Opportunities to reduce the radiation exposure during computed tomography to assess the changes in the lungs in patients with COVID-19: use of adaptive statistical iterative reconstruction

Daria A. Filatova , Valentin E. Sinitsin , Elena A. Mershina

Digital Diagnostics ›› 2021, Vol. 2 ›› Issue (2) : 94 -104.

PDF (1154KB)
Digital Diagnostics ›› 2021, Vol. 2 ›› Issue (2) :94 -104. DOI: 10.17816/DD62477
Original Study Articles
research-article

Opportunities to reduce the radiation exposure during computed tomography to assess the changes in the lungs in patients with COVID-19: use of adaptive statistical iterative reconstruction

Author information +
History +
PDF (1154KB)

Abstract

BACKGROUND: Several COVID-19 patients are subjected to multiple imaging examinations during hospitalization, the cumulative effect of which can significantly increase the total dose of radiation received. The effective radiation dose can be reduced by lowering the current and voltage of the X-ray tube, but this reduces image quality. One possible solution is to use adaptive statistical iterative reconstruction technology on the «raw» CT data. Recently, data on the efficacy of low-dose CT (LDCT) in the diagnosis of COVID-19 have appeared in the literature.

AIM: To analyze the quality and diagnostic value of LDCT images of the lungs after applying an iterative processing algorithm and to assess the possibility of reducing the radiation load on the patient when diagnosing COVID-19.

MATERIALS AND METHODS: Patients from the Infectious Diseases Department of the Moscow State University Hospital participated in the prospective study. CT examinations were performed at the time of patient admission and discharge and were repeated as needed during hospitalization. In the first study, a standard CT protocol with a tube voltage of 120 kV and automatic current modulation in the range of 200–400 mA was used; in repeated CT scans, the LDCT protocol was used with reduced tube voltage parameters (100 or 110 kV) and automatic current modulation in the range of 40–120 mA. To assess the diagnostic value of LDCT in comparison with standard CT, a survey was conducted among doctors from the Department of Radiation Diagnostics at Moscow State University Hospital. The questionnaire included a comparison of the two methods for identifying the following pathological processes: «ground-glass» opacities, compaction of the lung tissue with reticular changes, areas of lung tissue consolidation, and lymphadenopathy.

RESULTS: The study included 151 patients. The average age was 58±14.2 years, with men accounting for 53.6% of the population. During LDCT the radiation load was reduced by 2.96 times on average, CTDI by 2.6 times, DLP by 3.1 times, the current on the tube by 1.83 times, and the voltage on the tube by 1.2 times. The results indicate that the effectiveness of detecting the main signs of viral pneumonia and assessing the dynamics of the patient’s condition does not differ significantly from CT performed according to the standard protocol.

CONCLUSIONS: The results of a comparison of standard and low-dose CT show that there is no significant loss of diagnostic information and image quality as the radiation load is reduced. Thus, chest LDCT can be used to successfully diagnose COVID-19 in routine practice.

Keywords

COVID-19 / X-ray computed tomography / lung / radiation protection / SARS-CoV-2

Cite this article

Download citation ▾
Daria A. Filatova, Valentin E. Sinitsin, Elena A. Mershina. Opportunities to reduce the radiation exposure during computed tomography to assess the changes in the lungs in patients with COVID-19: use of adaptive statistical iterative reconstruction. Digital Diagnostics, 2021, 2(2): 94-104 DOI:10.17816/DD62477

登录浏览全文

4963

注册一个新账户 忘记密码

References

[1]

Ministry of Health of the Russian Federation. Temporary guidelines: prevention, diagnosis and treatment of new coronavirus infection. Version 8 (03.09.2020). Moscow; 2020. (In Russ). Available from: https://base.garant.ru/74596434/

[2]

Министерство здравоохранения Российской Федерации. Временные методические рекомендации: профилактика, диагностика и лечение новой коронавирусной инфекции. Версия 8 (03.09.2020). Москва, 2020. Режим доступа: https://base.garant.ru/74596434/. Дата обращения: 14.03.2021.

[3]

Ministry of Health of the Russian Federation. Temporary guidelines: prevention, diagnosis and treatment of new coronavirus infection. Version 8 (03.09.2020). Moscow; 2020. (In Russ). Available from: https://base.garant.ru/74596434/

[4]

Romanov BK. Coronavirus infection COVID-19. Safety and Risc of Farmacotherapy. 2020;8(1):3–8. (In Russ.)

