Hemolysis in Pulsed Field Ablation for Atrial Fibrillation: A Narrative Review

Angelica Cersosimo , Nicola Pierucci , Marco Valerio Mariani , Andrea Matteucci , Andrea Giovanni Parato , Vincenzo Mirco La Fazia , Andrea Natale

Reviews in Cardiovascular Medicine ›› 2026, Vol. 27 ›› Issue (3) : 46485

PDF (5325KB)
Reviews in Cardiovascular Medicine ›› 2026, Vol. 27 ›› Issue (3) :46485 DOI: 10.31083/RCM46485
Review
review-article
Hemolysis in Pulsed Field Ablation for Atrial Fibrillation: A Narrative Review
Author information +
History +
PDF (5325KB)

Abstract

Pulsed field ablation (PFA) has emerged as a promising non-thermal energy source for treating atrial fibrillation (AF), demonstrating comparable efficacy to traditional thermal ablation techniques while offering an improved safety profile. However, recent evidence suggests that PFA may be associated with intravascular hemolysis, a complication that can potentially lead to acute kidney injury (AKI). This review aims to provide a comprehensive overview of the mechanisms of hemolysis associated with PFA and to summarize current strategies to mitigate the risk of AKI. The delivery of high-voltage electrical pulses during PFA can induce red blood cell lysis, resulting in elevated circulating free hemoglobin. The extent of hemolysis has been shown to correlate with several procedural variables, including peak output voltage, catheter-tissue contact quality, and, particularly, the number of energy applications delivered. Recent studies have highlighted that adequate preprocedural hydration may effectively reduce the incidence of AKI by promoting renal clearance of hemolytic products. Although hemolysis appears to be an unavoidable effect of pulsed field ablation, the clinical consequences associated with hemolysis, particularly AKI, can be significantly reduced with preventive measures.

Graphical abstract

Keywords

acute kidney injury / hemolysis / pulsed field ablation / atrial fibrillation

Cite this article

Download citation ▾
Angelica Cersosimo, Nicola Pierucci, Marco Valerio Mariani, Andrea Matteucci, Andrea Giovanni Parato, Vincenzo Mirco La Fazia, Andrea Natale. Hemolysis in Pulsed Field Ablation for Atrial Fibrillation: A Narrative Review. Reviews in Cardiovascular Medicine, 2026, 27(3): 46485 DOI:10.31083/RCM46485

登录浏览全文

4963

注册一个新账户 忘记密码

1. Introduction

Catheter ablation for atrial fibrillation has conventionally been performed with thermal energy but is limited by the risk of collateral tissue injury [1, 2, 3, 4]. These safety concerns have led to the development of non-thermal approaches, particularly pulsed field ablation (PFA) [5]. This delivers ultra-rapid (nano and microseconds), high-voltage electrical pulses that induce irreversible electroporation, increasing membrane permeability and disrupting cellular homeostasis, ultimately leading to programmed cell death [6, 7, 8, 9, 10]. Different tissues have significantly different electroporation thresholds: cardiomyocytes have the lowest threshold among all tissues, requiring only 400 V/cm [11]. As opposite, nerve cells need 3800 V/cm [12], and vascular smooth muscle cells require 1750 V/cm [13].

Various devices utilizing PFA technology have been developed that could essentially be categorized into two groups: large footprint devices such as Farapulse (Boston Scientific), Pulseselect (Medtronic), Varipulse (Johnson&Johnson), and focal point-by-point devices such as Affera (Medtronic).

Large clinical studies have shown an association between pulsed field ablation and acute kidney injury related to procedure-induced hemolysis [14].

