Design and implement of the signal generator in the magnetic focused conductivity tomography system

FU Lin, HUANG Kama

PDF(475 KB)
PDF(475 KB)
Front. Electr. Electron. Eng. ›› 2007, Vol. 2 ›› Issue (2) : 230-233. DOI: 10.1007/s11460-007-0042-7

Design and implement of the signal generator in the magnetic focused conductivity tomography system

  • FU Lin, HUANG Kama
Author information +
History +

Abstract

To improve stability and performance of the signal source and sweeping detection, as well as to extract abundant and reliable signal, the direct digital synthesis technology was employed to design the generator of the source which formed sweeping frequencies of sine wave output from 1 to 20 MHz. The planar spiral coil was connected as an amplitude modulation circuit. The same coil adopted differential architecture for signal detection and extraction. The MC1595 was utilized to compose a phase detector in which difference of phases varies with the change of frequencies. A low pass filter was designed to filter the carry waves of the sweeping source. Thereby the system gained abundant data and its stability was improved. Further, the spatial resolution of the system was enhanced. All of the above favors the use of software in the magnetic focused conductivity tomography system (MFCT) to reconstruct the image of conductivity within the human body.

Cite this article

Download citation ▾
FU Lin, HUANG Kama. Design and implement of the signal generator in the magnetic focused conductivity tomography system. Front. Electr. Electron. Eng., 2007, 2(2): 230‒233 https://doi.org/10.1007/s11460-007-0042-7