[5]

Романов Б.К. Коронавирусная инфекция COVID-2019//Безопасность и риск фармакотерапии. 2020. Т. 8, № 1. С. 3–8. doi: 10.30895/2312-7821-2020-8-1-3-8

[6]

Romanov BK. Coronavirus infection COVID-19. Safety and Risc of Farmacotherapy. 2020;8(1):3–8. (In Russ.)

[7]

Morozov SP, Protsenko DN, Smetanina SV, et al. Radiation diagnostics of coronavirus disease (COVID-19): organization, methodology, interpretation of results: preprint No. CDT-Version 2 of 17.04.2020. Moscow; 2020. 78 р. (In Russ.)

[8]

Морозов С.П., Проценко Д.Н., Сметанина С.В., и др. Лучевая диагностика коронавирусной болезни (COVID-19): организация, методология, интерпретация результатов: препринт № ЦДТ – Версия 2 от 17.04.2020. Москва; 2020. 78 с.

[9]

Morozov SP, Protsenko DN, Smetanina SV, et al. Radiation diagnostics of coronavirus disease (COVID-19): organization, methodology, interpretation of results: preprint No. CDT-Version 2 of 17.04.2020. Moscow; 2020. 78 р. (In Russ.)

[10]

Udugama B, Kadhiresan P, Kozlowski HN, et al. Diagnosing COVID-19: the disease and tools for detection. ACS Nano. 2020;14(4):3822–3835. doi: 10.1021/acsnano.0c02624

[11]

Udugama B., Kadhiresan P., Kozlowski H.N., et al. Diagnosing COVID-19: The disease and tools for detection//ACS Nano. 2020. Vol. 14, N 4. P. 3822–3835. doi: 10.1021/acsnano.0c02624.

[12]

Udugama B, Kadhiresan P, Kozlowski HN, et al. Diagnosing COVID-19: the disease and tools for detection. ACS Nano. 2020;14(4):3822–3835. doi: 10.1021/acsnano.0c02624

[13]

Zhao W, Zhong Z, Xie X, et al. Relation between chest ct findings and clinical conditions of coronavirus disease (COVID-19) pneumonia: a multicenter study. AJR Am J Roentgenol. 2020;214(5):1072–1077. doi: 10.2214/AJR.20.22976

[14]

Zhao W., Zhong Z., Xie X., et al. Relation between chest ct findings and clinical conditions of coronavirus disease (COVID-19) pneumonia: a multicenter study//AJR Am J Roentgenol. 2020. Vol. 214, N 5. P. 1072–1077. doi: 10.2214/AJR.20.22976

[15]

Zhao W, Zhong Z, Xie X, et al. Relation between chest ct findings and clinical conditions of coronavirus disease (COVID-19) pneumonia: a multicenter study. AJR Am J Roentgenol. 2020;214(5):1072–1077. doi: 10.2214/AJR.20.22976

[16]

Beregi JP, Greffier J. Low and ultra-low dose radiation in CT: Opportunities and limitations. Diagn Interv Imaging. 2019;100(2):63–64. doi: 10.1016/j.diii.2019.01.007

[17]

Beregi J.P., Greffier J. Low and ultra-low dose radiation in CT: Opportunities and limitations//Diagn Interv Imaging. 2019. Vol. 100, N 2. P. 63–64. doi: 10.1016/j.diii.2019.01.007.

[18]

Beregi JP, Greffier J. Low and ultra-low dose radiation in CT: Opportunities and limitations. Diagn Interv Imaging. 2019;100(2):63–64. doi: 10.1016/j.diii.2019.01.007

[19]

Cheng L, Chen Y, Fang T, et al. Fast iterative adaptive reconstruction in low-dose CT imaging. In: 2006 International Conference on Image Processing. Atlanta, GA: IEEE; 2006. P. 889–892. Available from: https://ieeexplore.ieee.org/document/4106673/

[20]

Cheng L., Fang T., Tyan J. Fast Iterative adaptive reconstruction in low-dose CT imaging//2006 International Conference on Image Processing. Atlanta: GA: IEEE; 2006. P. 889–892. Режим доступа: https://ieeexplore.ieee.org/document/4106673/. Дата обращения: 14.03.2021.

[21]

Cheng L, Chen Y, Fang T, et al. Fast iterative adaptive reconstruction in low-dose CT imaging. In: 2006 International Conference on Image Processing. Atlanta, GA: IEEE; 2006. P. 889–892. Available from: https://ieeexplore.ieee.org/document/4106673/

[22]

Hara AK, Paden RG, Silva AC, et al. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol. 2009;193(3):764–771. doi: 10.2214/AJR.09.2397

[23]

Hara A.K., Paden R.G., Silva A.C., et al. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study//AJR Am J Roentgenol. 2009. Vol. 193, N 3. P. 764–771. doi: 10.2214/AJR.09.2397.