2. Hemolysis: Pathophysiological Mechanisms, Occurrence and Marker

Hemolysis refers to the destruction of red blood cells (RBCs), resulting in the release of free hemoglobin (Hb) into the circulation [15]. When this process occurs within the vascular system, it is termed intravascular hemolysis. Under normal physiological conditions, free Hb released into the bloodstream is rapidly bound by haptoglobin, a plasma protein that acts as a hemoglobin scavenger [15]. The resulting Hb-haptoglobin complex is then transported to the liver and spleen, where it is internalized and metabolized via macrophages expressing the CD163+ scavenger receptor. However, when hemolysis is excessive and the haptoglobin-binding capacity becomes saturated, free Hb remains unbound and circulates freely in the plasma [16]. In this scenario, the renal glomerulus becomes the primary route for Hb clearance. This filtration of free Hb can have nephrotoxic effects, particularly when the iron within ferrous hemoglobin (Fe2+) undergoes oxidation to form ferric hemoglobin (Fe3+), releasing free heme [16]. Free heme is a potent pro-oxidant and can trigger significant oxidative stress within the renal tubular epithelium. In detail, free heme and free hemosiderin, drive the majority of tubular oxidative stress, generating mitochondrial dysfunction, lipid peroxidation, and inflammatory injury. These pathways underlie hemoglobinuria-associated acute kidney injury (AKI) [17]. Importantly, the toxicity of free heme is not kidney-limited; oxidative injury may also affect the vascular endothelium, myocardial tissue, and reticuloendothelial organs, highlighting the systemic nature of intravascular hemolysis.

In cardiology intravascular hemolysis is observed particularly in patients using post mechanical circulatory support devices. This condition arises when mechanical stress on circulating red blood cells (RBCs) leads to their fragmentation and the release of cellular content into the bloodstream. Contributing factors to this phenomenon include inadequate preload and device thrombosis, both of which increase the mechanical stress on RBCs [18]. Therefore, the mechanism of PFA induced hemolysis is not completely understood but is suspected to be the exposure of red blood cells to high-voltage electrical pulses results in their lysis and subsequent release of free hemoglobin through a stepwise mechanism: the high voltage leads to leakage of ions, which causes colloidal hemolysis of the RBCs due to an induced osmotic imbalance [19].

Red blood cells begin to release their contents into the bloodstream when pulses of intensity ranging from 1100 to 1220 V/cm are applied. A significant percentage of red blood cells undergo lysis when the electroporation field reaches 1500 V/cm, and at levels exceeding 1600 V/cm, nearly all red blood cells are destroyed [20]. To achieve a lesion of 3 mm depth with optimal tissue-catheter contact, 700 V/cm is required. However, when the contact is suboptimal, achieving the same lesion depth necessitates increasing the energy up to 1500 V/cm [21]. Moreover, variability in catheter–tissue contact can increase the heterogeneity of the electric field and promote energy dispersion into the circulating blood pool [22]. Large-footprint or multielectrode catheters may accentuate this effect, especially in anatomically irregular regions such as the posterior wall and mitral isthmus, where stable contact is difficult to achieve. Biphasic waveforms, nanosecond–microsecond pulse durations, and higher peak voltages increase the degree of membrane permeabilization [23]. Repetitive pulse trains may further potentiate hemolysis by cumulative charge delivery [23]. Although PFA is designed to target tissue selectively, RBCs circulating in high-field regions remain vulnerable when exposed to suprathreshold intensities [24].

Considering the amount of applications required during the atrial fibrillation (AF) ablation in particular if applications beyond PV are required, becomes clear that hemolysis is almost unavoidable. The two single shot PFA devices, Farapulse and PulseSelect, release a bipolar pulse train with a maximum peak of 2000 V and 1600 V respectively [25]. The peak voltage of the PF waveform also contributes to the degree of hemolysis. In the presence of hemolysis, an increase in reticulocytes is expected as the bone marrow’s normal response to the peripheral loss of red blood cells [26]. An increase in unconjugated bilirubin and LDH and a decrease in haptoglobin levels accompany the destruction of red blood cells. Then once haptoglobin’s capacity to bind free hemoglobin is saturated, the free hemoglobin is filtered by the glomerulus, leading to hemoglobinuria. This condition causes red-brown urine and is indicated by a positive urine test for heme in the absence of red blood cells [27].

3. Current Evidence

The current literature provides limited but growing insight into the relationship between hemolysis and AKI following PFA. Although PFA has been widely recognized for its myocardial selectivity and favorable safety profile, emerging data indicate that intravascular hemolysis is a consistent occurrence during the procedure and may, in some circumstances, contribute to renal injury.

