Pulse Field Ablation: An Update on Energy-Specific Adverse Events

Minghao Zhou , Hongliang Yang , Huan Sun , Ming Yu , Daoyuan Si , Yuquan He

Reviews in Cardiovascular Medicine ›› 2025, Vol. 26 ›› Issue (10) : 44442

PDF (7014KB)
Reviews in Cardiovascular Medicine ›› 2025, Vol. 26 ›› Issue (10) :44442 DOI: 10.31083/RCM44442
Review
review-article
Pulse Field Ablation: An Update on Energy-Specific Adverse Events
Author information +
History +
PDF (7014KB)

Abstract

Pulsed field ablation (PFA) represents a paradigm shift in cardiac ablation technology. Indeed, PFA, a non-thermal technique, achieves homogeneous tissue effects by delivering ultrashort, high-frequency electric pulses. Moreover, PFA has emerged as a prominent ablation strategy in electrophysiology laboratories worldwide. While current clinical evidence demonstrated promising outcomes and expanded applications for existing PFA platforms, energy-specific difficulties complicated the definition of safety boundaries, particularly in patients with clinical complications. Furthermore, substantial heterogeneity among commercial PFA systems impeded procedural standardization. Therefore, this review aimed to synthesize contemporary perspectives on mitigating these energy-specific adverse events, with emphasis on modifiable perioperative factors. The evidence of safety across commercially available PFA systems with divergent catheter design philosophies was concurrently evaluated to examine how design strategies influence procedural safety profiles.

Graphical abstract

Keywords

atrial fibrillation / pulsed field ablation / energy-specific adverse event / vagal ganglionated plexus / coronary artery spasm / hemolysis

Cite this article

Download citation ▾
Minghao Zhou, Hongliang Yang, Huan Sun, Ming Yu, Daoyuan Si, Yuquan He. Pulse Field Ablation: An Update on Energy-Specific Adverse Events. Reviews in Cardiovascular Medicine, 2025, 26(10): 44442 DOI:10.31083/RCM44442

登录浏览全文

4963

注册一个新账户 忘记密码

1. Introduction

Pulsed field ablation (PFA) technique has emerged as an advanced therapeutic approach for arrhythmias, representing a significant improvement in cardiac ablation technique [1, 2]. It has attracted much attention as a new form of ablation energy and for its preferential myocardial selectivity and effectiveness in forming homogeneous lesions. However, registry data based on the real-world applications for atrial fibrillation (AF) treatment indicated that 0.2% of patients experience energy-specific adverse events (AEs) [3], which represent a critical aspect of PFA safety assessment. Recent statistical analyses of data from Farapulse (Boston Scientific) and Pulseselect (Medtronic) devices revealed that 21% (13/61) of patient complications associated with the Farapulse system are attributable to energy-specific AEs [4]. These complications have caused anxiety among PFA operators and have raised concerns regarding the safety associated with different ablation modalities. However, there is no standardized perioperative management strategy to avoid the potential risk of energy-specific AEs due to the strong heterogeneity of contemporary PFA ablation systems in catheter design and ablation parameter settings. Thus, there is an urgent need for a generalized perioperative intervention plan to avoid the occurrence of unpredictable perioperative AEs due to the increasing frequency of PFA application and the continuous expansion of its indications. Therefore, the standardized PFA ablation procedure should advance by summarizing the existing clinical/device factors associated with energy-specific AEs.

2. The Classification of Energy-Specific Complications

Available meta-analyses indicate that PFA exhibited a superior safety profile when compared to thermal ablation modalities (e.g., radiofrequency ablation [RFA] and cryoballoon ablation [CBA]), particularly regarding serious complications (Table 1) [5, 6, 7, 8, 9, 10, 11, 12, 13, 14]. Nevertheless, PFA is associated with distinct energy-specific AEs, including nerve dysfunction, coronary artery spasm, intravascular hemolysis, and hemolysis-induced acute kidney injury (AKI) [3]. Analysis of large PFA registries (primarily using the Farapulse system) quantified the percentage of the most relevant AEs, as illustrated in Fig. 1 [3, 15, 16]. Specifically, the Multi-National Survey on the Safety of the Post-Approval Clinical Use of Pulsed Field Ablation in 17,000+ Patients (MANIFEST-17K) established definitions for energy-specific AEs identifying risks including transient neurological disturbance (e.g., phrenic nerve palsy and excessive vagal response), vasospasm, and hemolysis-induced renal failure. The analysis of the Food and Drug Administration (FDA) Manufacturer and User Facility Device Experience (MAUDE) database (encompassing 1237 PFA and RFA reports) further delineated energy-specific AEs, thus supporting the above analysis. According to these results, the most frequently reported PFA AEs are pericardial effusion, vasovagal reaction, and hemolysis. This profile is in contrast with that of RFA, where pericardial effusion, ischemic stroke, and esophageal injury are the predominant ones. Vagal reactions (14.1% vs. 0%), coronary spasm (5.8% vs. 0.6%), and hemolysis (9.0% vs. 0%) show significantly higher incidence rates with PFA compared to RFA [17].

Debates regarding the cerebrovascular risks of PFA persist. Despite most studies report comparable rates of stroke/transient ischemic attack (TIA) to thermal ablation, concerns remain on the potential undefined cerebral injury risks [7, 14]. In vitro studies revealed that gas microbubbles (20–200 µm in diameter) formed selectively in the anode region of PFA devices, contribute to acute ischemic events, and significant differences in bubble formation are observed among current commercial systems [18, 19]. The real-time carotid echocardiography that detects the formation of microbubbles revealed that Varipulse generates substantially more microembolic signals (MESs), silent cerebral events, and silent cerebral lesions than other PFA platforms. However, further investigation is needed to confirm that MESs induce impaired neurological or cognitive function. Forleo GB et al. [20] noted that microemboli from catheter ablation approach levels seen in major cardiac surgery, which may drive subtle neuropsychological changes, including higher rates of cognitive impairment. However, as most ischemic signals resolve within 90 days without chronic glial scarring, microbubble formation alone cannot fully define the neurosafety profile of PFA. RFA emboli typically occur during sheath/catheter exchange, while CBA emboli occur during catheter insertion or balloon inflation. However, sheath replacement in PFA may introduce abundant gas emboli that likely result in extensive MES formation, suggesting this step as a potential target for mitigating post-ablation neural injury. Since the specific mechanisms underlying potential PFA-related cerebral injury remain incompletely elucidated, the discussion on energy-specific AEs is focused on those with established associations: vagal responses, coronary spasm, hemolysis, and hemolysis-induced AKI.

