Invasive Versus Echocardiographic Transvalvular Pressure Gradients After Transcatheter Aortic Valve Implantation: A Long-Term Retrospective Cohort
James A. Brown , Eishan Ashwat , Michel Pompeu Sá , Andres Martinez Plotnikow , Sarah Yousef , Nishant Agrawal , Dustin Kliner , Catalin Toma , Amber Makani , Yisi Wang , Floyd W. Thoma , Derek Serna-Gallegos , Ibrahim Sultan
The Heart Surgery Forum ›› 2026, Vol. 29 ›› Issue (2) : 50776
Evidence regarding the reliability of transthoracic echocardiography (TTE) in assessing valvular gradients immediately following transcatheter aortic valve implantation (TAVI) remains conflicting. Thus, this study aimed to compare post-procedural TTE and invasive mean aortic valve (AV) pressure gradients immediately after TAVI.
This was a retrospective, single-institution cohort study of consecutive TAVIs between 2012 and 2023. Immediate post-procedural mean AV pressure gradients were measured invasively via pigtail catheter and non-invasively via TTE. The Spearman coefficient was used to assess the correlation between TTE and invasive gradients. Multivariable Cox proportional-hazards regression was performed for long-term survival.
A total of 1589 patients underwent TAVI with available TTE and invasive pressure gradients. A total of 49.2% received self-expanding valves (SEVs), and 50.8% received balloon-expanding valves (BEVs); 7.2% underwent valve-in-valve (ViV); 17.6% received a small valve (Evolut ≤26 mm, Sapien ≤23 mm, and Portico/Navitor ≤25 mm). For the entire cohort, the TTE and invasive mean gradients showed a moderate correlation (Spearman ρ = 0.401), with a median absolute difference of 2.2 [1.0–4.0] mmHg. The SEV gradients were more strongly correlated than the BEV gradients (ρ = 0.447 vs. 0.345). Similarly, the small valve gradients were more strongly correlated than the large valve gradients (ρ = 0.455 vs. 0.375), while the ViV gradients were more strongly correlated than the native TAVI gradients (ρ = 0.575 vs. 0.357). A total of 1.6% of the patients had a difference >10 mmHg between the invasive and TTE gradients. In the multivariable Cox regression, a >10 mmHg discordance was not significantly associated with an increased hazard of death after TAVI (p = 0.326).
Following TAVI, we observed a moderate correlation between TTE and invasive measurements of mean AV pressure gradients, with only modest discordance noted between measurement modalities.
invasive AV gradient / TTE / transvalvular AV gradient / TAVI
| [1] |
Thyregod HGH, Ihlemann N, Jørgensen TH, Nissen H, Kjeldsen BJ, Petursson P, et al. Five-Year Clinical and Echocardiographic Outcomes From the NOTION Randomized Clinical Trial in Patients at Lower Surgical Risk. Circulation. 2019; 139: 2714–2723. https://doi.org/10.1161/CIRCULATIONAHA.118.036606. |
| [2] |
Leon MB, Mack MJ, Hahn RT, Thourani VH, Makkar R, Kodali SK, et al. Outcomes 2 Years After Transcatheter Aortic Valve Replacement in Patients at Low Surgical Risk. Journal of the American College of Cardiology. 2021; 77: 1149–1161. https://doi.org/10.1016/j.jacc.2020.12.052. |
| [3] |
Yousef S, Bianco V, Kliner D, Toma C, Serna-Gallegos D, West D, et al. Outcomes of Transcatheter Aortic Valve Replacement in Patients With Concomitant Aortic Regurgitation. The Annals of Thoracic Surgery. 2023; 116: 728–734. https://doi.org/10.1016/j.athoracsur.2023.02.008. |
| [4] |
Baumgartner H, Hung J, Bermejo J, Chambers JB, Edvardsen T, Goldstein S, et al. Recommendations on the Echocardiographic Assessment of Aortic Valve Stenosis: A Focused Update from the European Association of Cardiovascular Imaging and the American Society of Echocardiography. Journal of the American Society of Echocardiography: Official Publication of the American Society of Echocardiography. 2017; 30: 372–392. https://doi.org/10.1016/j.echo.2017.02.009. |
| [5] |
Leon MB, Piazza N, Nikolsky E, Blackstone EH, Cutlip DE, Kappetein AP, et al. Standardized endpoint definitions for transcatheter aortic valve implantation clinical trials: a consensus report from the Valve Academic Research Consortium. European Heart Journal. 2011; 32: 205–217. https://doi.org/10.1093/eurheartj/ehq406. |
| [6] |
Baumgartner H, Khan S, DeRobertis M, Czer L, Maurer G. Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an in vitro study of normal St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves. Journal of the American College of Cardiology. 1992; 19: 324–332. https://doi.org/10.1016/0735-1097(92)90486-7. |