The first report was provided by Venier et al. [28], who studied 68 consecutive patients undergoing AF PFA with the Farapulse system between June and October 2023. In this cohort, 28% of patients demonstrated laboratory evidence of hemolysis, as evidenced by a significant reduction in haptoglobin levels below 0.04 g/L. Additionally, 27% of patients showed signs of hemoglobinuria, 77% had elevated lactate dehydrogenase (LDH) levels, and 35% exhibited increased levels of free hemoglobin [28]. Despite these laboratory abnormalities indicating red blood cell lysis, no cases of AKI were observed post-procedure in this specific cohort. However, the same group reported two additional cases of AKI due to intravascular hemolysis following PFA with the Farapulse system, which occurred in May and June 2023 [28].

In another study by Mohanty et al. [22], 28 patients underwent PFA, and 21 (75%) developed hemoglobinuria. Four of these patients progressed to AKI, reinforcing the potential renal implications of extensive hemolysis. Both the Mohanty et al. [22] and Venier et al. [28] studies clearly demonstrated a positive correlation between the number of PFA applications and hemolysis severity. Patients with hemoglobinuria had significantly higher LDH and total bilirubin levels, along with lower haptoglobin concentrations. In the first study [28], patients who developed hemoglobinuria had received an average of 57.38 ± 22.78 PFA applications compared to 32.43 ± 4.27 in those who did not. Among the four patients who developed AKI, the number of applications was notably higher. Similarly, in the second study [22], patients with hemoglobinuria received between 62 and 127 applications, while those without hemoglobinuria had between 40 and 60. The two patients who developed AKI had undergone 174 and 126 applications, respectively. Furthermore, a positive correlation has been reported between the number of PFA applications and the concentration of RBCs [29], which is indicative of hemolysis.

Recent evidence suggests that the degree of hemolysis associated with PFA may vary depending significantly on the type and size of the ablation catheter used [30]. In particular, larger, multi-electrode catheters appear to be linked with a higher incidence of hemolysis compared to smaller, point-by-point catheters. This was highlighted in the NEMESIS-PFA study [31], which systematically investigated tissue injury and hemolysis across different catheter platforms used in PFA procedures. In detail, these authors reported that patients treated with larger lattice-style or multi-electrode catheters demonstrated elevated levels of hemolysis markers, including lactate dehydrogenase (LDH) and plasma-free hemoglobin, when compared to those treated with smaller, focal-tip catheters [31]. The difference in hemolysis burden may be attributable to catheter-tissue interface characteristics. Larger catheters, while designed for anatomical efficiency and faster pulmonary vein isolation, may not achieve consistent wall contact in anatomically complex regions such as the posterior wall or ridge between the left pulmonary veins. In these regions, incomplete or unstable contact can result in unintended energy dispersion into the circulating blood pool rather than focused delivery to myocardial tissue. Indeed, dispersion increases the likelihood of red blood cell damage due to the electroporation effect extending into the blood rather than being confined to the tissue [32]. A recent study by Martinez et al. [33] found a correlation between the number of PFA applications and creatinine increase, estimating 129–140 applications to observe a 0.3 mg/dL rise, while no association was observed between hemolysis markers and AKI. Importantly, even patients receiving high volumes and high numbers of PFA pulses without pre-procedural saline loading did not experience significant renal deterioration.

These findings collectively underscore that the number of PFA applications is a key determinant of hemolysis severity and the associated risk of AKI. Based on the current evidence, a threshold of approximately 100 applications may represent a critical cut-off for identifying patients at higher risk for renal complications. Importantly, patients with pre-existing renal impairment may develop AKI even with a lower number of applications, highlighting the need for tailored procedural strategies, careful monitoring, and the use of pre-procedural hydration to mitigate renal risk. Table 1 (Ref. [27, 29, 30, 31, 34, 35]) summarizes the current evidence about hemolysis during PFA procedure.

4. Hemolysis in Thermal Energy Sources

With thermal energy sources severe renal dysfunction as a complication is rare and mostly anecdotal [36]. The available data comparing the impact of hemolysis in AF ablation procedures using thermal sources (radiofrequency (RF) and cryoablation) and non-thermal sources (PFA) are limited and somewhat conflicting.