3. Available Factors for Energy-Specific Aes Perioperative Intervention

3.1 Intraoperative Vagal Response

3.1.1 Competitive Muscarinic Antagonist

PFA possesses a unique neurotropic selectivity that preserves vagal ganglionic cell viability while eliciting transient autonomic stimulation [21], evidenced by the absence of significant heart rate increase [22] and elevated post-procedural S100 protein levels [23], the former being a biomarker associated with denervation, as shown in Fig. 2. Experience based on current CBA application revealed that, preoperative intravenous administration of 1 mg atropine reduces severe vagal reactions from 92.3% (12/13) to 33.3% (4/12; p < 0.01) [24]. Current PFA protocols do not require their prophylactic administration, while competitive muscarinic antagonists such as atropine and glycopyrrolate are frequently used to mitigate excessive vagal response. This situation requires critical re-evaluation due to the procedural demands of PFA, particularly the repeated pulsed field applications required to achieve sufficient lesion formation in the targeted anatomical regions [25, 26], substantially increasing the risk of vagal activation through electrical stimulation of the cardiac plexuses. Future procedural guidelines should explicitly address the management of intraoperative vagal response by incorporating standardized strategies such as preprocedural or intraprocedural application of competitive muscarinic antagonists and selective use of temporary ventricular pacing leads.

3.1.2 Initial Ablation Site

The vagal ganglia localized in the subepicardial fat pads adjacent to the myocardial layer constitute critical mediators of neurocardiac regulation and represent strategic therapeutic targets for autonomic reflex modulation during cardiac ablation. These neural aggregates function as interconnected networks in the cardiac autonomic architecture (Fig. 3A). Hu F et al. [27]. confirmed that prioritizing intervention and regulation of the right anterior ganglionic plexus (RAGP) during PFA (Fig. 3B,C) significantly reduces the incidence of intraoperative vagus nerve involvement (2.5% vs. 62.5%). This RAGP-centric approach has equivalent efficacy across RFA modalities. Current prospective PFA data reveal a remarkable disparity in autonomic responses: an left superior pulmonary vein (LSPV)-first strategy elicit vagal reactions in 78% of cases versus 13% with right superior pulmonary vein (RSPV) prioritization. Furthermore, the RSPV-focused protocol significantly reduces temporary pacing requirements (35% vs. 8%; p < 0.01) [28].

3.2 Coronary Spasm

3.2.1 Prophylactic Nitroglycerin Administration

The coronary spasm induced by PFA is primarily due to direct stimulation of the vascular smooth muscle by pulsed electric fields (PEF) and the cascade of local inflammatory responses [29, 30], as shown in Fig. 4. The reported incidence of such acute vascular reactions during PFA is from 0.08% to 0.14% [3, 15, 16, 31]. Intravenous nitroglycerin is frequently used as a prophylactic measure during the ablation of the anatomic site adjacent to the coronary artery [32]. However, the administration of this pharmacological treatment does not universally preclude the intracoronary administration to alleviate coronary spasm in all patients. Even with the prophylactic use of nitroglycerin, a subset of patient still experiences delayed and, clinically significant coronary spasm. A pertinent case report documented the occurrence of clinically apparent vasospasm induced by a pentaspline PFA catheter during cavotricuspid isthmus (CTI) ablation, despite the prophylactic administration of high-dose intravenous nitroglycerin prior to PFA. This episode was characterized by a significant vasospasm accompanied by a ST-segment elevation in leads II, III, and aVF on the electrocardiogram. The clinical scenario was further complicated by the onset of hemodynamically unstable nonsustained ventricular tachycardia, culminating in cardiac arrest in need of temporary pacing and an additional 4 mg nitroglycerin [33]. Our hypothesis was that the transient cellular electroshock effect induced by PEF further prolonged the metabolic clearance time of the aforementioned hyperphosphorylated myosin light chain, while reducing vascular smooth muscle sensitivity to nitric oxide, consequently exacerbating vasoconstrictor responses and increasing nitric oxide dosage requirements. As regards the increase of the efficacy of nitrate administration, Malyshev Y et al. [34] propose two innovative nitroglycerin delivery protocols: (1) multiple bolus injections (2–3 mg every 2 minutes) into the right atrium, and (2) a single bolus injection (3 mg) into the right atrium combined with continuous peripheral intravenous infusion (1 mg/min). These strategies show a significant effect in preventing severe spasm, although the optimal dosing regimen remains to be established due to the limited sample size in their study.

3.2.2 Anatomical Sites

PFA-induced coronary spasm shows distinct clinical characteristics, including acute onset patterns and specific anatomical predilections [32, 35, 36, 37, 38, 39, 40]. The mean distance between the endocardium and the right coronary artery at the CTI is 4.2 ± 2.1 mm, while the minimum distance at the mitral isthmus (MI) region is 3.8 ± 2.3 mm (Fig. 5) [41]. However, the ideal tissue proximity is identified as a significant factor influencing the formation of PFA-induced electrical isolation [42]. Clinical experience from pentaspline PFA catheter confirms that coronary spasm frequently occurs during the linear ablation of the CTI and superior MI [32, 43]. Most such cases involve subclinical vasospasm induced by localized electric field stimulation. However, optical coherence tomography observations by Tam MTK et al. [44] reveal significant vascular remodeling three months post-PFA, as the median vessel wall area increases by 17.1% and the lumen area decreases by 10.1%. Potential hemodynamic consequences need attention due to the established correlation between atherosclerosis severity and coronary spasm [45], particularly regarding PFA exacerbating flow impairment in patients with pre-existing coronary abnormalities such as atherosclerotic plaques or myocardial bridges.

3.3 Intravascular Hemolysis and Subsequent Acute Renal Injury

The exposure of erythrocytes to high-voltage pulses (several kV/cm) generates a transmembrane potential, leading to pore formation in the cell membrane at critical thresholds (Fig. 6). This membrane permeabilization results in colloid osmotic hemolysis, characterized by erythrocyte swelling and rupture due to hemoglobin-induced osmotic pressure [46, 47]. Severe hemolysis can precipitate AKI through heme-mediated proximal tubule epithelial damage and intratubular protein condensation, manifesting as oliguria and increased creatinine levels [48]. Current studies related to PFA hemolysis and AKI are summarized in Table 2 (Ref. [49, 50, 51, 52, 53, 54, 55, 56, 57, 58, 59]), which helps to determine reasonable perioperative management strategies.

3.3.1 Patient Condition

The reported incidence of AKI following hemolysis in PFA procedures ranges from 0% to 5.26% among clinical series [49, 50, 51, 52]. A prospective analysis by De Smet MAJ et al. [49] identifies AKI-prone patients as older individuals with comorbidities including chronic kidney disease, heart failure, diabetes, and hypertension, typically requiring extended duration of the ablation. Among them, a patient with stage 3 AKI with a baseline glomerular filtration rate of 14 mL/min developed progressive kidney injury (serum creatinine from 3.58 to 3.79 mg/dL) with hyperkalemia after only 32 pentaspline PFA ablations and started to receive renal replacement therapy. Previous studies demonstrated that patients with a baseline glomerular filtration rate <50 mL/min exhibit a statistically significant increase in serum creatinine levels after PFA (Δcrea: 27.0 ± 103.1 vs. –0.2 ± 12.1 µmol/L; p = 0.010) [50]. This finding suggests that individuals with pre-existing significant renal impairment may represent a high-risk population for PFA-related complications rather than obtaining clinical benefit from the procedure. More importantly, patients with significant preoperative renal dysfunction may experience substantial renal deterioration even with standard PFA protocols.

