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW polyethylene. Biomaterials, 1986, 7(1): 20-24.

[57]

Streicher R, Weber H, Schoen R, et al. Wear resistant couplings for longer lasting articulating total joint replacements. Adv Biomater, 1993, 10: 179-186.

[58]

Liew KW, Kok CK, Efzan ME. Effect of EDM dimple geometry on friction reduction under boundary and mixed lubrication. Tribol Int, 2016, 101: 1-9.

[59]

Dresel W. Lubricants and lubrication, 2007, New York: Wiley

[60]

Houdková Š, Šperka P, Repka M, et al. Shifted laser surface texturing for bearings applications. J Phys Conf Ser, 2017, 843(1): 012076

[61]

Czichos H, Habig K. Tribologie Handbuch: Systemanalyse, Prüftechnik, Werkstoffe und Konstruktionselemente, 1992, Wiesbaden: Vieweg Verlag

[62]

Nečas D, Sawae Y, Fujisawa T, et al. The influence of proteins and speed on friction and adsorption of metal/UHMWPE contact pair. Biotribology, 2017, 11: 51-59.

[63]

Saikko V. Effect of lubricant protein concentration on the wear of ultra-high molecular weight polyethylene sliding against a CoCr counterface. J Tribol, 2003, 125(3): 638-642.

[64]

Lu Z, McKellop H. Frictional heating of bearing materials tested in a hip joint wear simulator. Proc Inst Mech Eng Part H J Eng Med, 1997, 211(1): 101-108.

[65]

Scholes S, Unsworth A. The effects of proteins on the friction and lubrication of artificial joints. Proc Inst Mech Eng Part H J Eng Med, 2006, 220(6): 687-693.

[66]

Sonntag R, Reinders J, Rieger JS, et al. Hard-on-hard lubrication in the artificial hip under dynamic loading conditions. Plos One, 2013, 8(8): e71622

[67]

Mazzucco D, Spector M. Effects of contact area and stress on the volumetric wear of ultrahigh molecular weight polyethylene. Wear, 2003, 254(5–6): 514-522.

[68]

Xiong DS. Friction and wear properties of UHMWPE composites reinforced with carbon fiber. Mater Lett, 2005, 59(2–3): 175-179.

[69]

Saikko V. Effect of contact pressure on wear and friction of ultra-high molecular weight polyethylene in multidirectional sliding. Proc Inst Mech Eng Part H J Eng Med, 2006, 220(7): 723-731.

[70]

Wang A, Essner A, Klein R. Effect of contact stress on friction and wear of ultra-high molecular weight polyethylene in total hip replacement. Proc Inst Mech Eng Part H J Eng Med, 2001, 215(2): 133-139.

[71]

Myant C, Underwood R, Fan J, et al. Lubrication of metal-on-metal hip joints: The effect of protein content and load on film formation and wear. J Mech Behav Biomed Mater, 2012, 6: 30-40.

[72]

Leslie I, Williams S, Brown C, et al. Effect of bearing size on the long-term wear, wear debris, and ion levels of large diameter metal-on-metal hip replacements—an in vitro study. J Biomed Mater Res B Appl Biomater, 2008, 87(1): 163-172.

[73]

Chyr A, Qiu M, Speltz JW, et al. A patterned microtexture to reduce friction and increase longevity of prosthetic hip joints. Wear, 2014, 315(1–2): 51-57.

[74]

Zhang H, Qin LG, Hua M, et al. A tribological study of the petaloid surface texturing for Co-Cr-Mo alloy artificial joints. Appl Surf Sci, 2015, 332: 557-564.

[75]

Guezmil M, Bensalah W, Mezlini S. Effect of bio-lubrication on the tribological behavior of UHMWPE against M30NW stainless steel. Tribol Int, 2016, 94: 550-559.

[76]

Yan Y, Dowson D, Neville A. In-situ electrochemical study of interaction of tribology and corrosion in artificial hip prosthesis simulators. J Mech Behav Biomed Mater, 2013, 18: 191-199.

[77]

Sedlaček M, Podgornik B, Vižintin J. Correlation between standard roughness parameters skewness and kurtosis and tribological behaviour of contact surfaces. Tribol Int, 2012, 48: 102-112.

[78]

Kovalchenko A, Ajayi O, Erdemir A, et al. Friction and wear behavior of laser textured surface under lubricated initial point contact. Wear, 2011, 271(9–10): 1719-1725.

[79]

Liu F, Fisher J. Effect of an edge at cup rim on contact stress during micro-separation in ceramic-on-ceramic hip joints. Tribol Int, 2017, 113: 323-329.

[80]

Dowson D, Taheri S, Wallbridge N. The role of counterface imperfections in the wear of polyethylene. Wear, 1987, 119(3): 277-293.

[81]

Fisher J, Firkins P, Reeves E, et al. The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc Inst Mech Eng Part H J Eng Med, 1995, 209(4): 263-264.

[82]

Dowling JM, Atkinson J, Dowson D, et al. The characteristics of acetabular cups worn in the human body. J Bone Joint Surg Br, 1978, 60(3): 375-382.

[83]

Isaac G, Atkinson J, Dowson D, et al. The role of cement in the long term performance and premature failure of Charnley low friction arthroplasties. Eng Med, 1986, 15(1): 19-22.