Modest alterations in hemolysis markers, specifically, only borderline changes in LDH and haptoglobin, and a mild increase in RBC microparticles (RBCµ) are observed following RF ablation, although these changes appear less pronounced than those associated with PFA. Additionally, no significant variation in serum creatinine concentration is detected after RF ablation. This limited degree of hemolysis is likely attributable both to the direct thermal effects of RF application (i.e., red blood cells in vitro undergo budding and fragmentation at temperatures above 49 °C) and to indirect effects mediated by coagulation activation, which generates mechanical stress [29]. A comparative study investigating the occurrence of hemolysis in patients undergoing PFA versus radiofrequency (RF) or cryoballoon ablation reported conflicting results [35]. The authors observed that acute AKI occurred less frequently in the PFA group, and fewer patients showed altered markers of intravascular hemolysis compared to those treated with thermal energy sources. However, significant differences in group sizes may have influenced the findings: 1707 patients in the RF group, 557 in the cryoballoon group, and only 306 in the PFA group. Additionally, the average number of applications per patient was higher in the PFA cohort (32 applications), potentially introducing procedural variability [35]. The hemolysis markers used (lactate dehydrogenase and bilirubin) are nonspecific and may not accurately reflect the true incidence of hemolysis. Moreover, hypotension during general anesthesia, a known contributor to renal injury, was not uniformly controlled for and could have confounded the results. These factors collectively raise concerns about potential misinterpretation of the extent of hemolysis among the different treatment groups [24].

5. How to Minimize the Risk of AKI

Hemolysis is an almost inevitable consequence of PFA procedures for AF. While this phenomenon cannot be entirely prevented due to the intrinsic nature of the energy delivery mechanism, its physiological repercussions. Particularly, its impact on renal function, must be carefully managed. The breakdown of red blood cells releases free hemoglobin and other intracellular contents into the circulation, which, if not promptly cleared, can accumulate in the renal tubules and contribute to AKI. Therefore, strategies aimed at mitigating the renal impact of hemolysis are crucial, especially in patients undergoing a high number of applications during the procedure.

One of the most effective and straightforward approaches to protecting renal function is the implementation of an adequate hydration protocol before and during the procedure. Pre-procedural intravenous administration of fluids has been shown to play a protective role by enhancing renal perfusion and diluting circulating hemoglobin and cellular debris.

The high-voltage electrical pulses cause hemolysis, releasing free hemoglobin into the bloodstream. Hemoglobin-induced tubular-barrier deregulation and oxidative cell damage result in AKI.

The severity of hemolysis increases with higher peak voltage and a higher number of applications (90–100), correlating with an increased risk of AKI. Free hemoglobin can be neutralized by haptoglobin. However, when the capacity to bind free hemoglobin is exceeded, it is filtered by the glomerulus, leading to hemoglobinuria. Given the potential for hemolysis to cause downstream complications, including AKI, strategies to minimize this risk are essential (Fig. 1).

Adequate periprocedural hydration is a key preventive strategy to reduce the risk of acute kidney injury after pulsed field ablation. Pre-procedural intravenous fluid administration has been shown to lower AKI incidence by improving renal perfusion and facilitating clearance of hemolysis-related byproducts [37]. However, signs of hemolysis have been reported even in hydrated patients. Therefore, hydration should be considered part of a broader risk-mitigation for AKI.

In addition to maintaining adequate hydration, it is essential to avoid periprocedural hypotension, which represents another important and potentially preventable risk factor for renal injury [38]. Hypotension can compromise renal perfusion, exacerbating the effects of hemolysis [39]. One common cause of periprocedural hypotension in AF ablation is a vagal response, particularly during energy delivery near the left superior pulmonary vein. To prevent this, administration of 1 mg of atropine after successful transseptal access is recommended. Atropine, an anticholinergic agent, effectively blunts the parasympathetic response, thus reducing the likelihood of vagally mediated bradycardia and hypotension during ablation [40].