3.3.2 Application Frequence

The cumulative number of PFA applications is considered as the most significant intraprocedural determinant of postprocedural hemolysis and renal dysfunction. Venier S et al. [51] pioneered the investigation of AKI after PFA, demonstrating a significant inverse correlation between plasma haptoglobin levels and total application count (median [Interquartile Range]: 75 [62–127] vs. 62 [54–71] in hemolytic-positive vs. negative cohorts, p = 0.011). Subsequent validation studies further corroborated these findings, establishing a predictive application threshold range from 54 to 74 discharges for hemolysis risk [50, 51, 53]. Receiver operating characteristic (ROC) curve analysis identifies 70 applications as an optional predictive threshold (area under the curve (AUC): 0.709) with balanced sensitivity and specificity. Procedural safety of Farapulse system can be stratified by discharge frequency (Fig. 7). Hemolysis risk is minimized with less than 54 PFA applications, whereas applications between 54 and 74 markedly increase hemoglobinuria incidence. This dose-dependent effect is particularly significant in patients with persistent AF requiring extensive ablation [49, 52, 54, 60]. Tamirisa KP et al. [61] show that application thresholds must be individualized according to the specific PFA system used. Since catheter-induced hemolysis correlates with non-contact electrode exposure in the atrial blood pool, discharge thresholds derived from pentaspline experience cannot be broadly extrapolated to standardize PFA procedures. Although clinical guidelines currently lack specific discharge thresholds for PFA, our recommendations are the following: (1) prioritize focal ablation or single catheters with minimal footprint when anticipating extra ablation; (2) use sensitive biomarkers as red blood cell microparticles to evaluate hemolysis severity. Since red blood cell microparticles are not routinely used in clinical practice, lactate dehydrogenase, haptoglobin, indirect bilirubin and plasma-free hemoglobin may serve as the preferred alternative parameters [61].

3.3.3 Catheter-Tissue Proximity

Nies M et al. [56] performed in vitro experiments to investigate the relationship between tissue contact and hemolysis using a pentaspline catheter. Analysis of 76 blood samples from four swine indicates that non-contact applications induce significantly greater hemolysis, with free plasma hemoglobin (fHB) levels becoming statistically significant after four applications (0.12 ± 0.03 vs. 0.05 ± 0.03 g/dL; p = 0.008). This phenomenon was also observed with loop catheters, where both contact and non-contact groups show a linear dose-response relationship between fHB levels and PFA applications. The fHB increase per application is 0.0003 g/dL in contact and 0.00056 g/dL in non-contact conditions [57]. Although previous experience suggested that the damage by PFA is not entirely dependent on contact force, poor adherence to tissue can further predict the degree of intraoperative hemolysis [58].

3.3.4 Preoperative Drug Pretreatment and Postoperative Hydration

The perioperative treatment is considered the basis for avoiding severe renal insufficiency caused by hemolysis. The gadolinium-diethylenetriamine-penta-acetic acid, a clinically approved magnetic resonance imaging (MRI) contrast medium, was introduced for reducing PFA-induced hemolysis through erythrocyte membrane stabilization. Preclinical studies reveal that the above acid modulates erythrocyte membrane stability, improving the resistance to electroporation-mediated membrane disruption during PFA [59]. However, this approach is still in the stage of animal model-based validation due to the lack of reliable evidence. Postoperative hydration is the only treatment established in clinical practice to reduce hemolysis-associated kidney injury. Mohanty S et al. [55] demonstrated that, the planned intravenous infusion of 0.9% sodium chloride 2 L effectively prevents serum creatinine increase in 75 patients after ablation. Despite this approach probably complicated the postoperative management due to the risk of volume overload and subsequent clinical deterioration, this measure is necessary especially for patients with preoperative renal insufficiency.

4. Limitation of Contemporary PFA Ablation Catheters

Pulse parameters in PFA systems, including voltage, waveform, pulse duration, and frequency, are predefined by manufacturers during device development, resulting in significant inter-device variability. This inherent variability, coupled with the nonlinear relationship between pulse parameters and procedural risks, complicates the establishment of optimal ablation protocols. Contemporary PFA systems predominantly use microsecond-duration pulses, typically applying voltages around 2000 V with bidirectional waveforms. However, the spatial distribution of PEF generated by multielectrode ablation catheters critically depend on the activation modality, which governs whether PEF superpose constructively or cancel each other out. Contemporary pentaspline PFA catheters use two principal energy delivery strategies: simultaneous activation and sequential activation (Fig. 8). Computer simulations showed that the simultaneous activation produces myocardial lesions with a mean depth of 1.6 mm (maximum 3.2 mm). In contrast, sequential activation achieves greater penetration depth (mean 2.7 mm, maximum 5.1 mm) but needs a higher total energy delivery due to the increased application frequency [62]. This escalated energy exposure increases the risk of hemoglobinuria and subsequent renal injury resulting from cumulative erythrocyte destruction.

Farapulse (Boston Scientific), Pulseselect (Medtronic), and Varipulse (Biosense Webster) are the PFA systems most widely used worldwide by electrophysiology centers. Among these, Farapulse was subjected to systematic safety and efficacy tests, thus extending its applications beyond pulmonary vein isolation, as validated in the atrial fibrillation study evaluating new technology (ADVENT) trial, to include left atrial posterior wall isolation and CTI linear ablation [63, 64, 65]. As regards further ablation treatments carried out in other anatomical regions, Superior Vena Cava isolation and MI linear ablation have both been proven to be feasible under the PFA procedure [66, 67, 68]. The current clinical research progress on contemporary PFA systems is summarized in Fig. 9, and technical characteristics and safety profiles of the new PFA catheter are described in subsequent sections. Varipulse is the only PFA system among the aforementioned devices that incorporates real-time force sensing during the ablation procedure. This suggests its potential advantages in a further optimization of the safety of the procedure.

5. Research Status of Novel PFA Ablation Catheters

5.1 Balloon-in-Basket PFA

The Volt system (Abbott) exemplifies the balloon-type PFA philosophy, integrating real-time three-dimensional (3D) electroanatomical mapping by the EnSite X EP System to achieve precise lesion creation. A prospective premarketing trial involving 32 patients revealed, an acute procedural success rate of 99.2% of this system with no major AEs [69]. Generally, balloon inflation mechanically removes the residual blood between electrodes and myocardium, theoretically reducing the hemolysis risk. However, current evidence remains limited to pulmonary vein isolation, with no data addressing the safety profile of the system during mitral/tricuspid isthmus ablation. This critical knowledge gap is probably being addressed in the ongoing VOLT-AF IDE clinical trial (NCT06223789), in which 435 patients were enrolled to comprehensively evaluate the system’s safety and efficacy among broader ablation targets.