| [7] |
Abbas AE, Mando R, Hanzel G, Gallagher M, Safian R, Hanson I, et al. Invasive Versus Echocardiographic Evaluation of Transvalvular Gradients Immediately Post-Transcatheter Aortic Valve Replacement. Circulation. Cardiovascular Interventions. 2019; 12: e007973. https://doi.org/10.1161/CIRCINTERVENTIONS.119.007973. |
| [8] |
Abbas AE, Khalili H, Madanat L, Elmariah S, Shannon F, Al-Azizi K, et al. Echocardiographic Versus Invasive Aortic Valve Gradients in Different Clinical Scenarios. Journal of the American Society of Echocardiography: Official Publication of the American Society of Echocardiography. 2023; 36: 1302–1314. https://doi.org/10.1016/j.echo.2023.06.016. |
| [9] |
Zoghbi WA, Jone PN, Chamsi-Pasha MA, Chen T, Collins KA, Desai MY, et al. Guidelines for the Evaluation of Prosthetic Valve Function With Cardiovascular Imaging: A Report From the American Society of Echocardiography Developed in Collaboration With the Society for Cardiovascular Magnetic Resonance and the Society of Cardiovascular Computed Tomography. Journal of the American Society of Echocardiography: Official Publication of the American Society of Echocardiography. 2024; 37: 2–63. https://doi.org/10.1016/j.echo.2023.10.004. |
| [10] |
Herrmann HC, Pibarot P, Wu C, Hahn RT, Tang GHL, Abbas AE, et al. Bioprosthetic Aortic Valve Hemodynamics: Definitions, Outcomes, and Evidence Gaps: JACC State-of-the-Art Review. Journal of the American College of Cardiology. 2022; 80: 527–544. https://doi.org/10.1016/j.jacc.2022.06.001. |
| [11] |
Herrmann HC, Mehran R, Blackman DJ, Bailey S, Möllmann H, Abdel-Wahab M, et al. Self-Expanding or Balloon-Expandable TAVR in Patients with a Small Aortic Annulus. The New England Journal of Medicine. 2024; 390: 1959–1971. https://doi.org/10.1056/NEJMoa2312573. |
| [12] |
Rodés-Cabau J, Ribeiro HB, Mohammadi S, Serra V, Al-Atassi T, Iñiguez A, et al. Transcatheter or Surgical Aortic Valve Replacement in Patients With Severe Aortic Stenosis and Small Aortic Annulus: A Randomized Clinical Trial. Circulation. 2024; 149: 644–655. https://doi.org/10.1161/CIRCULATIONAHA.123.067326. |
| [13] |
Hatle L, Angelsen BA, Tromsdal A. Non-invasive assessment of aortic stenosis by Doppler ultrasound. British Heart Journal. 1980; 43: 284–292. https://doi.org/10.1136/hrt.43.3.284. |
| [14] |
Faggiano P, Ghizzoni G, Sorgato A, Sabatini T, Simoncelli U, Gardini A, et al. Rate of progression of valvular aortic stenosis in adults. The American Journal of Cardiology. 1992; 70: 229–233. https://doi.org/10.1016/0002-9149(92)91280-h. |
| [15] |
Saikrishnan N, Kumar G, Sawaya FJ, Lerakis S, Yoganathan AP. Accurate assessment of aortic stenosis: a review of diagnostic modalities and hemodynamics. Circulation. 2014; 129: 244–253. https://doi.org/10.1161/CIRCULATIONAHA.113.002310. |
| [16] |
Bach DS, Schmitz C, Dohmen G, Aaronson KD, Steinseifer U, Kleine P. In vitro assessment of prosthesis type and pressure recovery characteristics: Doppler echocardiography overestimation of bileaflet mechanical and bioprosthetic aortic valve gradients. The Journal of Thoracic and Cardiovascular Surgery. 2012; 144: 453–458. https://doi.org/10.1016/j.jtcvs.2011.12.036. |
| [17] |
Mascherbauer J, Schima H, Maurer G, Baumgartner H. Doppler assessment of mechanical aortic valve prostheses: effect of valve design and size of the aorta. The Journal of Heart Valve Disease. 2004; 13: 823–830. |
| [18] |
Aalaei-Andabili SH, Park KE, Choi CY, Manning EW, Stinson WW, Van Woerkom R, et al. Relationship between Invasive and Echocardiographic Transvalvular Gradients after Transcatheter Aortic Valve Replacement. Cardiology and Therapy. 2020; 9: 201–206. https://doi.org/10.1007/s40119-020-00161-y. |
| [19] |
Abdel-Wahab M, Mehilli J, Frerker C, Neumann FJ, Kurz T, Tölg R, et al. Comparison of balloon-expandable vs self-expandable valves in patients undergoing transcatheter aortic valve replacement: the CHOICE randomized clinical trial. JAMA. 2014; 311: 1503–1514. https://doi.org/10.1001/jama.2014.3316. |
| [20] |
Johnson NP, Zelis JM, Tonino PAL, Houthuizen P, Bouwman RA, Brueren GRG, et al. Pressure gradient vs. flow relationships to characterize the physiology of a severely stenotic aortic valve before and after transcatheter valve implantation. European Heart Journal. 2018; 39: 2646–2655. https://doi.org/10.1093/eurheartj/ehy126. |
/
| 〈 |
|
〉 |