Features of general surgical operations in patients with implanted pacemakers
Aleksandr Evgen'evich Tyagunov , E V Pervova , T V Nechay , A N Aleksandrov , K A Rogov , A D Zlotnikova , A M Zhdanov , A V Sazhin , A E Tyagunov , E V Pervova , T V Nechai , A N Aleksandrov , K A Rogov , A D Zlotnikova , A M Zhdanov , A V Sazhin
Russian Medicine ›› 2011, Vol. 17 ›› Issue (5) : 22 -28.
Features of general surgical operations in patients with implanted pacemakers
The operation of an implanted pacemaker was evaluated in 99 patients aged 62.4 ± 9.6 years during non-cardiac operations. Electrocoagulation (EC) caused inhibited EC stimuli in 9 (9.1%) patients, short-term episodes of ineffective stimulation with impaired synchronization in 2 (2%), and an altered stimulation regimen in 2 (2%). The abnormalities were short-term and depended on EC duration. Four patients were found to have EC-unassociated episodes of myopotential inhibition. The revealed abnormalities require meticulous patient examination prior to surgery, intraoperative ECG monitoring, and time-limited use of monopolar EC.
pacemaker / electrocoagulation during surgery / pacemaker disturbances
| [1] |
Базаев В. А. Функциональная диагностика в кардиологии / Под ред. Л. А. Бокерия и др. - М., 2002. - Т. 1. - С. 311-333. |
| [2] |
Бокерия Л. А., Ревишвили А. Ш., Ступаков И. Н., Самородская И. В. // Качественная Клиническая Практика. - 2003. - № 1. - С. 75-82. |
| [3] |
Бокерия Л. А., Ревишвили А. Ш., Дубровский И. А. Состояние электрокоагуляции в России в 2006 году. URL: http://www.electrophysiology.ru/library/article.php?new_ article_id = 23, свободный. - Заглавие с экрана. - Яз. англ. (18.01.2008). |
| [4] |
Григоров С. С., Вотчал Ф. Б., Костылева О. В. Электрокардиограмма при искусственном водителе ритма сердца. - М., 1990. |
| [5] |
Комплекс суточного мониторирования ЭКГ автоматизированный "Astrocard®) - холтеровский анализ" (Astrocard® - Holtersystem-2F for Windows). Паспорт. - М., 1996. |
| [6] |
Костылева О. В., Вотчал Ф. Б., Моргунов В. С. Руководство по применению имплантируемых электрокардиостимуляторов ЭКС-511 и ЭКС-530. - Тверь, 1999. |
| [7] |
Федоров В. Д., Дрогайцев А. Д., Рябов А., Полежаев В. В. // Хирургия. - 1985. - № 8. - С. 116-120. |
| [8] |
Bourke M. E. // Can. J. Anaesth. - 1996. - Vol. 43, N 5, pt 2. - P. R24-R41. |
| [9] |
Caramella J. P., Mentre B., Jattiot F. et al. // Ann. Fr. Anesth. Reanim. - 1987. - Vol. 6, N 3. - P. 214-216. |
| [10] |
Caramella J. P., Aliot E., Claude E. et al. // Ann. Fr. Anesth. Reanim. - 1988. - Vol. 7, N 4. - P. 309-319. |
| [11] |
Caramella J. P., Dodinot B. // Ann. Fr. Anesth. Reanim. - 1989. - Vol. 8, N 3. - P. 290-292. |
| [12] |
Chauvin M., Crenner F., Brechenmacher C. et al. // Arch. Mal. Coeur. - 1990. - Vol. 83, N 7. - P. 997-1002. |
| [13] |
Chauvin M., Crenner F., Brechenmacher C. // Arch. Mal. Coeur. - 1992. - Vol. 85, N 9. - P. 1323-1327. |
| [14] |
Chauvin M., Crenner F., Brechenmacher C. // Pace. - 1992. - Vol. 15. - P. 2028-2033. |
| [15] |
Delhumeau A., Ronceray S., Moreau X. // Ann. Fr. Anesth. Reanim. - 1988. - Vol. 7, N 2. - P. 162-164. |
| [16] |
Dreifus L. S., Fisch C., Griffin J. C. et al. // J. Am. Coll. Cardiol. - 1991. - Vol. 18. - P. 1-13. |
| [17] |
Godin J. F., Petitot J. C. // Pacing Clin. Electrophysiol. - 1989. - Vol. 12, N 6. - P. 1011. |
| [18] |
Mangar D., Atlas G. M., Kane P. B. // Can. J. Anaesth. - 1991. - Vol. 38, N 5. - P. 616-618. |
| [19] |
Peters R. W., Gold M. R. // J. Interv. Card. Electrophysiol. - 1998. - Vol. 2, N 4. - P. 343-344. |
| [20] |
Salukhe T. V., Dob D., Sutton R. et al. // Br. J. Anaesth. - 2004. - Vol. 93, N 1. - P. 95-104. |
| [21] |
Shibata Hil-oshi, Yano M., Okahisa T. et al. // Dig. Endosc. - 1999. - Vol. 11. - P. 150-157. |
| [22] |
Wilson S., Neustein S. N., Camunas J. et al. // Mt Sinai J. Med. - 2006. - Vol. 73, N 6. - P. 880-883. |
Eco-Vector
/
| 〈 |
|
〉 |