5.2 Focal Linear PFA Ablation Catheter

Next-generation focal ablation catheters, using a point-by-point ablation strategy, demonstrate improved safety profiles due to the ameliorated tissue contact. The CENTAURI system (Galvanize Therapeutics) currently represents the only clinically available single-tip linear catheter. A clinical study demonstrates promising outcomes, with 100% acute procedural success in complex atrial tachycardia cases and 73% freedom from atrial tachyarrhythmias at 6-month follow-up (82% for paroxysmal AF, 68% for persistent AF) [70]. As regards energy-specific complications, analysis of two studies involving 32 isthmus ablation procedures (15 anterior mitral lines, 2 lateral MI ablations, and 15 CTI ablations) revealed only one case of nitroglycerin-responsive right coronary artery spasm [71, 72]. Despite these preliminary results suggest favorable safety characteristics, a larger-scale investigation is needed to comprehensively evaluate energy-specific adverse events associated with focal PFA catheter because of the limitations posed by the small cohort size in current prospective studies.

5.3 PFA/RFA Catheter

The Sphere-9 catheter (Medtronic) is the first clinically available dual-energy system, features a unique catheter tip design consisting of a 9 mm diameter expandable NiTi-based lattice electrode with a spherical surface containing nine microelectrodes (each 0.7 mm in diameter). This lattice head end design allows for a larger single ablation area than conventional linear catheters.

A recent randomized trial demonstrated the noninferiority of Sphere-9 to RFA [73]. Adverse event analysis revealed transient phrenic nerve injury (0.9%) and bradycardia in the PFA group, versus hematuria (0.5%) in the radiofrequency group. However, the unique pulsed-field energy waveform of the Sphere-9 catheter, combined with its catheter design, increases the anisotropy of its emitted electric field, potentially contributing to coronary spasm mediated by atypical ablation sites. Del Monte A et al. [74] report that the Sphere-9 catheter induces progressive ST-segment increase in the inferior leads during the creation of a posterior MI line positioned immediately below the mitral annulus and subsequent coronary angiography reveals the subocclusion of the distal left circumflex coronary artery at a site corresponding to the prior pulsed-field application. The absence of routine pre-procedural nitroglycerin administration in this case may have increased the risk of coronary spasm, this observation emphasizes the need to recognize catheter-mediated coronary spasm occurring in atypical sites.

The ThermoCool SmartTouch Surroundflow-Dual-Energy catheter (Biosense Webster) uses advanced contact force monitoring and real-time 3D mapping abilities to quantitatively predict lesion depth with high accuracy in a preclinical model. This catheter represents a shift in ablation technology as the integrated PFA/RFA catheter system is able to perform real-time energy impact assessment through quantitative parameter analysis. This technological advancement enables precise control the distribution of PEF, potentially minimizing collateral tissue effects. Preliminary evidence from the SmartfIRE study demonstrated favorable safety profiles, with no coronary spasm events reported among 12 patients who underwent CTI ablation (n = 6 PFA-only, n = 6 RF-only) [75]. The unique abilities of the system, including quantitative lesion assessment through exponential modeling, alternative energy source availability, and stable tissue contact, may significantly reduce energy-specific AEs by optimizing the delivery parameters of PEF and minimizing unnecessary energy exposure.

5.4 Spherical Multielectrode Array PFA

Most of the first-generation PFA systems used in clinical practice lack the detection of real-time pressure, relying instead on fluoroscopic guidance combined with electroanatomical mapping for catheter positioning. This approach risks causing suboptimal tissue apposition, potentially resulting in unnecessary energy delivery through incorrectly positioned electrodes. The new map-and-ablate spherical array PFA catheter (Kardium Inc.) is ameliorated by protocols of contact force-guided selective activation protocols. This system integrates 122 gold-plated electrodes among 16 articulating ribs, synchronized with a proprietary Globe Pulsed Field System to enable the real-time assessment of tissue adherence assessment and selective electrode activation. A premarketing clinical study reveals an acceptable safety profile of the catheter demonstrated acceptable safety profile, as 11 patients who underwent extended ablation protocols (pulmonary vein isolation plus posterior wall and mitral isthmus ablation), did not show any instances of coronary spasm [76]. Our hypothesize that catheters incorporating this design may mitigate hemolysis by minimizing unnecessary discharges in the atrial blood pool. Such targeted energy delivery can reduce erythrocyte damage, though empirical validation through larger-scale studies should be performed.

5.5 Nanosecond PFA Technique

Nanosecond pulsed-field ablation (nsPFA) represents a transformative design in parameter optimization, distinct from conventional millisecond and microsecond PFA in its mechanism of action [77, 78, 79, 80]. This modality induces targeted apoptosis through intracellular effect while avoiding thermal effects and muscle contractions [78]. The underlying mechanism involves high-frequency nanosecond pulses causing sustained membrane potential distortion, preventing the restoration of Ca2+ concentration restoration during interstimulus intervals. This leads to the inhibition of voltage-gated sodium/calcium channel and subsequent nanoelectroporation of sarcolemmal and sarcoplasmic reticulum membranes [81, 82, 83, 84]. Although no clinical experiences in the use of nsPFA for cardiac ablation are available, preclinical studies demonstrate exceptional systemic tolerance. The CellFX Percutaneous Electrode System achieves therapeutic levels three times above clinical thresholds in porcine liver, kidney, and skeletal muscle models, with no significant alterations in urinalysis or serum chemistry [85]. The development of the CellFX Nano-PFA 360 Catheter Endocardial Ablation System (Pulse Biosciences) marks a milestone in nsPFA technology, with ongoing first-in-human trials (NCT06696170) evaluating the safety and efficacy in patients with paroxysmal AF.

6. Conclusions

PFA carries distinct energy-specific complications. Although standardized management protocols for these AEs remain undefined, the amelioration of a comprehensive strategy that integrates preoperative patient assessment, intraoperative precautions, and controlled energy delivery significantly mitigates serious risks. Next-generation PFA catheters featuring enhanced tip designs, optimized discharge parameters, and advanced ablation modalities offer substantial potential for a further reduction of AE incidence. Procedural safety may be established by detailed perioperative protocols specifically tailored to energy-specific AEs, requiring synergistic integration with evolving PFA systems.

References

[1]

Rottner L, Metzner A. Atrial Fibrillation Ablation: Current Practice and Future Perspectives. Journal of Clinical Medicine. 2023; 12: 7556. https://doi.org/10.3390/jcm12247556.

[2]

Chun KRJ, Miklavčič D, Vlachos K, Bordignon S, Scherr D, Jais P, et al. State-of-the-art pulsed field ablation for cardiac arrhythmias: ongoing evolution and future perspective. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2024; 26: euae134. https://doi.org/10.1093/europace/euae134.

[3]

Ekanem E, Neuzil P, Reichlin T, Kautzner J, van der Voort P, Jais P, et al. Safety of pulsed field ablation in more than 17,000 patients with atrial fibrillation in the MANIFEST-17K study. Nature Medicine. 2024; 30: 2020–2029. https://doi.org/10.1038/s41591-024-03114-3.