[24]

Hara AK, Paden RG, Silva AC, et al. Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol. 2009;193(3):764–771. doi: 10.2214/AJR.09.2397

[25]

Prakash P, Kalra M, Kambadakone A, et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol. 2010;45(4):202–210. doi: 10.1097/RLI.ob013e3181dzfeec

[26]

Prakash P., Kalra M., Kambadakone A., et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique//Invest Radiol. 2010. Vol. 45, N 4. P. 202–210. doi: 10.1097/RLI.ob013e3181dzfeec

[27]

Prakash P, Kalra M, Kambadakone A, et al. Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol. 2010;45(4):202–210. doi: 10.1097/RLI.ob013e3181dzfeec

[28]

Chen LG, Wu PA, Sheu MH, et al. Automatic current selection with iterative reconstruction reduces effective dose to less than 1 mSv in low-dose chest computed tomography in persons with normal BMI. Medicine (Baltimore). 2019;98(28):e16350. doi: 10.1097/MD.0000000000016350

[29]

Chen L.G., Wu P.A., Sheu M.H., et al. Automatic current selection with iterative reconstruction reduces effective dose to less than 1 mSv in low-dose chest computed tomography in persons with normal BMI//Medicine (Baltimore). 2019. Vol. 98, N 28. P. e16350. doi: 10.1097/MD.0000000000016350.

[30]

Chen LG, Wu PA, Sheu MH, et al. Automatic current selection with iterative reconstruction reduces effective dose to less than 1 mSv in low-dose chest computed tomography in persons with normal BMI. Medicine (Baltimore). 2019;98(28):e16350. doi: 10.1097/MD.0000000000016350

[31]

Dangis A, Gieraerts C, De Brueker Y, et al. Accuracy and reproducibility of low-dose submillisievert chest CT for the diagnosis of COVID-19. Radiology Cardiothoracic Imaging. 2020;2(2):e200196. doi: 10.1148/ryct.2020200196

[32]

Dangis A., Gieraerts C., De Brueker Y., et al. Accuracy and reproducibility of low-dose submillisievert chest CT for the diagnosis of COVID-19//Radiology Cardiothoracic Imaging. 2020. Vol. 2, N 2. P. e200196. doi: 10.1148/ryct.2020200196

[33]

Dangis A, Gieraerts C, De Brueker Y, et al. Accuracy and reproducibility of low-dose submillisievert chest CT for the diagnosis of COVID-19. Radiology Cardiothoracic Imaging. 2020;2(2):e200196. doi: 10.1148/ryct.2020200196

[34]

Sethuraman N, Jeremiah SS, Ryo A. Interpreting diagnostic tests for SARS-CoV-2. JAMA. 2020;323(22):2249–2251. doi: 10.1001/jama.2020.8259

[35]

Sethuraman N., Jeremiah S.S., Ryo A. Interpreting diagnostic tests for SARS-CoV-2//JAMA. 2020. Vol. 323, N 22. P. 2249–2251. doi: 10.1001/jama.2020.8259.

[36]

Sethuraman N, Jeremiah SS, Ryo A. Interpreting diagnostic tests for SARS-CoV-2. JAMA. 2020;323(22):2249–2251. doi: 10.1001/jama.2020.8259

[37]

Long C, Xu H, Shen Q, et al. Diagnosis of the Coronavirus disease (COVID-19): rRT-PCR or CT? Eur J Radiol. 2020;126:108961. doi: 10.1016/j.ejrad.2020.108961

[38]

Long C., Xu H., Shen Q., et al. Diagnosis of the Coronavirus disease (COVID-19): rRT-PCR or CT?//Eur J Radiology. 2020. Vol. 126. P. 108961. doi: 10.1016/j.ejrad.2020.108961

[39]

Long C, Xu H, Shen Q, et al. Diagnosis of the Coronavirus disease (COVID-19): rRT-PCR or CT? Eur J Radiol. 2020;126:108961. doi: 10.1016/j.ejrad.2020.108961

[40]

Fang Y, Zhang H, Xie J, et al. Sensitivity of chest CT for COVID-19: comparison to RT-PCR. Radiology. 2020;296(2):E115–E117. doi: 10/1148/radiol.2020200432

[41]