Furthermore, ensuring optimal catheter-tissue contact is critical in maximizing the efficiency of energy delivery and minimizing the unintended release of energy into the blood pool, which can increase hemolysis. Applications delivered with poor contact may result in excessive interaction between the pulsed field energy and circulating blood, leading to higher levels of red blood cell destruction. To avoid this, operators should aim to confirm stable and effective contact between the ablation catheter and the targeted myocardial tissue. Intracardiac echocardiography enables real-time visualization of catheter position and stability, allowing operators to adjust and confirm proper apposition before energy delivery. This minimizes applications in regions with suboptimal contact, where energy could otherwise be directed into the bloodstream, increasing the risk of hemolysis. For instance, Mohanty et al. [41] recently showed that the use of ICE significantly improves the quality of contact between the ablation catheter and atrial tissue.

Although hemolysis during AF PFA is largely unavoidable, its downstream effects, especially on renal function, can be significantly attenuated through a combination of proactive hydration, hemodynamic stability, and careful procedural technique. These measures are essential in optimizing patient outcomes and reducing the incidence of post-procedural AKI (Fig. 2).

6. Hemolysis Across Different PFA Systems

Available evidence indicates that hemolysis is a frequent but heterogeneous phenomenon following PFA, with substantial variability related both to the methods used for its assessment and to the catheter technologies under investigation. Kawamura et al. [30] evaluated hemolysis using a panel of biochemical markers, including free hemoglobin, LDH, total bilirubin, and haptoglobin, measured before and after the procedure, demonstrating significant differences between single-shot PFA systems despite a comparable number of applications. Their findings suggested that system-specific characteristics contribute to hemolytic burden, while the absence of acute kidney injury supported a predominantly subclinical manifestation of hemolysis.

Gianni et al. [42] conducted a large retrospective analysis that systematically evaluated the extent of hemolysis associated with different PFA technologies. In this study, haptoglobin was used as the primary marker of intravascular hemolysis, complemented by free hemoglobin, LDH, indirect bilirubin, and renal function parameters, allowing for a comprehensive biological and clinical characterization [42]. Although a reduction in haptoglobin was observed in the vast majority of cases, the severity of hemolysis differed markedly across catheter types. Notably, the focal Sphere-9 catheter, characterized by a smaller size allowing better contact, was associated with significantly lower rates of significant and severe hemolysis compared with larger single-shot catheters such as Farawave, PulseSelect, and Varipulse. These findings suggest that catheter geometry and size may play a key role in limiting erythrocyte exposure to the electric field, beyond the effects of application number or nominal voltage.

Consistent results were reported by Bruss et al. [37], who compared all PFA platforms commercially available in Europe using haptoglobin and LDH as markers of hemolysis. Their analysis confirmed that focal lattice-tip catheters induced a significantly smaller decrease in haptoglobin both per procedure and per application than larger footprint catheters, further supporting the relevance of catheter design in modulating hemolytic effects.

Taken together, these studies indicate that PFA-related hemolysis is strongly influenced by system-specific factors. In particular, available evidence highlights that smaller focal catheters are associated with a more favorable hemolytic profile, a finding with potential implications for tailoring ablation strategies in patients at increased risk of hemolysis-related complications.

7. Conclusion

Exposing red blood cells to high-voltage pulses during PFA inevitably leads to hemolysis. With a high number of applications (90–100), the severity of hemolysis increases, correlating with a higher risk of developing AKI. A strategy to minimizes the risk of AKI includes limiting applications to only those essential for therapeutic goals and ensuring adequate hydration before the procedure.

References

[1]

Van Gelder IC, Rienstra M, Bunting KV, Casado-Arroyo R, Caso V, Crijns HJGM, et al. 2024 ESC Guidelines for the management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). European Heart Journal. 2024; 45: 3314–3414. https://doi.org/10.1093/eurheartj/ehae176.

[2]

Haïssaguerre M, Jaïs P, Shah DC, Takahashi A, Hocini M, Quiniou G, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. The New England Journal of Medicine. 1998; 339: 659–666. https://doi.org/10.1056/NEJM199809033391003.

[3]

Bonelli A, Degiovanni A, Cersosimo A, Spinoni EG, Bosco M, Dell’Era G, et al. Determinants of invasive left atrial pressure in patients with atrial fibrillation. European Heart Journal. Cardiovascular Imaging. 2024; 25: 1590–1598. https://doi.org/10.1093/ehjci/jeae194.