[4]

Futela P, Kowlgi GN, DeSimone CV, Killu AM, Siontis KC, Noseworthy PA, et al. Early insights on adverse events associated with PulseSelect™ and FARAPULSE™: analysis of the MAUDE database. Journal of Interventional Cardiac Electrophysiology: an International Journal of Arrhythmias and Pacing. 2024. https://doi.org/10.1007/s10840-024-01935-3. (online ahead of print)

[5]

Aldaas OM, Malladi C, Han FT, Hoffmayer KS, Krummen D, Ho G, et al. Pulsed field ablation versus thermal energy ablation for atrial fibrillation: a systematic review and meta-analysis of procedural efficiency, safety, and efficacy. Journal of Interventional Cardiac Electrophysiology: an International Journal of Arrhythmias and Pacing. 2024; 67: 639–648. https://doi.org/10.1007/s10840-023-01660-3.

[6]

Zhang H, Zhang H, Lu H, Mao Y, Chen J. Meta-analysis of pulsed-field ablation versus cryoablation for atrial fibrillation. Pacing and Clinical Electrophysiology: PACE. 2024; 47: 603–613. https://doi.org/10.1111/pace.14971.

[7]

Iqbal M, Kamarullah W, Pranata R, Putra ICS, Karwiky G, Achmad C, et al. Meta-analysis of Pulsed Field Ablation Versus Thermal Ablation for Pulmonary Vein Isolation in AF: A Broad Overview Focusing on Efficacy, Safety and Outcomes. Arrhythmia & Electrophysiology Review. 2024; 13: e13. https://doi.org/10.15420/aer.2024.05.

[8]

Qamar U, Agarwal S, Krishan S, Deshmukh A, DeSimone CV, Stavrakis S, et al. Efficacy and safety of pulsed field ablation for atrial fibrillation: A systematic review and meta-analysis. Pacing and Clinical Electrophysiology: PACE. 2024; 47: 474–480. https://doi.org/10.1111/pace.14947.

[9]

de Campos MCAV, Moraes VRY, Daher RF, Micheleto JPC, de Campos LAV, Barros GFA, et al. Pulsed-field ablation versus thermal ablation for atrial fibrillation: A meta-analysis. Heart Rhythm O2. 2024; 5: 385–395. https://doi.org/10.1016/j.hroo.2024.04.012.

[10]

Rudolph I, Mastella G, Bernlochner I, Steger A, von Olshausen G, Hahn F, et al. Efficacy and safety of pulsed field ablation compared to cryoballoon ablation in the treatment of atrial fibrillation: a meta-analysis. European Heart Journal Open. 2024; 4: oeae044. https://doi.org/10.1093/ehjopen/oeae044.

[11]

Amin AM, Nazir A, Abuelazm MT, Ibrahim AA, Elbenawi H, Aboutaleb A, et al. Efficacy and safety of pulsed-field versus conventional thermal ablation for atrial fibrillation: A systematic review and meta-analysis. Journal of Arrhythmia. 2024; 40: 1059–1074. https://doi.org/10.1002/joa3.13118.

[12]

Waseem MH, Abideen ZU, Ahmed A, Sajid B, Cheema AH, Ramzan NUH, et al. Pulsed Field Versus High-Power Short-Duration Radiofrequency Ablation in Atrial Fibrillation: A Meta-Analysis. Pacing and Clinical Electrophysiology: PACE. 2025; 48: 402–413. https://doi.org/10.1111/pace.15166.

[13]

Xue J, Huang Q, Yu F, Mao Y. Meta-Analysis of Pulsed-Field Ablation Versus- High-Power Short-Duration Ablation for Atrial Fibrillation. Pacing and Clinical Electrophysiology: PACE. 2025; 48: 180–191. https://doi.org/10.1111/pace.15141.

[14]

Kaddoura R, Abushanab D, Asaad N. Pulsed-field ablation versus radiofrequency or cryoballoon thermal ablation in atrial fibrillation: a systematic review and meta-analysis. Postgraduate Medical Journal. 2025; 101: 690–703. https://doi.org/10.1093/postmj/qgaf049.

[15]

Turagam MK, Neuzil P, Schmidt B, Reichlin T, Neven K, Metzner A, et al. Safety and Effectiveness of Pulsed Field Ablation to Treat Atrial Fibrillation: One-Year Outcomes From the MANIFEST-PF Registry. Circulation. 2023; 148: 35–46. https://doi.org/10.1161/CIRCULATIONAHA.123.064959.

[16]

Schmidt B, Bordignon S, Neven K, Reichlin T, Blaauw Y, Hansen J, et al. EUropean real-world outcomes with Pulsed field ablatiOn in patients with symptomatic atRIAl fibrillation: lessons from the multi-centre EU-PORIA registry. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2023; 25: euad185. https://doi.org/10.1093/europace/euad185.

[17]

Cho MS, Lee SR, Black-Maier E, Jackson KP, Friedman DJ, Pokorney SD, et al. Complications associated with pulsed field ablation vs radiofrequency catheter ablation of atrial fibrillation. Heart Rhythm. 2024. https://doi.org/10.1016/j.hrthm.2024.10.063. (in press)

[18]

van Es R, Groen MHA, Stehouwer M, Doevendans PA, Wittkampf FHM, Neven K. In vitro analysis of the origin and characteristics of gaseous microemboli during catheter electroporation ablation. Journal of Cardiovascular Electrophysiology. 2019; 30: 2071–2079. https://doi.org/10.1111/jce.14091.

[19]

Shiomi S, Tokuda M, Takato U, Sakurai R, Yamazaki Y, Matsumoto T, et al. Real-time detection of microembolic signals using carotid echocardiography: A comparative study of 3 pulsed-field systems. Heart Rhythm. 2025. https://doi.org/10.1016/j.hrthm.2025.06.046. (in press)

[20]

Forleo GB, Della Rocca DG, Lavalle C, Mantica M, Papavasileiou LP, Ribatti V, et al. A Patient With Asymptomatic Cerebral Lesions During AF Ablation: How Much Should We Worry? Journal of Atrial Fibrillation. 2016; 8: 1323. https://doi.org/10.4022/jafib.1323.

[21]

van Zyl M, Khabsa M, Tri JA, Ladas TP, Yasin OZ, Ladejobi AO, et al. Open-chest Pulsed Electric Field Ablation of Cardiac Ganglionated Plexi in Acute Canine Models. The Journal of Innovations in Cardiac Rhythm Management. 2022; 13: 5061–5069. https://doi.org/10.19102/icrm.2022.130704.

[22]

Musikantow DR, Neuzil P, Petru J, Koruth JS, Kralovec S, Miller MA, et al. Pulsed Field Ablation to Treat Atrial Fibrillation: Autonomic Nervous System Effects. JACC. Clinical Electrophysiology. 2023; 9: 481–493. https://doi.org/10.1016/j.jacep.2022.10.028.

[23]

Tohoku S, Schmidt B, Schaack D, Bordignon S, Hirokami J, Chen S, et al. Impact of Pulsed-Field Ablation on Intrinsic Cardiac Autonomic Nervous System After Pulmonary Vein Isolation. JACC. Clinical Electrophysiology. 2023; 9: 1864–1875. https://doi.org/10.1016/j.jacep.2023.05.035.