Fang Y., Zhang H., Xie J., et al. Sensitivity of chest CT for COVID-19: comparison to RT-PCR//Radiology. 2020. Vol. 296, N 2. P. E115–E117. doi: 10/1148/radiol.2020200432

[42]

Fang Y, Zhang H, Xie J, et al. Sensitivity of chest CT for COVID-19: comparison to RT-PCR. Radiology. 2020;296(2):E115–E117. doi: 10/1148/radiol.2020200432

[43]

Yang Y, Yang M, Shen C, et al. Evaluating the accuracy of different respiratory specimens in the laboratory diagnosis and monitoring the viral shedding of 2019-nCoV infections. medRxiv. 2020. doi: 10.1101/2020.02.11.20021493

[44]

Yang Y., Yang M., Shen C., et al. Evaluating the accuracy of different respiratory specimens in the laboratory diagnosis and monitoring the viral shedding of 2019-nCoV infections//medRxiv. 2020. doi: 10.1101/2020.02.11.20021493

[45]

Yang Y, Yang M, Shen C, et al. Evaluating the accuracy of different respiratory specimens in the laboratory diagnosis and monitoring the viral shedding of 2019-nCoV infections. medRxiv. 2020. doi: 10.1101/2020.02.11.20021493

[46]

Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277–2284. doi: 10.1056/NEJMra072149

[47]

Brenner D.J., Hall E.J. Computed tomography — an increasing source of radiation exposure//N Engl J Med. 2007. Vol. 357, N 22. P. 2277–2284. doi: 10.1056/NEJMra072149

[48]

Brenner DJ, Hall EJ. Computed tomography--an increasing source of radiation exposure. N Engl J Med. 2007;357(22):2277–2284. doi: 10.1056/NEJMra072149

[49]

Pierce DA, Preston DL. Radiation-related cancer risks at low doses among atomic bomb survivors. Radiat Res. 2000;154(2):178–186. doi: 10.1667/0033-7587(2000)154[0178:rrcral]2.0.co;2

[50]

Pierce D.A., Preston D.L. Radiation-related cancer risks at low doses among atomic bomb survivors//Radiat Res. 2000. Vol. 154, N 2. P. 178–186. doi: 10.1667/0033-7587(2000)154[0178:rrcral]2.0.co;2

[51]

Pierce DA, Preston DL. Radiation-related cancer risks at low doses among atomic bomb survivors. Radiat Res. 2000;154(2):178–186. doi: 10.1667/0033-7587(2000)154[0178:rrcral]2.0.co;2

[52]

Matkevich E, Sinitsyn V, Mershina E. Comparative analysis of radiation doses of patients with computed tomography in a Federal medical institution. Journal of Radiology and Nuclear Medicine. 2016;97(1):33–39. (In Russ). doi: 10.20862/0042-4676-2016-97-1-33-40

[53]

Маткевич Е.И., Синицын В.Е., Мершина Е.А. Сравнительный анализ доз облучения пациентов при компьютерной томографии в федеральном лечебном учреждении//Вестник рентгенологии и радиологии. 2016. Т. 97, № 1. С. 33–39. doi: 10.20862/0042-4676-2016-97-1-33-40

[54]

Matkevich E, Sinitsyn V, Mershina E. Comparative analysis of radiation doses of patients with computed tomography in a Federal medical institution. Journal of Radiology and Nuclear Medicine. 2016;97(1):33–39. (In Russ). doi: 10.20862/0042-4676-2016-97-1-33-40

[55]

Naidich DP, Marshall CH, Gribbin C, et al. Low-dose CT of the lungs: preliminary observations. Radiology. 1990;175(3):729–731. doi: 10.1148/radiology.175.3.2343122

[56]

Naidich D.P., Marshall C., Gribbin C., et al. Low-dose CT of the lungs: preliminary observations//Radiology. 1990. Vol. 175, N 3. P. 729–731. doi: 10.1148/radiology.175.3.2343122

[57]

Naidich DP, Marshall CH, Gribbin C, et al. Low-dose CT of the lungs: preliminary observations. Radiology. 1990;175(3):729–731. doi: 10.1148/radiology.175.3.2343122

[58]

Prasad SR, Wittram C, Shepard JA, et al. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality. AJR Am J Roentgenol. 2002;179(2):461–465. doi: 10.2214/ajr.179.2.1790461

[59]

Prasad S.R., Wittram C., Sherard J.A., et al. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality//AJR Am J Roentgenol. 2002. Vol. 179, N 2. P. 461–465. doi: 10.2214/ajr.179.2.1790461