[4]

Joglar JA, Chung MK, Armbruster AL, Benjamin EJ, Chyou JY, Cronin EM, et al. 2023 ACC/AHA/ACCP/HRS Guideline for the Diagnosis and Management of Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2024; 149: e1–e156. https://doi.org/10.1161/CIR.0000000000001193.

[5]

Verma A, Haines DE, Boersma LV, Sood N, Natale A, Marchlinski FE, et al. Pulsed Field Ablation for the Treatment of Atrial Fibrillation: PULSED AF Pivotal Trial. Circulation. 2023; 147: 1422–1432. https://doi.org/10.1161/CIRCULATIONAHA.123.063988.

[6]

Schmidt B, Bordignon S, Neven K, Reichlin T, Blaauw Y, Hansen J, et al. EUropean real-world outcomes with Pulsed field ablatiOn in patients with symptomatic atRIAl fibrillation: lessons from the multi-centre EU-PORIA registry. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2023; 25: euad185. https://doi.org/10.1093/europace/euad185.

[7]

Verma A, Asivatham SJ, Deneke T, Castellvi Q, Neal RE. Primer on Pulsed Electrical Field Ablation: Understanding the Benefits and Limitations. Circulation. Arrhythmia and Electrophysiology. 2021; 14: e010086. https://doi.org/10.1161/CIRCEP.121.010086.

[8]

Xie F, Varghese F, Pakhomov AG, Semenov I, Xiao S, Philpott J, et al. Ablation of Myocardial Tissue With Nanosecond Pulsed Electric Fields. PloS One. 2015; 10: e0144833. https://doi.org/10.1371/journal.pone.0144833.

[9]

Ruzgys P, Novickij V, Novickij J, Šatkauskas S. Influence of the electrode material on ROS generation and electroporation efficiency in low and high frequency nanosecond pulse range. Bioelectrochemistry (Amsterdam, Netherlands). 2019; 127: 87–93. https://doi.org/10.1016/j.bioelechem.2019.02.002.

[10]

Verma A, Zhong P, Castellvi Q, Girouard S, Mediratta V, Neal RE, 2nd. Thermal Profiles for Focal Pulsed Electric Field Ablation. JACC. Clinical Electrophysiology. 2023; 9: 1854–1863. https://doi.org/10.1016/j.jacep.2023.05.005.

[11]

Kaminska I, Kotulska M, Stecka A, Saczko J, Drag-Zalesinska M, Wysocka T, et al. Electroporation-induced changes in normal immature rat myoblasts (H9C2). General Physiology and Biophysics. 2012; 31: 19–25. https://doi.org/10.4149/gpb_2012_003.

[12]

Li W, Fan Q, Ji Z, Qiu X, Li Z. The effects of irreversible electroporation (IRE) on nerves. PloS One. 2011; 6: e18831. https://doi.org/10.1371/journal.pone.0018831.

[13]

Maor E, Ivorra A, Rubinsky B. Non thermal irreversible electroporation: novel technology for vascular smooth muscle cells ablation. PloS One. 2009; 4: e4757. https://doi.org/10.1371/journal.pone.0004757.

[14]

Ekanem E, Neuzil P, Reichlin T, Kautzner J, van der Voort P, Jais P, et al. Safety of pulsed field ablation in more than 17,000 patients with atrial fibrillation in the MANIFEST-17K study. Nature Medicine. 2024; 30: 2020–2029. https://doi.org/10.1038/s41591-024-03114-3.

[15]

Vallelian F, Buehler PW, Schaer DJ. Hemolysis, free hemoglobin toxicity, and scavenger protein therapeutics. Blood. 2022; 140: 1837–1844. https://doi.org/10.1182/blood.2022015596.

[16]

Schaer DJ, Buehler PW, Alayash AI, Belcher JD, Vercellotti GM. Hemolysis and free hemoglobin revisited: exploring hemoglobin and hemin scavengers as a novel class of therapeutic proteins. Blood. 2013; 121: 1276–1284. https://doi.org/10.1182/blood-2012-11-451229.

[17]

Deuel JW, Schaer CA, Boretti FS, Opitz L, Garcia-Rubio I, Baek JH, et al. Hemoglobinuria-related acute kidney injury is driven by intrarenal oxidative reactions triggering a heme toxicity response. Cell Death & Disease. 2016; 7: e2064. https://doi.org/10.1038/cddis.2015.392.