[24]

Xin Y, Wei Z, Wang SS, Wu LL, Wang Y, Wang GY. Preventive effect of atropine premedication on vagal reflex in patients undergoing suspension laryngoscopy during anesthesia induction. Zhonghua Yi Xue Za Zhi. 2023; 103: 1892–1896. https://doi.org/10.3760/cma.j.cn112137-20221124-02488. (In Chinese)

[25]

Giannopoulos G, Kossyvakis C, Angelidis C, Panagopoulou V, Tsiachris D, Vrachatis DA, et al. Coincidental ganglionated plexus modification during radiofrequency pulmonary vein isolation and post-ablation arrhythmia recurrence. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2017; 19: 1967–1972. https://doi.org/10.1093/europace/euw309.

[26]

Qin M, Liu X, Jiang WF, Wu SH, Zhang XD, Po S. Vagal response during pulmonary vein isolation: Re-recognized its characteristics and implications in lone paroxysmal atrial fibrillation. International Journal of Cardiology. 2016; 211: 7–13. https://doi.org/10.1016/j.ijcard.2016.02.116.

[27]

Hu F, Zheng L, Liu S, Shen L, Liang E, Ding L, et al. Avoidance of Vagal Response During Circumferential Pulmonary Vein Isolation: Effect of Initiating Isolation From Right Anterior Ganglionated Plexi. Circulation. Arrhythmia and Electrophysiology. 2019; 12: e007811. https://doi.org/10.1161/CIRCEP.119.007811.

[28]

Del Monte A, Della Rocca DG, Pannone L, Vetta G, Cespón Fernández M, Marcon L, et al. Pulsed field ablation of the right superior pulmonary vein prevents vagal responses via anterior right ganglionated plexus modulation. Heart Rhythm. 2024; 21: 780–787. https://doi.org/10.1016/j.hrthm.2024.01.040.

[29]

Katsumata N, Shimokawa H, Seto M, Kozai T, Yamawaki T, Kuwata K, et al. Enhanced myosin light chain phosphorylations as a central mechanism for coronary artery spasm in a swine model with interleukin-1beta. Circulation. 1997; 96: 4357–4363. https://doi.org/10.1161/01.cir.96.12.4357.

[30]

Shimokawa H, Tomoike H, Nabeyama S, Yamamoto H, Ishii Y, Tanaka K, et al. Coronary artery spasm induced in miniature swine: angiographic evidence and relation to coronary atherosclerosis. American Heart Journal. 1985; 110: 300–310. https://doi.org/10.1016/0002-8703(85)90148-6.

[31]

Kueffer T, Bordignon S, Neven K, Blaauw Y, Hansen J, Adelino R, et al. Durability of Pulmonary Vein Isolation Using Pulsed-Field Ablation: Results From the Multicenter EU-PORIA Registry. JACC. Clinical Electrophysiology. 2024; 10: 698–708. https://doi.org/10.1016/j.jacep.2023.11.026.

[32]

Reddy VY, Petru J, Funasako M, Kopriva K, Hala P, Chovanec M, et al. Coronary Arterial Spasm During Pulsed Field Ablation to Treat Atrial Fibrillation. Circulation. 2022; 146: 1808–1819. https://doi.org/10.1161/CIRCULATIONAHA.122.061497.

[33]

Gunawardene MA, Hartmann J, Tigges E, Jezuit J, Willems S. Word of caution: clinically apparent coronary spasm following pulsed field cavotricuspid isthmus ablation despite nitroglycerin prophylaxis - a case report. European Heart Journal. Case Reports. 2024; 8: ytae553. https://doi.org/10.1093/ehjcr/ytae553.

[34]

Malyshev Y, Neuzil P, Petru J, Funasako M, Hala P, Kopriva K, et al. Nitroglycerin to Ameliorate Coronary Artery Spasm During Focal Pulsed-Field Ablation for Atrial Fibrillation. JACC. Clinical Electrophysiology. 2024; 10: 885–896. https://doi.org/10.1016/j.jacep.2023.12.015.

[35]

Hishikari K, Kuwahara T, Takahashi A, Isobe M. Severe coronary artery spasm during radiofrequency ablation for atrial fibrillation. International Journal of Cardiology. 2014; 172: e513–e515. https://doi.org/10.1016/j.ijcard.2014.01.090.

[36]

Yamashita E, Tada H, Tadokoro K, Hashimoto T, Kaseno K, Miyaji K, et al. Left atrial catheter ablation promotes vasoconstriction of the right coronary artery. Pacing and Clinical Electrophysiology: PACE. 2007; 30: S98–S102. Erratum in: Pacing and Clinical Electrophysiology: PACE. 2007; 30: 1426. https://doi.org/10.1111/j.1540-8159.2007.00615.x.

[37]

Blessberger H, Kammler J, Wichert-Schmitt B, Steinwender C. Exercise-induced vasospastic angina after left atrial catheter ablation: a case report. Journal of Electrocardiology. 2013; 46: 649–652. https://doi.org/10.1016/j.jelectrocard.2013.07.015.

[38]

Fujiwara R, Yoshida A, Hirata KI. Ventricular fibrillation associated with multi-vessel coronary spasms following radiofrequency ablation of atrial fibrillation and atrial flutter. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2014; 16: 1060. https://doi.org/10.1093/europace/eut404.

[39]

Hasegawa K, Miyazaki S, Kaseno K, Tada H. Late-onset lethal arrhythmia after catheter ablation. Journal of Cardiovascular Electrophysiology. 2018; 29: 485–486. https://doi.org/10.1111/jce.13409.

[40]

Nakamura T, Takami M, Fukuzawa K, Kiuchi K, Kono H, Kobori A, et al. Incidence and Characteristics of Coronary Artery Spasms Related to Atrial Fibrillation Ablation Procedures - Large-Scale Multicenter Analysis. Circulation Journal: Official Journal of the Japanese Circulation Society. 2021; 85: 264–271. https://doi.org/10.1253/circj.CJ-20-1096.

[41]

Holda MK, Koziej M, Holda J, Tyrak K, Piatek K, Krawczyk-Ozóg A, et al. Spatial relationship of blood vessels within the mitral isthmus line. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2018; 20: 706–711. https://doi.org/10.1093/europace/euw423.

[42]

Gasperetti A, Assis F, Tripathi H, Suzuki M, Gonuguntla A, Shah R, et al. Determinants of acute irreversible electroporation lesion characteristics after pulsed field ablation: the role of voltage, contact, and adipose interference. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2023; 25: euad257. https://doi.org/10.1093/europace/euad257.

[43]

Zhang C, Neuzil P, Petru J, Funasako M, Hala P, Kopriva K, et al. Coronary Artery Spasm During Pulsed Field vs Radiofrequency Catheter Ablation of the Mitral Isthmus. JAMA Cardiology. 2024; 9: 72–77. https://doi.org/10.1001/jamacardio.2023.4405.