[84]

Kubo K, Clarke I, Sorimachi T, et al. Aggressive 3rd-body wear challenge to highly crosslinked polyethylene: A hip simulator model. Wear, 2009, 267(5–8): 734-742.

[85]

Cooper J, Dowson D, Fisher J, et al. Ceramic bearing surfaces in total artificial joints: Resistance to third body wear damage from bone cement particles. J Med Eng Technol, 1991, 15(2): 63-67.

[86]

Atkinson J, Dowson D, Isaac G, et al. Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: II: A microscopical study of the surfaces of Charnley polyethylene acetabular sockets. Wear, 1985, 104(3): 217-224.

[87]

Atkinson J, Dowson D, Isaac J, et al. Laboratory wear tests and clinical observations of the penetration of femoral heads into acetabular cups in total replacement hip joints: III: The measurement of internal volume changes in explanted Charnley sockets after 2–16 years in vivo and the determination of wear factors. Wear, 1985, 104(3): 225-244.

[88]

Caravia L, Dowson D, Fisher J, et al. The influence of bone and bone cement debris on counterface roughness in sliding wear tests of ultra-high molecular weight polyethylene on stainless steel. Proc Inst Mech Eng Part H J Eng Med, 1990, 204(1): 65-70.

[89]

Zietz C, Fabry C, Reinders J, et al. Wear testing of total hip replacements under severe conditions. Expert Rev Med Devices, 2015, 12(4): 393-410.

[90]

Hembus J, Lux L, Jackszis M, et al. Wear analysis of cross-linked polyethylene inserts articulating with alumina and ion-treated cobalt-chromium femoral heads under third-body conditions. Wear, 2018, 402: 216-223.

[91]

Affatato S, Bersaglia G, Foltran I, et al. The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: Role of the third-body wear process. Biomaterials, 2002, 23(24): 4839-4846.

[92]

Lundberg HJ, Liu SS, Callaghan JJ, et al. Association of third body embedment with rim damage in retrieved acetabular liners. Clin Orthop Relat Res, 2007, 465: 133-139.

[93]

Oberbach T, Begand S, Kaddick C. Surface resistance of dispersion ceramics against third body abrasion. Key Eng Mater, 2009, 396–398: 161-164.

[94]

Tetreault DM, Kennedy FE. Friction and wear behavior of ultrahigh molecular weight polyethylene on Co-Cr and titanium alloys in dry and lubricated environments. Wear, 1989, 133(2): 295-307.

[95]

Wang A, Essner A, Stark C, et al. Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditions. Biomaterials, 1996, 17(9): 865-871.

[96]

Cooper J, Dowson D, Fisher J. The effect of transfer film and surface roughness on the wear of lubricated ultra-high molecular weight polyethylene. Clinical Materials, 1993, 14(4): 295-302.

[97]

Onate J, Comin M, Braceras I, et al. Wear reduction effect on ultra-high-molecular-weight polyethylene by application of hard coatings and ion implantation on cobalt chromium alloy, as measured in a knee wear simulation machine. Surf Coat Technol, 2001, 142: 1056-1062.

[98]

Geigy D. Scientific tables, 1970, 7 Basel: Geigy Pharmaceuticals

[99]

Wang A, Essner A, Schmidig G. The effects of lubricant composition on in vitro wear testing of polymeric acetabular components. J Biomed Mater Res Part B Appl Biomater, 2004, 68(1): 45-52.

[100]

Dowson D. Bio-tribology. Faraday Discuss, 2012, 156(1): 9-30.

[101]

Nečas D, Vrbka M, Rebenda D, et al. In situ observation of lubricant film formation in THR considering real conformity: The effect of model synovial fluid composition. Tribol Int, 2018, 117: 206-216.

[102]

Aurora A, DesJardins J, Joseph P, et al. Effect of lubricant composition on the fatigue properties of ultra-high molecular weight polyethylene for total knee replacement. Proc Inst Mech Eng Part H J Eng Med, 2006, 220(4): 541-551.

[103]

Hu P, Liu R, Liu J, et al. Advanced Stellite alloys with improved metal-on-metal bearing for hip implants. Mater Des, 2014, 60: 424-432.

[104]

Blunt L, Bills P, Jiang X, et al. The role of tribology and metrology in the latest development of bio-materials. Wear, 2009, 266(3–4): 424-431.

[105]

Bills P, Brown L, Jiang X, et al. A metrology solution for the orthopaedic industry. J Phys Conf Ser, 2005, 13: 316-319.

[106]

Fan J, Myant C, Underwood R, et al. Inlet protein aggregation: A new mechanism for lubricating film formation with model synovial fluids. Proc Inst Mech Eng Part H J Eng Med, 2011, 225(7): 696-709.

[107]

Nečas D, Vrbka M, Křupka I, et al. Lubrication within hip replacements: implication for ceramic-on-hard bearing couples. J Mech Behav Biomed Mater, 2016, 61: 371-383.

Funding

Science Foundation Ireland(15/RP/B3208)

National Natural Science Foundation of China http://dx.doi.org/10.13039/501100001809(51320105009)

AI Summary AI Mindmap
PDF

204

Accesses

0

Citation

Detail

Sections
Recommended

AI思维导图

/