[60]

Prasad SR, Wittram C, Shepard JA, et al. Standard-dose and 50%-reduced-dose chest CT: comparing the effect on image quality. AJR Am J Roentgenol. 2002;179(2):461–465. doi: 10.2214/ajr.179.2.1790461

[61]

Zwirewich CV, Mayo JR, Müller NL. Low-dose high-resolution CT of lung parenchyma. Radiology. 1991;180(2):413–417. doi: 10.1148/radiology.180.2.2068303

[62]

Zwirewich C.V., Mayo J.R., Müller N.L. Low-dose high-resolution CT of lung parenchyma//Radiology. 1991. Vol. 180, N 2. P. 413–417. doi: 10.1148/radiology.180.2.2068303.

[63]

Zwirewich CV, Mayo JR, Müller NL. Low-dose high-resolution CT of lung parenchyma. Radiology. 1991;180(2):413–417. doi: 10.1148/radiology.180.2.2068303

[64]

Zhu X, Yu J, Huang Z. Low-dose chest CT: optimizing radiation protection for patients. AJR Am J Roentgenol. 2004;183(3):809–816. doi: 10.2214/ajr.183.3.1830809

[65]

Zhu X., Yu J., Huang Z. Low-dose chest CT: optimizing radiation protection for patients//AJR Am J Roentgenol. 2004. Vol. 183, N 3. P. 809–816. doi: 10.2214/ajr.183.3.1830809

[66]

Zhu X, Yu J, Huang Z. Low-dose chest CT: optimizing radiation protection for patients. AJR Am J Roentgenol. 2004;183(3):809–816. doi: 10.2214/ajr.183.3.1830809

[67]

Kubo T, Ohno Y, Takenaka D, et al. Standard-dose vs. low-dose CT protocols in the evaluation of localized lung lesions: Capability for lesion characterization – iLEAD study. Eur J Radiol Open. 2016;3:67–73. doi: 10.1016/j.ejro.2016.03.002

[68]

Kubo T., Ohno Y., Takenaka D., et al. Standard-dose vs. low-dose CT protocols in the evaluation of localized lung lesions: Capability for lesion characterization – iLEAD study//Eur J Radiol Open. 2016. Vol. 3. P. 67–73. doi: 10.1016/j.ejro.2016.03.002

[69]

Kubo T, Ohno Y, Takenaka D, et al. Standard-dose vs. low-dose CT protocols in the evaluation of localized lung lesions: Capability for lesion characterization – iLEAD study. Eur J Radiol Open. 2016;3:67–73. doi: 10.1016/j.ejro.2016.03.002

[70]

Gombolevsky VA, Chernina VY, Blokhin IA. Main achievements of low-dose computed tomography in lung cancer screening. Tuberculosis and Lung Diseases. 2021;99(1):61–70. (In Russ). doi: 10.21292/2075-1230-2021-99-1-61-7025.

[71]

Гомболевский В.А., Чернина В.Ю., Блохин И.А. Основные достижения низкодозной компьютерной томографии в скрининге рака легкого//Туберкулез и болезни легких. 2021. Т. 99. № 1. С. 61–70. doi: 10.21292/2075-1230-2021-99-1-61-70

[72]

Gombolevsky VA, Chernina VY, Blokhin IA. Main achievements of low-dose computed tomography in lung cancer screening. Tuberculosis and Lung Diseases. 2021;99(1):61–70. (In Russ). doi: 10.21292/2075-1230-2021-99-1-61-7025.

[73]

Rampinelli C, De Marco P, Origgi D, et al. Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ. 2017;356:j347. doi: 10.1136/bmj.j347

[74]

Rampinelli C., De Marco P., Origgi D., et al. Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis//BMJ. 2017. Vol. 356. P. j347. doi: 10.1136/bmj.j347

[75]

Rampinelli C, De Marco P, Origgi D, et al. Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ. 2017;356:j347. doi: 10.1136/bmj.j347

[76]

Chiles C. Lung cancer screening with low dose CT. Radiol Clin North Am. 2014;52(1):27–46. doi: 10.1016/j.rcl.2013.08.006.

[77]

Chiles C. Lung cancer screening with low dose CT//Radiol Clin North Am. 2014. Vol. 52, N 1. P. 27–46. doi: 10.1016/j.rcl.2013.08.006.

[78]

Chiles C. Lung cancer screening with low dose CT. Radiol Clin North Am. 2014;52(1):27–46. doi: 10.1016/j.rcl.2013.08.006.