[44]

Tam MTK, Chan JYS, Chan CP, Wu EB, Lai A, Au ACK, et al. Effect of Pulsed-Field Ablation on Human Coronary Arteries: A Longitudinal Study With Intracoronary Imaging. JACC. Clinical Electrophysiology. 2025; 11: 1478–1488. https://doi.org/10.1016/j.jacep.2025.03.014.

[45]

Peeters D, Woelders E, Jansen T, Konst R, Crooijmans C, van de Hoef T, et al. Association Between Coronary Artery Spasm and Atherosclerotic Disease. JACC. Cardiovascular Imaging. 2025; 18: 226–239. https://doi.org/10.1016/j.jcmg.2024.05.024.

[46]

Kinosita K, Jr, Tsong TT. Hemolysis of human erythrocytes by transient electric field. Proceedings of the National Academy of Sciences of the United States of America. 1977; 74: 1923–1927. https://doi.org/10.1073/pnas.74.5.1923.

[47]

Chang SK. Electric-field-induced volume and membrane ionic permeability changes of red blood cells. IEEE Transactions on Bio-medical Engineering. 1993; 40: 1054–1059. https://doi.org/10.1109/10.247804.

[48]

Goto S, Hosojima M, Kabasawa H, Arai K, Takemoto K, Aoki H, et al. Megalin-related mechanism of hemolysis-induced acute kidney injury and the therapeutic strategy. The Journal of Pathology. 2024; 263: 315–327. https://doi.org/10.1002/path.6284.

[49]

De Smet MAJ, François C, De Becker B, Tavernier R, le Polain de Waroux JB, Knecht S, et al. Intravascular haemolysis and acute kidney injury following atrial fibrillation ablation: a report using two different systems for pulsed field ablation. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2024; 26: euae251. https://doi.org/10.1093/europace/euae251.

[50]

Popa MA, Venier S, Menè R, Della Rocca DG, Sacher F, Derval N, et al. Characterization and Clinical Significance of Hemolysis After Pulsed Field Ablation for Atrial Fibrillation: Results of a Multicenter Analysis. Circulation. Arrhythmia and Electrophysiology. 2024; 17: e012732. https://doi.org/10.1161/CIRCEP.124.012732.

[51]

Venier S, Vaxelaire N, Jacon P, Carabelli A, Desbiolles A, Garban F, et al. Severe acute kidney injury related to haemolysis after pulsed field ablation for atrial fibrillation. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2023; 26: euad371. https://doi.org/10.1093/europace/euad371.

[52]

Osmancik P, Bacova B, Herman D, Hozman M, Fiserova I, Hassouna S, et al. Periprocedural Intravascular Hemolysis During Atrial Fibrillation Ablation: A Comparison of Pulsed Field With Radiofrequency Ablation. JACC. Clinical Electrophysiology. 2024; 10: 1660–1671. https://doi.org/10.1016/j.jacep.2024.05.001.

[53]

Stojadinović P, Ventrella N, Alfredová H, Wichterle D, Peichl P, Čihák R, et al. Prediction of major intravascular hemolysis during pulsed electric field ablation of atrial fibrillation using a pentaspline catheter. Journal of Cardiovascular Electrophysiology. 2024; 35: 2405–2410. https://doi.org/10.1111/jce.16468.

[54]

Jordan F, Knecht S, Isenegger C, Arnet R, Krisai P, Völlmin G, et al. Acute kidney injury after catheter ablation of atrial fibrillation: Comparison between different energy sources. Heart Rhythm. 2024; 21: 1248–1249. https://doi.org/10.1016/j.hrthm.2024.04.046.

[55]

Mohanty S, Casella M, Compagnucci P, Torlapati PG, Della Rocca DG, La Fazia VM, et al. Acute Kidney Injury Resulting From Hemoglobinuria After Pulsed-Field Ablation in Atrial Fibrillation: Is it Preventable? JACC. Clinical Electrophysiology. 2024; 10: 709–715. https://doi.org/10.1016/j.jacep.2023.12.008.

[56]

Nies M, Koruth JS, Mlček M, Watanabe K, Tibenská VC, Královec Š et al. Hemolysis After Pulsed Field Ablation: Impact of Lesion Number and Catheter-Tissue Contact. Circulation. Arrhythmia and Electrophysiology. 2024; 17: e012765. https://doi.org/10.1161/CIRCEP.124.012765.

[57]

Mattison L, Verma A, Tarakji KG, Sigg DC. Hemolysis After Pulsed Field Ablation: The Role of Dose and Contact in an Acute Porcine Model. Circulation. Arrhythmia and Electrophysiology. 2024; 17: e013317. https://doi.org/10.1161/CIRCEP.124.013317.

[58]

Fiserova I, Fiser O, Novak M, Trnka J, Gibalova A, Kvapil D, et al. Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro. Heart Rhythm. 2025; 22: 466–474. https://doi.org/10.1016/j.hrthm.2024.08.019.

[59]

Yuan J, Wang Z, Yu Y, Peng H, Shi S, He F, et al. The Protective Impact of Gd-DTPA on Hemolysis in Pulsed Field Ablation for Atrial Fibrillation. JACC. Clinical Electrophysiology. 2024; 10: 2741–2743. https://doi.org/10.1016/j.jacep.2024.09.028.

[60]

La Fazia VM, Mohanty S, Torlapati PG, Gianni C, Della Rocca DG, Bode W, et al. Hydration to Prevent Kidney Injury After Pulsed Field Ablation: Importance of Timing and Fluids Amount. JACC. Clinical Electrophysiology. 2025; 11: 599–601. https://doi.org/10.1016/j.jacep.2024.11.001.

[61]

Tamirisa KP, Sanchez JE, La Fazia VM, Gianni C, Serpa F, Mohanty S, et al. Pulsed field ablation-related hemoglobinuria and acute kidney injury: Insights and strategies for effective management. American Heart Journal. 2025; 290: 339–346. https://doi.org/10.1016/j.ahj.2025.07.013.

[62]

Ji X, Zhang H, Zang L, Yan S, Wu X. The Effect of Discharge Mode on the Distribution of Myocardial Pulsed Electric Field-A Simulation Study for Pulsed Field Ablation of Atrial Fibrillation. Journal of Cardiovascular Development and Disease. 2022; 9: 95. https://doi.org/10.3390/jcdd9040095.

[63]

Reddy VY, Gerstenfeld EP, Natale A, Whang W, Cuoco FA, Patel C, et al. Pulsed Field or Conventional Thermal Ablation for Paroxysmal Atrial Fibrillation. The New England Journal of Medicine. 2023; 389: 1660–1671. https://doi.org/10.1056/NEJMoa2307291.

[64]

Reddy VY, Gerstenfeld EP, Schmidt B, Nair D, Natale A, Saliba W, et al. Pulsed Field Ablation for Persistent Atrial Fibrillation: 1-Year Results of ADVANTAGE AF. Journal of the American College of Cardiology. 2025; 85: 1664–1678. https://doi.org/10.1016/j.jacc.2025.03.515.

