Studies on injector pump-ultrasonic nebulizer feeding sample system for MPT-AES

Feng Guodong, Han Songbai, Yu Aimin

PDF(376 KB)
PDF(376 KB)
Front. Chem. China ›› 2006, Vol. 1 ›› Issue (2) : 232-237. DOI: 10.1007/s11458-006-0010-4

Studies on injector pump-ultrasonic nebulizer feeding sample system for MPT-AES

  • Feng Guodong, Han Songbai, Yu Aimin
Author information +
History +

Abstract

The feeding-sample system used by microwave plasma torch atomic emission spectroscopy (MPT-AES) is the pneumatic nebulization system; its efficiency, however, is not good. A newly built injector pump ultrasonic nebulizer combination feeding-sample system has been designed. Its performance was tested and compared with that of the pneumatic nebulization system. It can be concluded that the newly built feeding-sample system can increase the spectral line intensity by about two to three times and decrease the detection limit by about 2 to 10 times. Moreover, this newly built system can reduce the time taken washing the sample cell from 30 min or so to about 10 min.