A novel heart pump model was obtained by improving the traditional axial pump design theory with the consideration of working and hydraulic situations for artificial hearts. The pump head range and the velocity triangle were introduced and an iterative approach was utilized for the initial model. Moreover, computational fluid dynamics (CFD) simulations were performed to determine relevant model parameters. The results show that this procedure can be used for designing a series of high-efficiency artificial heart pumps.
WANG Yingpeng, SONG Xinwei, YING Chuntong
. Applications of traditional pump design theory
to artificial heart and CFD simulation[J]. Frontiers in Energy, 2008
, 2(4)
: 504
-507
.
DOI: 10.1007/s11708-008-0059-5
1. Sezai Y . Progressand future perspectives in mechanical circulatory support. Artificial Organs, 2001, 25(5): 318–322. doi:10.1046/j.1525-1594.2001.06804.x
2. Troskolanski A T, Tazarkiewicz S . Calculation and Constructionof Impeller. Beijing: Mechanical Industry Press, 1981 (in Chinese)
3. Stepanoff A J . Centrifugal and Axial Flow Pumps: Theory, Design and Application.2nd ed.Florida: Krieger Publishing Company, 1992
4. Brennen C E . Hydrodynamics of Pumps. VT: Concepts NREC & Oxford University Press, 1994
5. Grirnkoy L P, Parpila A H . Vane Pump. Beijing: National Defense IndustryPress, 1982 (in Chinese)
6. Balje O E . Turbomachines: A Guide to Design, Selection and Theory. New York: John Wiley & Sons Inc, 1981
7. Apel J, Neudel F, Reul H . Computational fluid dynamics and experimental validationof a microaxial blood pump. ASAIO J, 2001, 47(5): 552–558. doi:10.1097/00002480-200109000-00031
8. Apel J, Paul R, Klaus S, et al.. Assessment of hemolysis related quantities ina microaxial blood pump by computational fluid dynamics. Artificial Organs, 2001, 25(5): 341–347. doi:10.1046/j.1525-1594.2001.025005341.x