[65]

Reddy VY, Gerstenfeld EP, Schmidt B, Andrade JG, Nair D, Natale A, et al. Pulsed Field Ablation of Persistent Atrial Fibrillation With Continuous Electrocardiographic Monitoring Follow-Up: ADVANTAGE AF Phase 2. Circulation. 2025; 152: 27–40. https://doi.org/10.1161/CIRCULATIONAHA.125.074485.

[66]

Pierucci N, La Fazia VM, Mohanty S, Schiavone M, Doty B, Gabrah K, et al. Results of ICE-Guided Isolation of the Superior Vena Cava With Pulsed Field Ablation. JACC. Clinical Electrophysiology. 2025; 11: 752–760. https://doi.org/10.1016/j.jacep.2024.11.009.

[67]

Kueffer T, Seiler J, Madaffari A, Mühl A, Asatryan B, Stettler R, et al. Pulsed-field ablation for the treatment of left atrial reentry tachycardia. Journal of Interventional Cardiac Electrophysiology: an International Journal of Arrhythmias and Pacing. 2023; 66: 1431–1440. https://doi.org/10.1007/s10840-022-01436-1.

[68]

Hirokami J, Chun KRJ, Bordignon S, Tohoku S, Neven K, Reichlin T, et al. Pulsed-Field Ablation for Persistent Atrial Fibrillation in EU-PORIA Registry. Journal of Cardiovascular Electrophysiology. 2025; 36: 1710–1720. https://doi.org/10.1111/jce.16583.

[69]

Sanders P, Healy S, Emami M, Kotschet E, Miller A, Kalman JM. Initial clinical experience with the balloon-in-basket pulsed field ablation system: acute results of the VOLT CE mark feasibility study. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2024; 26: euae118. https://doi.org/10.1093/europace/euae118.

[70]

Wenzel JP, Lemoine MD, Rottner L, My I, Moser F, Obergassel J, et al. Nonthermal Point-by-Point Pulmonary Vein Isolation Using a Novel Pulsed Field Ablation System. Circulation. Arrhythmia and Electrophysiology. 2023; 16: e012093. https://doi.org/10.1161/CIRCEP.123.012093.

[71]

Weyand S, Adam V, Biehler P, Hägele P, Hanger S, Heinzmann D, et al. Focal Pulsed Field Ablation for Atrial Arrhythmias: Efficacy and Safety under Deep Sedation. Journal of Clinical Medicine. 2024; 13: 576. https://doi.org/10.3390/jcm13020576.

[72]

Erhard N, Englert F, Prommersberger S, Popa M, Bourier F, Reents T, et al. Focal pulsed field ablation in complex atrial tachycardia: First clinical experience and 1-year outcome. Heart Rhythm. 2025; 22: 57–64. https://doi.org/10.1016/j.hrthm.2024.07.024.

[73]

Anter E, Mansour M, Nair DG, Sharma D, Taigen TL, Neuzil P, et al. Dual-energy lattice-tip ablation system for persistent atrial fibrillation: a randomized trial. Nature Medicine. 2024; 30: 2303–2310. https://doi.org/10.1038/s41591-024-03022-6.

[74]

Del Monte A, Della Rocca DG, Pannone L, Vetta G, Doundoulakis I, Cespón Fernández M, et al. Coronary Vasospasm During Isthmus Pulsed Field Ablation With Wide Area Focal Catheter. Circulation. Arrhythmia and Electrophysiology. 2024; 17: e012923. https://doi.org/10.1161/CIRCEP.124.012923.

[75]

Duytschaever M, Račkauskas G, De Potter T, Hansen J, Knecht S, Phlips T, et al. Dual energy for pulmonary vein isolation using dual-energy focal ablation technology integrated with a three-dimensional mapping system: SmartfIRE 3-month results. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2024; 26: euae088. https://doi.org/10.1093/europace/euae088.

[76]

Turagam MK, Neuzil P, Petru J, Funasako M, Koruth JS, Skoda J, et al. AF ablation using a novel “single-shot” map-and-ablate spherical array pulsed field ablation catheter: 1-Year outcomes of the first-in-human PULSE-EU trial. Heart Rhythm. 2024; 21: 1218–1226. https://doi.org/10.1016/j.hrthm.2024.04.102.

[77]

Sano MB, Arena CB, DeWitt MR, Saur D, Davalos RV. In-vitro bipolar nano- and microsecond electro-pulse bursts for irreversible electroporation therapies. Bioelectrochemistry (Amsterdam, Netherlands). 2014; 100: 69–79. https://doi.org/10.1016/j.bioelechem.2014.07.010.

[78]

Sugrue A, Maor E, Del-Carpio Munoz F, Killu AM, Asirvatham SJ. Cardiac ablation with pulsed electric fields: principles and biophysics. Europace: European Pacing, Arrhythmias, and Cardiac Electrophysiology: Journal of the Working Groups on Cardiac Pacing, Arrhythmias, and Cardiac Cellular Electrophysiology of the European Society of Cardiology. 2022; 24: 1213–1222. https://doi.org/10.1093/europace/euac033.

[79]

Tekle E, Oubrahim H, Dzekunov SM, Kolb JF, Schoenbach KH, Chock PB. Selective field effects on intracellular vacuoles and vesicle membranes with nanosecond electric pulses. Biophysical Journal. 2005; 89: 274–284. https://doi.org/10.1529/biophysj.104.054494.

[80]

Schoenbach KH, Beebe SJ, Buescher ES. Intracellular effect of ultrashort electrical pulses. Bioelectromagnetics. 2001; 22: 440–448. https://doi.org/10.1002/bem.71.

[81]

Semenov I, Xiao S, Pakhomova ON, Pakhomov AG. Recruitment of the intracellular Ca2+ by ultrashort electric stimuli: the impact of pulse duration. Cell Calcium. 2013; 54: 145–150. https://doi.org/10.1016/j.ceca.2013.05.008.

[82]

Semenov I, Xiao S, Pakhomov AG. Primary pathways of intracellular Ca(2+) mobilization by nanosecond pulsed electric field. Biochimica et Biophysica Acta. 2013; 1828: 981–989. https://doi.org/10.1016/j.bbamem.2012.11.032.

[83]

Nesin V, Pakhomov AG. Inhibition of voltage-gated Na+ current by nanosecond pulsed electric field (nsPEF) is not mediated by Na+ influx or Ca2+ signaling. Bioelectromagnetics. 2012; 33: 443–451. http://doi.org/10.1002/bem.21703.

[84]

Nesin V, Bowman AM, Xiao S, Pakhomov AG. Cell permeabilization and inhibition of voltage-gated Ca(2+) and Na(+) channel currents by nanosecond pulsed electric field. Bioelectromagnetics. 2012; 33: 394–404. https://doi.org/10.1002/bem.21696.

[85]

Nuccitelli R, Litt J, Moss K, Hinman D, Getsin I, Uecker D, et al. The CellFX Percutaneous Electrode System for Nanosecond Pulsed Field Ablation. Bioelectricity. 2024; 6: 174–180. https://doi.org/10.1089/bioe.2024.0015.

PDF (7014KB)

0

Accesses

0

Citation

Detail

Sections
Recommended

/