[18]

Balthazar T, Bennett J, Adriaenssens T. Hemolysis during short-term mechanical circulatory support: from pathophysiology to diagnosis and treatment. Expert Review of Medical Devices. 2022; 19: 477–488. https://doi.org/10.1080/17434440.2022.2108319.

[19]

Kinosita K, Jr, Tsong TT. Hemolysis of human erythrocytes by transient electric field. Proceedings of the National Academy of Sciences of the United States of America. 1977; 74: 1923–1927. https://doi.org/10.1073/pnas.74.5.1923.

[20]

Bao N, Le TT, Cheng JX, Lu C. Microfluidic electroporation of tumor and blood cells: observation of nucleus expansion and implications on selective analysis and purging of circulating tumor cells. Integrative Biology: Quantitative Biosciences from Nano to Macro. 2010; 2: 113–120. https://doi.org/10.1039/b919820b.

[21]

Meckes D, Emami M, Fong I, Lau DH, Sanders P. Pulsed-field ablation: Computational modeling of electric fields for lesion depth analysis. Heart Rhythm O2. 2022; 3: 433–440. https://doi.org/10.1016/j.hroo.2022.05.009.

[22]

Mohanty S, Casella M, Compagnucci P, Torlapati PG, Della Rocca DG, La Fazia VM, et al. Acute Kidney Injury Resulting From Hemoglobinuria After Pulsed-Field Ablation in Atrial Fibrillation: Is it Preventable? JACC. Clinical Electrophysiology. 2024; 10: 709–715. https://doi.org/10.1016/j.jacep.2023.12.008.

[23]

Xiao S, Guo S, Nesin V, Heller R, Schoenbach KH. Subnanosecond electric pulses cause membrane permeabilization and cell death. IEEE Transactions on Bio-medical Engineering. 2011; 58: 1239–1245. https://doi.org/10.1109/TBME.2011.2112360.

[24]

La Fazia VM, Mohanty S, Torlapati PG, Gianni C, Della Rocca DG, Bode W, et al. Hydration to Prevent Kidney Injury After Pulsed Field Ablation: Importance of Timing and Fluids Amount. JACC. Clinical Electrophysiology. 2025; 11: 599–601. https://doi.org/10.1016/j.jacep.2024.11.001.

[25]

Ezzeddine FM, Asirvatham SJ, Nguyen DT. Pulsed Field Ablation: A Comprehensive Update. Journal of Clinical Medicine. 2024; 13: 5191. https://doi.org/10.3390/jcm13175191.

[26]

Piva E, Brugnara C, Chiandetti L, Plebani M. Automated reticulocyte counting: state of the art and clinical applications in the evaluation of erythropoiesis. Clinical Chemistry and Laboratory Medicine. 2010; 48: 1369–1380. https://doi.org/10.1515/CCLM.2010.292.

[27]

Popa MA, Venier S, Menè R, Della Rocca DG, Sacher F, Derval N, et al. Characterization and Clinical Significance of Hemolysis After Pulsed Field Ablation for Atrial Fibrillation: Results of a Multicenter Analysis. Circulation. Arrhythmia and Electrophysiology. 2024; 17: e012732. https://doi.org/10.1161/CIRCEP.124.012732.

[28]

Venier S, Vaxelaire N, Jacon P, Carabelli A, Desbiolles A, Garban F, et al. Severe acute kidney injury related to haemolysis after pulsed field ablation for atrial fibrillation. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2023; 26: euad371. https://doi.org/10.1093/europace/euad371.

[29]

Osmancik P, Bacova B, Herman D, Hozman M, Fiserova I, Hassouna S, et al. Periprocedural Intravascular Hemolysis During Atrial Fibrillation Ablation: A Comparison of Pulsed Field With Radiofrequency Ablation. JACC. Clinical Electrophysiology. 2024; 10: 1660–1671. https://doi.org/10.1016/j.jacep.2024.05.001.

[30]

Kawamura I, Miyazaki S, Kato R, Takahashi K, Negishi M, Honda M, et al. Comparison of hemolysis with different pulsed field ablation systems. Heart Rhythm. 2026; 23: 388–395. https://doi.org/10.1016/j.hrthm.2025.05.072.

