Experimental Analysis of Strength Characteristics of Devices for Acetabular Osteosynthesis
E. I Solod , A. F Lazarev , N. S Gavryushenko , L. V Fomin , I. N Sakharnykh , S. S Stoyukhin
N.N. Priorov Journal of Traumatology and Orthopedics ›› 2015, Vol. 22 ›› Issue (1) : 91 -94.
Three common elementary acetabular fractures, i.e. transverse fracture, fractures of anterior and posterior column, were simulated on plastic pelvis models. Comparative testing of three types of osteosynthesis - with reconstructive plate, cannulated screws and pins or using our new CITO flattened fine thread screws 3 mm in diameter were performed under similar conditions. Strength of cannulated screw and CITO screw was evaluated. Testing was performed on servo-hydraulic universal testing machine. The most stable fixation was achieved with cannulated screw and CITO screw. The latter was inferior in strength however ensured required rigid fixation.
acetabular fracture / testing / rigidity / metal fixator
| [1] |
Judet R., Judet J. Leturnel E. Fracture of Acetabulum classification and surgical approaches for open reduction. J. Bone Joint Surg. 1964; 46 (8): 1615-46. |
| [2] |
Солод Э.И., Лазарев А.Ф., Лазарев А.А., Гудушаури Я.Г., Какабадзе М.Г., Роскидайло А.С., Дан И.М. Возможности оперативного лечения переломов вертлужной впадины с использованием малоинвазивных технологий. Вестник травматологии и ортопедии им. Н.Н. Приорова. 2009; 2: 3-9. |
| [3] |
Лазарев А.Ф. Оперативное лечение повреждений таза: Автореф. дис. … д-ра мед. наук. М.; 1992. |
| [4] |
Лазарев А.Ф., Солод Э.И. Способ закрытого остеосинтеза переломов вертлужной впадины. Патент RU № 2327427 от 27.06.2008. |
| [5] |
Gary J.L., Lefaivre K.A., Gerold F., Hay M.T., Reinert C.M., Starr A.J. Survivorship of the native hip joint after percutaneous repair of acetabular fractures in the elderly. Injury. 2011; 42 (10): 1144-51. |
| [6] |
Mouhsine E., Garofalo R., Borens O., Wettstein M., Blanc C.H., Fischer J.F. et al. Percutaneous retrograde screwing for stabilisation of acetabular fractures. Injury. 2005; 36 (11): 1330-6. |
| [7] |
Starr A.J., Jones A.L., Reinert C.M., Borer D.S. Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001; 32: 45-50. |
| [8] |
Gay S.B., Sistrom C., Wang G.J., Kahler D.A., Boman T., McHugh N., Goitz H.T. Percutaneous screw fixation of acetabular fractures with CT guidance: preliminary results of a new technique. Am. J. Roentgenol. 1992; 158 (4): 819-22. |
| [9] |
Attias N., Lindsey R.W., Starr A.J., Borer D., Bridges K., Hipp J.A. The use of a virtual three-dimensional model to evaluate the intraosseous space available for percutaneous screw fixation of acetabular fractures. J. Bone Joint Surg. Br. 2005; 87 (11): 1520-3. |
| [10] |
Chen K.N., Wang G., Cao L.G., Zhang M.C. Differences of percutaneous retrograde screw fixation of anterior column acetabular fractures between male and female: A study of 164 virtual three-dimensional models. Injury. 2009; 40 (10): 1067-72. |
| [11] |
Rommens P.M. Is there a role for percutaneous pelvic and acetabular reconstruction? Injury. 2007; 38: 463-77. |
| [12] |
Starr A.J., Reinert C.M., Jones A.L. Percutaneous fixation of the columns of the acetabulum: a new technique. J. Orthop. Trauma. 1998; 12 (1): 51-8. |
| [13] |
Routt M.L., Simonia P.T., Grujic L. The retrograde medullary superior pubic ramus screw for the treatment of anterior pelvic ring disruptions: a new technique. J. Orthop. Trauma. 1995; 9: 35-44. |