Study and Application of Electrochemical Behavior of Calcium–ARS on a Mercury Film Electrode
Chen Xiaotong, Cui Qianling, Hu Jingbo
Author information+
Department of Chemistry, Beijing Normal University, Beijing 100875, China;
Show less
History+
Published
05 Mar 2006
Issue Date
05 Mar 2006
Abstract
A sensitive complex absorptive wave of Ca ARS was obtained by using differential pulse voltammetry when a mercury film glass carbon electrode was immersed in 0.1 mol L-1 KOH and 4.5?10-4 mol L-1 ARS solution. The peak potential obtained was -1.17 V (vs Ag AgCl). The peak current was proportional to the concentration of calcium in the range of 5.0?10-8 4.2?10-5 mol L-1. The detection limit was 2.0?10-8 mol L-1. This method was applied successfully to determining traces of calcium in blood serum. The electrochemical behavior of the system was also studied by cyclic voltammetry, and the experiment results showed that the electrode process was an irreversible absorptive with two electrons participating.
Chen Xiaotong, Cui Qianling, Hu Jingbo.
Study and Application of Electrochemical Behavior of Calcium–ARS on a Mercury Film Electrode. Front. Chem. China, 2006, 1(1): 102‒107 https://doi.org/10.1007/s11458-005-0018-1