[31]

Lakkireddy D, Katapadi A, Garg J, Herink E, Klotz M, Ganta J, et al. NEMESIS-PFA: Investigating Collateral Tissue Injury Associated With Pulsed Field Ablation. JACC. Clinical Electrophysiology. 2025; 11: 1747–1756. https://doi.org/10.1016/j.jacep.2025.04.017.

[32]

Carta-Bergaz A, Ríos-Muñoz GR, Ávila P, Atienza F, González-Torrecilla E, Arenal Á. Pulsed Field Ablation of Atrial Fibrillation: A Novel Technology for Safer and Faster Ablation. Biomedicines. 2024; 12: 2232. https://doi.org/10.3390/biomedicines12102232.

[33]

Martinez J, Challapalli M, Hutchinson M, Ibrahim M, Shen C, Klewer J, et al. Renal safety of high-dose pulsed field ablation of atrial fibrillation: A prospective real-world analysis. Heart Rhythm. 2026; 23: 396–404. https://doi.org/10.1016/j.hrthm.2025.07.026.

[34]

Auf der Heiden C, Bejinariu AG, Kelm M, Spieker M, Rana O. Hemolysis after pulsed-field ablation in pulmonary vein isolation for atrial fibrillation: A prospective controlled trial. Heart Rhythm. 2026; 23: e209–e216. https://doi.org/10.1016/j.hrthm.2025.04.038.

[35]

Jordan F, Knecht S, Isenegger C, Arnet R, Krisai P, Völlmin G, et al. Acute kidney injury after catheter ablation of atrial fibrillation: Comparison between different energy sources. Heart Rhythm. 2024; 21: 1248–1249. https://doi.org/10.1016/j.hrthm.2024.04.046.

[36]

van Tilborg AAJM, Dresselaars HF, Scheffer HJ, Nielsen K, Sietses C, van den Tol PM, et al. RF Ablation of Giant Hemangiomas Inducing Acute Renal Failure: A Report of Two Cases. Cardiovascular and Interventional Radiology. 2016; 39: 1644–1648. https://doi.org/10.1007/s00270-016-1415-1.

[37]

Bruss J, Kueffer T, Thalmann G, Kozhuharov NA, Haeberlin A, Noti F, et al. Comparison of hemolysis induced by different pulsed-field ablation systems. Europace. 2025; 27: euaf085.444.

[38]

Kher V, Srisawat N, Noiri E, Benghanem Gharbi M, Shetty MS, Yang L, et al. Prevention and Therapy of Acute Kidney Injury in the Developing World. Kidney International Reports. 2017; 2: 544–558.

[39]

Hamrahian SM, Vilayet S, Herberth J, Fülöp T. Prevention of Intradialytic Hypotension in Hemodialysis Patients: Current Challenges and Future Prospects. International Journal of Nephrology and Renovascular Disease. 2023; 16: 173–181. https://doi.org/10.2147/IJNRD.S245621.

[40]

Del Monte A, Della Rocca DG, Pannone L, Vetta G, Cespón Fernández M, Marcon L, et al. Pulsed field ablation of the right superior pulmonary vein prevents vagal responses via anterior right ganglionated plexus modulation. Heart Rhythm. 2024; 21: 780–787. https://doi.org/10.1016/j.hrthm.2024.01.040.

[41]

Mohanty S, Casella M, Doty B, Schiavone M, Gabrah K, Valeri Y, et al. Ensuring catheter-tissue contact with intracardiac echocardiography during pulsed-field ablation improves procedure outcome in patients with atrial fibrillation. Heart Rhythm. 2025; 22: e875–e881. https://doi.org/10.1016/j.hrthm.2025.05.045.

[42]

Gianni C, Al-Ahmad A, Elchouemi M, La Fazia VM, Mohanty S, Allison JD, et al. Pulsed Field Ablation-Related Hemolysis: Comparison Between Technologies. Circulation. Arrhythmia and Electrophysiology. 2025; 18: e014021. https://doi.org/10.1161/CIRCEP.125.014021.

PDF (5325KB)

0

Accesses

0

Citation

Detail

Sections
Recommended

/